Skip to content

SOLUTION SYSTEM 2108-04-000Recalls & reported problems

Depuy Orthopaedics, Inc. · SOLUTION SYSTEM CEMENT PLUG DRILL 4.5mm DIA · Class 1

Quick summary

This DePuy cement plug drill is a surgical drill bit. The 24 reports mostly describe the drill or its tip breaking during surgery, sometimes leaving a fragment in the patient. There were 2 injuries and no deaths. No recalls name this device, and reports rose from 1 in the previous 12 months to 3 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 10603295086932), received between Oct 12, 2017 and Jun 12, 2026 — classified as 2 injuries, 22 malfunctions.

The most frequently reported specific device problem is “Break” (23 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break23
  • Entrapment of Device3
  • Illegible Information1
  • Material Deformation1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • No Consequences Or Impact To Patient6
  • Foreign Body In Patient3
  • Insufficient Information3
  • Not Applicable3
  • No Information1

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 12, 2026Break

    IT WAS REPORTED THAT THE ITEM BROKE WHILE INSIDE A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25485417 ↗
  • MalfunctionMay 21, 2026Break; Entrapment of Device

    TWO REAMER TIPS FROM THE FEMORAL REV ALLOGRAFT BROKE OFF IN THE PATIENT¿S FEMUR. ONE WAS RETRIEVED. HEALTHCARE PROFESSIONAL ORDER XRAY TO CONFIRM PLACEMENT OF THE SECOND ONE AND WAS UNABLE TO RETRIEVE THE REAMER TIP FROM THE PATIENT¿S FEMUR. THE PROCEDURE WAS CONTINUED AND THE FEMORAL STEM IMPLANTED TO COMPLETE THE PROCEDURE AS INTENDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 25241378 ↗
  • MalfunctionFeb 25, 2026Break

    IT WAS REPORTED THAT DRILL BIT BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24445834 ↗
  • MalfunctionJul 1, 2025Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT THE SOL SYS CEM PLUG DRILLS 6.0MM WAS BROKEN AND RETRIEVED. SOL SYS CEM PLUG DRILLS 4.5MM WAS BROKEN AND NOT RETRIEVED.

    Patient problems: Foreign Body In Patient

    MAUDE report 22374013 ↗
  • MalfunctionAug 21, 2024Break; Break

    IT WAS REPORTED THAT THE DRILL BIT BROKE IN FEMUR TRYING TO REMOVE CEMENT.

    Patient problems: Insufficient Information

    MAUDE report 20017895 ↗
  • MalfunctionApr 9, 2024Break; Break

    IT WAS REPORTED THAT THE TIP OF SOLUTION DRILL BIT BROKE OFF DURING SURGERY. ALL PIECES WERE RECOVERED. SURGICAL DELAY WAS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 19069796 ↗
  • MalfunctionMar 27, 2024Break; Break

    IT WAS REPORTED THAT SOLUTION REAMER WAS BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18988642 ↗
  • MalfunctionJan 18, 2024Break; Break

    IT WAS REPORTED THAT THE TIP OF THE DRILL BIT BROKE INTRA SURGERY. THEY RETRIEVED THE BROKEN PART. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? --> NO, WAS PROCEDURE SUCCESSFULLY COMPLETED? --> YES, WERE FRAGMENTS GENERATED? --> NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18534257 ↗
  • MalfunctionSep 21, 2023Break; Break

    IT WAS REPORTED THAT THE DRILL BIT THE TIP BROKE INTERMEDULLARY. THE SURGEON BROUGHT IN X RAY & WAS ABLE TO EXTRACT THE PIECE FROM THE FEMUR. THERE WAS A 5 MINUTE DELAY & NO HARM CAME TO PATIENT. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? : UNKNOWN, IF YES, NUMBER OF MINUTES?: 10, ACTION TAKEN WHEN EVENT OCCURRED?: X RAY CALLED AND BROUGHT INTO ROOM, WAS PROCEDURE SUCCESSFULLY COMPLETED?: YES, WERE FRAGMENTS GENERATED?: YES, IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION?: YES, PATIENT STATUS/ OUTCOME / CONSEQUENCES?: NO, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED:? UNKNOWN, IS THE PATIENT PART OF A CLINICAL STUDY?: UNKNOWN, IP: (B)(4). DEVICE PROPERTY OF?: NONE, DEVICE IN POSSESSION OF?: NONE BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST. : TRUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17790726 ↗
  • MalfunctionJul 28, 2023Break; Break

    IT WAS REPORTED THAT THE 4.5MM DRILL BROKE. ALL PIECES WERE RETRIEVED. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17419090 ↗
  • MalfunctionDec 30, 2021Break; Break

    ADDITIONAL INFORMATION RECEIVED STATES THAT IT WAS CLOSE TO THE TIP THAT BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13117283 ↗
  • MalfunctionSep 15, 2021Break; Break

    ADDITIONAL INFORMATION RECEIVED STATES THAT THE ITEM BROKE INTO 2 PIECES. THE LARGE PIECE HAS BEEN RETURNED. THE REMAINDER WAS DISCARDED AFTER THE CASE BY THE HOSPITAL STAFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12474458 ↗
  • MalfunctionAug 5, 2021Break; Break

    IT WAS REPORTED THAT THE TIP OF THE CEMENT DRILL BROKE OFF. SURGICAL DELAY OF 3 MINUTES.

    Patient problems: Insufficient Information

    MAUDE report 12279904 ↗
  • MalfunctionApr 9, 2021Material Deformation; Material Deformation

    SOLUTION DRILL WAS BENT INTRA-OP WHILST SURGEON WAS ATTEMPTING TO USE IN FEMORAL CANAL. THERE WAS NO INCIDENT WITH PATIENT. INSTRUMENT WAS DEEMED IRREPARABLE AND A NEW ONE WAS ORDERED. THERE IS NO OTHER INFORMATION TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11641278 ↗
  • MalfunctionSep 8, 2020Break; Break

    IT WAS REPORTED THAT THE REAMER TIPS WAS BROKEN OFF. ALL PARTS RECOVERED. NO IMPACT ON SURGERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10504942 ↗
  • MalfunctionAug 26, 2020Break; Illegible Information; Break; Illegible Information

    ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE LOT NUMBERS WERE ILLEGIBLE ON THE DRILL. THE BROKEN PIECE WAS REMOVED FROM THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10453981 ↗
  • MalfunctionMar 4, 2020Break; Break

    IT WAS REPORTED THAT THINSHAFT 4.5 DRILL BROKE DURING SURGERY PLEASE REPLACE TO (B)(6). NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9788006 ↗
  • MalfunctionNov 22, 2019Break; Break

    IT WAS REPORTED THAT THE DRILL BROKE DURING CANAL REAMING. THERE WAS 10 MINUTES SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9362871 ↗
  • InjuryJul 24, 2019Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT DURING HIP REVISION, A DRILL BIT BROKE OFF INSIDE THE FEMORAL CANAL OF THE PATIENT AND THE SURGEON WAS UNABLE TO RETRIEVE. THERE WAS A 1 HOUR SURGICAL DELAY.

    Patient problems: Not Applicable

    MAUDE report 8823538 ↗
  • MalfunctionFeb 13, 2019Break; Break

    DURING A REVISION OF A FAILED FEMORAL STEM THE SURGEON NEEDED TO IMPLANT A REVISION FEMORAL STEM, THERE A DENSE SCLEROTIC BONE. THE SURGEON TEMPTED TO REAM THROUGH THE HARD BONE IN ORDER TO PASS A REVISION STEM. DURING REAMING THE DISTAL TIP BECAME BOUND IN THE CANAL AND SHEARED OFF.

    Patient problems: Foreign Body In Patient; No Consequences Or Impact To Patient

    MAUDE report 8336944 ↗
  • MalfunctionSep 18, 2018Break; Break

    IT WAS REPORTED THAT THE TIP OF THE REAMER WAS BROKE OFF DURING SURGERY. ALL THE PIECES WERE RECOVERED FROM THE PATIENT. THERE WAS 5 MINUTES SURGICAL DELAY REPORTED.

    Patient problems: Not Applicable

    MAUDE report 7886930 ↗
  • MalfunctionAug 15, 2018Break; Material Fragmentation; Break; Material Fragmentation

    ITEM BROKE DURING SURGERY - MORELAND INSTRUMENT SET.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7786259 ↗
  • InjuryDec 22, 2017Break; Break

    HIP REVISION, SURGEON: (B)(6), (B)(6) 2017. DURING CEMENT REMOVAL OF HIP REVISION. ATTEMPTING TO REMOVE CEMENT PLUG, DRILLED INTO CEMENT AND BROKE ON THE VERY FIRST USE. ACTION TAKEN TO MANAGE PROBLEM: HAD TO DO ETO (AND CABLES) TO REMOVE BROKEN DRILL BIT. 45 MINUTE DELAY TO PROCEDURE.

    Patient problems: Not Applicable

    MAUDE report 7141792 ↗
  • MalfunctionOct 12, 2017Break; Break

    IT WAS REPORTED THAT A REAMER TIP BROKE DURING THE SURGERY.

    Patient problems: No Information

    MAUDE report 6944555 ↗

Other BIT, DRILL devices

All BIT, DRILL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.