PINNACLE 2274-25-800Recalls & reported problems
Depuy (Ireland) · PINNACLE ONE PC REVISION DOME DRILL 25mm · Class 1
All models of PINNACLE (198)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy Pinnacle revision dome drill is a surgical drill used in hip replacement procedures. Its 150 reports mostly describe the drill or its bit breaking during surgery, and a few mention a ratchet that did not work. The reports include no deaths and no injuries, and most say there was no effect on the patient. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 151
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 151 adverse event reports that list this exact device (DI 10603295109136), received between Sep 14, 2017 and Sep 30, 2026 — classified as 151 malfunctions.
The most frequently reported specific device problem is “Break” (130 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break130
- Material Twisted/Bent10
- Entrapment of Device8
- Material Deformation8
- Failure to Cut5
- Naturally Worn4
- Appropriate Device Problem Term/Code Not Available2
- Bent2
Patient problems
- No Clinical Signs, Symptoms or Conditions98
- No Consequences Or Impact To Patient36
- Insufficient Information8
- Not Applicable7
- No Information2
- No Code Available1
Recent adverse event reports
The 25 most recent of 151 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Break
It was reported that the drill bit broke while being used to drill holes for screws in the acetabulum during the surgery. The broken piece was retrieved easily. A new drill bit was opened to complete the procedure. There was no significant delay. There was no reported patient or user harm. All available information has been disclosed.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948224 ↗MalfunctionAug 6, 2026Break; Material Deformation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50164335 ↗MalfunctionMay 21, 2026Mechanical Jam
IT WAS REPORTED THAT PRIOR TO A TOTAL HIP REPLACEMENT SURGERY, THE RATCHET WAS NOT WORKING. IT WAS NOTICED BEFORE USE. THERE WAS NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25241078 ↗MalfunctionMay 20, 2026Material Deformation
WAS THE PRODUCT BEING USED IN A CLINICAL TRIAL? NO. DID THE EVENT HAPPEN DURING A PROCEDURE? YES. WERE YOU IN THE PROCEDURE AT THE TIME OF THE EVENT? NO. EVENT OUTCOME/HOW WAS IT MANAGED? A DIFFERENT DRILL ON THE SET WAS USED WAS THERE ANY CONSEQUENCE TO THE PATIENT DUE TO THE EVENT? NO. WAS THE SURGERY PROLONGED DUE TO THE EVENT? NO. HAS THE REPORTER FACILITY INDICATED THERE MAY BE LEGAL ACTION? NO. PLEASE GIVE A DETAILED DESCRIPTION OF THE EVENT: THE DRILL BENT DURING NORMAL USE SO A DIFFERENT LENGTH DRILL FROM THE SET WAS USED TO FINISH THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25234381 ↗MalfunctionApr 13, 2026Break
HIP DRILL INSTRUMENTS TO BE DAMAGED AND REQUIRE REPLACEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24865489 ↗MalfunctionMar 12, 2026Break
IT WAS REPORTED THAT THE DRILL BROKE WHILE IN USE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24584055 ↗MalfunctionFeb 12, 2026Break
IT WAS REPORTED THAT THE DRILL BIT BROKE DURING USE - BOTH PIECES REMOVED WITH NO EFFECT ON PATIENT. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24322247 ↗MalfunctionJan 9, 2026Break
IT WAS REPORTED THAT PINNACLE REVISION DOME DRILL BROKE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24024442 ↗MalfunctionDec 22, 2025Break; Break
IT WAS REPORTED THAT DURING SURGERY, THE BURIC BROKE INTO TWO PIECES. IT WAS REPORTED THAT THE SURGERY TIME WAS NOT EXTENDED AS A DIRECT RESULT OF INCIDENT. THE PROCEDURE WAS COMPLETED WITH SAME LIKE PRODUCT. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23865066 ↗MalfunctionDec 18, 2025Break; Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE THE PIN REV DOME DRILL BROKE IN HALF. THERE WERE NO PATIENT CONSEQUENCES AND NO SURGICAL DELAYS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23840978 ↗MalfunctionNov 13, 2025Break; Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE DRILL BIT BROKE INTO TWO PIECES. IT WAS REPORTED THAT ALL BROKEN PIECES WERE RECOVERED. THERE WAS NO REPORT OF PATIENT HARM OR DELAY IN SURGERY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23542386 ↗MalfunctionNov 4, 2025Break; Break
IT WAS REPORTED THAT THE DRILL BROKE AND SNAPPED IN HALF WHILE BEING USED. IT WAS REPORTED THAT THE FRAGMENTS WERE REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. IT WAS REPORTED THAT THERE WAS NO DELAY IN THE SURGERY AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: Insufficient Information
MAUDE report 23466718 ↗MalfunctionOct 27, 2025Break; Break
IT WAS REPORTED THAT THE DRILLS FROM THE PINNACLE SET SNAPPED, AND THE T HANDLE CRACKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23392436 ↗MalfunctionSep 17, 2025Break; Break
IT WAS REPORTED THAT HOSPITAL WAS MISSING TWO 25MM DRILL PIECES FROM THE PINNACLE SCREW TRAY AND FLAGGED WITH THEIR SCRUB TEAM AND THEN THE INSTRUMENT COORDINATOR. THE TEAM NOTED THAT THESE HAVE BEEN MISSING FOR SOME TIME. THE SURGEON HAD BEEN USING MODULAR DRILL PIECES OR 25MM DRILL PIECES IN SURGERY IN PLACE OF THE 35MM WITH NO ISSUES. STAFF THOUGHT THESE HAD BEEN BROKEN FOR APPROXIMATELY SIX MONTHS. BOTH ITEMS ARE CONSIGNMENT SCREW PIECES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23075498 ↗MalfunctionSep 17, 2025Break; Break
IT WAS REPORTED THAT HOSPITAL WAS MISSING TWO 25 MM DRILL PIECES FROM THE PINNACLE SCREW TRAY AND FLAGGED WITH THEIR SCRUB TEAM AND THEN THE INSTRUMENT COORDINATOR. THE TEAM NOTED THAT THESE HAVE BEEN MISSING FOR SOME TIME. THE SURGEON HAD BEEN USING MODULAR DRILL PIECES OR 25 MM DRILL PIECES IN SURGERY IN PLACE OF THE 35 MM WITH NO ISSUES. STAFF THOUGHT THESE HAD BEEN BROKEN FOR APPROXIMATELY SIX MONTHS. BOTH ITEMS ARE CONSIGNMENT SCREW PIECES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23075496 ↗MalfunctionJul 17, 2025Break; Break
IT WAS REPORTED THAT THERE WAS BREAKAGE AND REPAIR OF CONSIGNED ITEMS IN THE ACCOUNT FROM GENERAL USAGE. NO OTHER INFORMATION WAS PROVIDED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22532131 ↗MalfunctionJul 17, 2025Break; Break
IT WAS REPORTED THE DEVICES WERE BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22532097 ↗MalfunctionJul 17, 2025Break; Break
IT WAS REPORTED THE DEVICES WERE BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22531978 ↗MalfunctionJul 17, 2025Break; Break
IT WAS REPORTED THE DEVICES WERE BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22531880 ↗MalfunctionJul 17, 2025Break; Break
IT WAS REPORTED THE DEVICES WERE BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22531847 ↗MalfunctionJul 11, 2025Break; Break
IT WAS REPORTED THAT THE DRILL BIT BROKE IN HALF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22482034 ↗MalfunctionJul 10, 2025Break; Break
IT WAS REPORTED THAT THE DRILL BROKE INTO TWO PIECES, ALL WAS REMOVED FROM PATIENT. NO PATIENT HARM AND NO SURGICAL DELAY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22467797 ↗MalfunctionMay 14, 2025Break; Break
IT WAS REPORTED THAT THE DRILL BIT BROKE IN HALF INTO ONE PIECE DURING NORMAL USE. THERE WAS NO ADVERSE EVENTS TO THE PATIENT, NO DELAYS IN THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22021299 ↗MalfunctionMar 28, 2025Break; Break
IT WAS REPORTED THAT THE TWO 25MM DRILL BITS BROKE AS THE SURGEON WAS DRILLING HOLES FOR THE CUP. ALL BROKEN PIECES WERE RETRIEVED BUT WERE LOST DURING DECONTAMINATION BY SPD. NO DELAY IN CASE.
Patient problems: Insufficient Information
MAUDE report 21724058 ↗MalfunctionMar 18, 2025Break; Break
IT WAS REPORTED A 25 MM PINN REVISION BROKE DURING USE INTRAOPERATIVELY. NO PATIENT IMPACT. NO SURGICAL DELAY. NO FRAGMENTS. AS PER HOSPITAL POLICY THE DRILL BIT WAS DISCARDED .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21634886 ↗
Other BIT, DRILL devices
- QUICKSET 2274-25-500Depuy Orthopaedics, Inc. · BIT, DRILL1,985 reports
- QUICKSET 2274-35-500Depuy Orthopaedics, Inc. · BIT, DRILL770 reports
- QUICKSET 2274-30-500Depuy Orthopaedics, Inc. · BIT, DRILL761 reports
- NA 03.037.022Synthes GmbH · BIT, DRILL339 reports
- NA 323.062Synthes GmbH · BIT, DRILL309 reports
- QUICKSET 2274-56-000Depuy (Ireland) · BIT, DRILL275 reports
- QUICKSET 2274-45-500Depuy Orthopaedics, Inc. · BIT, DRILL211 reports
- LCS 2366-83-000Depuy (Ireland) · BIT, DRILL207 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page