DELTA XTEND 130760042Recalls & reported problems
Depuy (Ireland) · DELTA Xtend Eccentric Glenosphere Dia 42mm ECC · Class 2 · Implant
Quick summary
This DePuy Delta Xtend shoulder implant is a metal half-shoulder replacement with an eccentric glenosphere used in shoulder arthroplasty. The 66 reports mostly describe the implant coming loose or dislocating, joint dislocation, pain, and suspected infection, along with fractures and broken posts or screws that sometimes left pieces in the bone. Sixty reports describe injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 66
- Deaths reported
- 0
- Injuries reported
- 60
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 66 adverse event reports that list this exact device (DI 10603295027751), received between Aug 2, 2017 and Dec 31, 2025 — classified as 60 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (27 reports).
10 other recalls cover devices cleared under the same FDA submission (K062250, K120174, K192855); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated27
- Adverse Event Without Identified Device or Use Problem24
- Fracture10
- Insufficient Device Problem Information3
- Appropriate Device Problem Term/Code Not Available2
- Break2
- Connection Problem1
- Difficult to Insert1
Patient problems
- No Code Available24
- Joint Dislocation21
- Pain15
- Unspecified Infection13
- Not Applicable6
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
- Joint Laxity3
- Unspecified Tissue Injury3
Recent adverse event reports
The 25 most recent of 66 reports, in the reporter's own words as published by the FDA.
InjuryDec 31, 2025Fracture; Device Dislodged or Dislocated; Fracture; Device Dislodged or Dislocated
IT WAS REPORTED THE PATIENT WAS REVISED DUE TO PAINFUL RIGHT SHOULDER. IT WAS NOTED THE GLENOSPHERE APPEARED TO BE TILTED FROM METAGLENE. THE IMPLANTS WERE REMOVED AND REPLACED. IT WAS NOTED THERE WAS A POSSIBLE INFECTION IN THE SHOULDER AND A POSSIBLE BREAK ON THE POST AT THE BOTTOM OF GLENOSPHERE AND PIECE IN THE METAGLENE POST. DOI: (B)(6) 2019. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Pain; Unspecified Infection
MAUDE report 23935479 ↗InjuryJun 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS INFECTED. THE SURGEON REMOVED THE POLY, GLENOSPHERE, METAGLENE AND SCREWS. THE SURGEON REMOVED 3 LOCKING SCREWS; HOWEVER, THE PRODUCT NUMBERS ARE NOT AVAILABLE. ALSO THE POLY WAS WORN ON THE ANTERIOR FACE SO THE PRODUCT NUMBER OR LOT NUMBER WERE UNREADABLE. THE SURGEON CONVERTED IT TO A CTA HEMI. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Unspecified Infection
MAUDE report 22293255 ↗InjuryJun 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INFECTION. A COMPETITOR'S ANTIBIOTIC SPACER WAS BEING PLACED UNTIL INFECTION CLEARS. DOI: (B)(6) 2014, DOR: (B)(6) 2024 , AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Unspecified Infection
MAUDE report 19616001 ↗InjuryJun 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SUBJECT ID: (B)(6). STUDY NO DOTS. CLINICAL ADVERSE EVENT RECEIVED FOR SUSPICION OF PROSTHETIC INFECTION - LEFT SHOULDER. EVENT IS SERIOUS AND IS CONSIDERED MODERATE. EVENT IS NOT RELATED TO DEVICE OR PROCEDURE. DATE OF IMPLANT: (B)(6) 2024. DATE OF EVENT: (B)(6) 2024. (LEFT SHOULDER). TREATMENT: I&D. NO REVISION COMPLETED.
Patient problems: Unspecified Infection
MAUDE report 19562870 ↗InjuryApr 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
(B)(6). CLINICAL ADVERSE EVENT RECEIVED FOR INFECTION. EVENT IS SERIOUS AND IS CONSIDERED MODERATE. EVENT IS NOT RELATED TO DEVICE OR PROCEDURE. DATE OF IMPLANT: (B)(6) 2024 DATE OF EVENT: (B)(6) 2024 (RIGHT SHOULDER) TREATMENT: INJECTED MEDICATION
Patient problems: Unspecified Infection
MAUDE report 19203137 ↗InjuryMar 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS COMPLAINING OF PAIN. THE SURGEON CONVERTED THE 155 EPIPHYSIS TO 145 AND DOWNSIZED THE GLENOSPHERE TO 38. ALL COMPONENTS WERE WELL FIXED. SURGICAL DELAY WAS UNKNOWN. DOI: (B)(6) 2020 DOR: (B)(6) 2024 AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Pain
MAUDE report 18966078 ↗InjuryFeb 15, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT WAS REVISED DUE TO PAIN. HE HAS HAD MULTIPLE FALLS. NO OTHER INFORMATION. IT WAS UNKNOWN IF THERE WAS ANY SURGICAL DELAY. DOI: UNKNOWN, DOR: ON (B)(6) 2024, AFFECTED SIDE: LEFT.
Patient problems: Fall; Pain
MAUDE report 18715370 ↗MalfunctionFeb 5, 2024Fracture; Fracture
IT WAS REPORTED THAT THE SCREW OF GLENOSPHERE BROKE. SURGICAL DELAY UNKNOWN. DOI: UNKNOWN, DOR: ON (B)(6) 2024, AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Insufficient Information
MAUDE report 18638990 ↗InjuryNov 28, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
SUBJECT ID: (B)(6). STUDY NO: DOTS. CLINICAL ADVERSE EVENT RECEIVED FOR DISLOCATION OF THE REVERSED SHOULDER PROSTHESIS. EVENT IS SERIOUS AND IS CONSIDERED MODERATE. EVENT IS POSSIBLY RELATED TO DEVICE. EVENT IS NOT RELATED TO PROCEDURE. DATE OF IMPLANT: (B)(6) 2023. DATE OF EVENT: 07 NOV 2023. (RIGHT SHOULDER). TREATMENT: CLOSED REDUCTION/MANIPULATION.
Patient problems: Joint Dislocation
MAUDE report 18220597 ↗InjuryNov 27, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
STUDY NO: DOTS. CLINICAL ADVERSE EVENT RECEIVED FOR DISLOCATION OF THE REVERSED SHOULDER PROSTHESIS. EVENT IS SERIOUS AND IS CONSIDERED MODERATE. EVENT IS POSSIBLY RELATED TO DEVICE. EVENT IS NOT RELATED TO PROCEDURE. DATE OF IMPLANT: (B)(6) 2023. DATE OF REVISION: (B)(6) 2023. DATE OF EVENT: 08 NOV 2023. (RIGHT SHOULDER). TREATMENT: REVISION; GLENOSPHERE WAS REVISED.
Patient problems: Joint Dislocation
MAUDE report 18208963 ↗InjuryNov 21, 2023Fracture; Device Dislodged or Dislocated; Fracture; Device Dislodged or Dislocated
ADDITIONAL INFORMATION RECEIVED: A. PLEASE CLARIFY, WAS THE GLENOSPHERE FRACTURED? NO. B. WAS THERE ANY SUGGESTION BY ANYONE INDICATING THAT THERE WAS OR MAY BE A DEFICIENCY WITH THE DXTEND SCREW LOCK D4.5X36MM? IF YES, PLEASE PROVIDE DETAILS. POSSIBLY WASN'T SEATED ALL THE WAY.
Patient problems: Joint Dislocation
MAUDE report 18182566 ↗InjuryJun 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION REVERSE TOTAL SHOULDER REPLACEMENT. PATIENT PRESENTED TO SURGEON WITH A PAINFUL SHOULDER. DECISION WAS MADE TO REVISE THE SHOULDER DUE TO INFERIOR NOTCHING. THE REVISION SURGERY WAS UNDERTAKEN AND THE GLENOSPHERE WAS REMOVED AND REPLACED WITH A LATERALISED GLENOSPHERE. THE POLY LINER WAS REMOVED AND REPLACED WITH A NEW POLY LINER. RANGE OF MOTION WAS PERFORMED AND NO IMPINGEMENT WAS NOTED. DOI: UNKNOWN, DOE: (B)(6) 2023, AFFECTED SIDE: UNKNOWN.
Patient problems: Pain; Unspecified Tissue Injury
MAUDE report 17207945 ↗InjuryFeb 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT PRESENTED WITH INFECTION. SURGEON REMOVED IMPLANTS TO PLACE ANTIBIOTIC SPACER. NOT ABLE TO READ ANY LOT NUMBERS. DOI: UNKNOWN. DOR: (B)(6) 2023. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Unspecified Infection
MAUDE report 16434557 ↗MalfunctionOct 12, 2022Fracture; Fracture
IT WAS REPORTED THAT THE GLENOSPHERE THREAD BROKEN 2X.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15587307 ↗InjuryJun 22, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
LEFT SHOULDER REPLACEMENT REVISION FOR DISLOCATION - SURGEON NOTED THAT IN PRIMARY SHOULDER (COMPLETED BY HIM (B)(6) 2022) THE METAGLENE HAS BEEN POSITIONED QUITE HIGH ON THE FACE OF THE GLENOID, CAUSING THE HUMERAL CUP TO NOTCH ON THE INFERIOR NECK OF GLENOID. SURGEON REMOVED GLENOSPHERE AND HUMERAL CUP, REMOVED EXTRA BONE FROM NECK OF GLENOID AND REPLACED THE GLENOSPHERE AS WELL AS ADDING A RETENTIVE CUP DOI: (B)(6) 2022. DOR: (B)(6) 2022. LEFT SHOULDER.
Patient problems: Joint Dislocation; Unspecified Tissue Injury
MAUDE report 14773036 ↗InjuryMay 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ALL IMPLANTS WERE REMOVED DUE TO CONFIRMED INFECTION OF RIGHT SHOULDER. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2019; DOR: (B)(6) 2022; AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Unspecified Infection
MAUDE report 14381519 ↗InjuryApr 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION COMPLETED DUE TO LOOSENING OF THE METAGLENE AT THE BONE TO IMPLANT INTERFACE. IT WAS ALSO REPORTED THAT THE 12 YEAR OLD DELTA REVERSE BECAME UNSTABLE, POLY WEAR IS PRESENT, AND THIS IS A GLENOID DEFECT. PATIENT CONVERTED INTO A CTA, STEM WAS WELL FIXED. DOI- UNKNOWN, DOR- (B)(6) 2022, AFFECTED SIDE- LEFT SHOULDER.
Patient problems: Joint Laxity
MAUDE report 14136990 ↗InjuryJan 31, 2022Fracture; Migration; Fracture; Migration
THE GLENOSPHERE SCREWING SYSTEM HAD BROKEN RESULTING IN MEDIALIZATION OF THE IMPLANT CAUSING PAIN FOR THE PATIENT AS WELL AS A DECREASE IN SHOULDER MOBILITY. A REVISION HAS BEEN MADE FOR A REPLACEMENT OF PROSTHESIS ON (B)(6) 2022. THE DEVICE WAS IMPLANTED ON 2018. DOI: 2018, DOR: (B)(6) 2022, UNKNOWN SHOULDER.
Patient problems: Loss of Range of Motion; Pain
MAUDE report 13392580 ↗InjuryDec 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE INFECTION WAS CONFIRMED.
Patient problems: Inflammation; Pain; Unspecified Infection
MAUDE report 13117862 ↗InjuryNov 16, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT WAS REVISED DUE TO LOOSENING OF THE UNKNOWN COMPONENT AT BONE TO IMPLANT INTERFACE. PATIENT DISLOCATED HIS REVERSE LTSA MANY MONTHS AGO AND NEVER CAME BACK FOR SURGERY AFTER BEING SEEN IN CLINIC AND WAS TOLD TO COME BACK FOR SURGERY. PATIENT FINALLY SHOWS UP IN CLINIC OVER TWO YEARS LATER AND X-RAYS REVEALS THAT THE METAGLENE, FOUR SCREWS AND GLENOSPHERE HAD DETACHED FROM THE GLENOID AND WAS FREE FLOATING IN THE SHOULDER. REMOVED AND CONVERTED TO A CTA. DOI: (B)(6) 2019. DOR: (B)(6) 2021. UNKNOWN SHOULDER SIDE.
Patient problems: Joint Dislocation
MAUDE report 12820439 ↗InjuryMay 3, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INSTABILITY. DOI: (B)(6) 2019 DOR: (B)(6) 2021 LEFT SHOULDER.
Patient problems: Joint Laxity
MAUDE report 11761726 ↗InjuryApr 13, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A DEPUY SYNTHES SALES CONSULTANT WAS NOTIFIED PRIOR TO THE CASE BY A SURGEON THAT THE PATIENT HAD FALLEN THE PREVIOUS WEEK AND RE-INJURED THE OPERATIVE SHOULDER. SURGEON ADVISED THAT HE PLANNED TO REVISE THE PATIENT'S INJURED SHOULDER AND TREAT FOR POSSIBLE INFECTION. UPON FURTHER INTRAOPERATIVE EVALUATION, SURGEON DISCOVERED THAT THE METAGLENE, 4 LOCKING SCREWS AND THE GLENOSPHERE HAD ALL BECOME DISLODGED FROM THE PATIENT'S GLENOID. SURGEON WAS ABLE TO EXTRACT ALL OF THESE IMPLANTS AS ON UNIT COMPLETELY PUT TOGETHER AND UNDAMAGED. AT THIS POINT, SURGEON CHECKED THE HUMERAL STEM AND EPIPHYSIS WAS DETERMINED THAT THOSE IMPLANTS REMAINED WELL FIXED AND UNDAMAGED. SURGEON REPLACED THE HUMERAL CUP POLY AND FORMED AN ANTIBIOTIC CEMENT SPACER TO SERVE AS A TEMPORARY FIT PROSTHESIS WHILE THE PATIENT'S INFECTION WILL BE TREATED. DUE TO QUESTIONS ABOUT THE STABILITY AND HEALTH OF THE PATIENT'S GLENOID, SURGEON ASKED FOR ASSISTANCE FROM THE DEPUY SYNTHES TRUMATCH SHOULDER SOLUTIONS TEAM TO EVALUATE OPTIONS AND RECOMMENDATIONS FOR THE PATIENT'S GLENOID FOLLOWING TREATMENT OF THE INFECTION. DOI: (B)(6) 2017 DOR: (B)(6) 2021 LEFT SHOULDER.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11659416 ↗InjuryApr 5, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE PATIENT SHOULDER WAS REVISED TO ADDRESS DISLOCATION FROM REVERSE TSA, AFTER WORKING WITH HIS HORSES. WAS REVISED FROM A +0 TO +2 GLENOSPHERE AND A +6 TO +9 POLY. NO DELAY IN SURGERY. DOI: UNKNOWN, DOR: (B)(6) 2021, RIGHT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 11613125 ↗InjuryMar 16, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED TO ADDRESS INFECTION. DOI: (B)(6) 2017; DOR: (B)(6) 2021; LEFT SHOULDER.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11497615 ↗InjuryMar 15, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ORIGINAL PRIMARY OPERATION WAS IN (B)(6) 2013. PATIENT BECAME UNWELL AND THERE WAS A SUSPICION OF INFECTION. THE SHOULDER HAD PREVIOUS WASHOUTS WITH NO REAL BENEFIT. ALL IMPLANTS WERE REMOVED AND A TEMPORARY CEMENTED SPACER WAS IMPLANTED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11486218 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K062250, K120174, K192855. One submission often covers many models, so check the product and lot details in each record.
- Z-0834-2018 · Aug 31, 2017 · TerminatedDELTA XTEND" Reverse Shoulder System Modular Centered and Eccentric Epiphysis Implants intended for use in total shoulder or hemi-shoulder replacement procedures in patients with n
- Z-0833-2018 · Aug 31, 2017 · TerminatedDELTA XTEND" Reverse Shoulder System Modular Centered and Eccentric Epiphysis Implants are intended for use in total shoulder or hemi-shoulder replacement procedures in patients wi
- Z-0532-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard humeral PE Cup Dia 38/+3 STD, sterile, Depuy France; REF 130738203. The device is used as a shoulder implant.
- Z-0537-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+6 STD, sterile, Depuy France; REF 130742206. The device is used as a shoulder implant.
- Z-0536-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+3 STD, sterile, Depuy France; REF 130742203. The device is used as a shoulder implant.
- Z-0534-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 38/+9 STD, sterile, Depuy France; REF 130738209. The device is used as a shoulder implant.
- Z-0535-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 38/+6R RET, sterile, Depuy France; REF 130738106. The device is used as a shoulder implant.
- Z-0538-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+9 STD, sterile, Depuy France; REF 130742209. The device is used as a shoulder implant.
- Z-0539-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42+6R RET, sterile, Depuy France; REF 130742106. The device is used as a shoulder implant.
- Z-0533-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humerus PE Cup, Dia 38/+6 STD, sterile, Depuy France; REF 130738206. The device is used as a shoulder implant.
Other PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED devices
- DJO SURGICAL 509-02-032Encore Medical, L.P. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED364 reports
- DELTA XTEND 130790036Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED171 reports
- DELTA XTEND 130770018Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED159 reports
- TRIGEN 71118200Smith & Nephew, Inc. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED123 reports
- DJO SURGICAL 509-02-432Encore Medical, L.P. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED119 reports
- DELTA XTEND 130738206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED116 reports
- COMPREHENSIVE SHOULDER SYSTEM 113633Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED2 recalls4 reports
- DELTA XTEND 130742206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED99 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.