PINNACLE 2274-35-800Recalls & reported problems

Depuy (Ireland) · PINNACLE ONE PC REVISION DOME DRILL 35mm · Class 1

All models of PINNACLE (198)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Pinnacle One PC revision dome drill is a surgical drill bit used in joint replacement surgery. Its 104 reports, 1 of them an injury, mostly describe the drill breaking or snapping in two, with pieces retrieved in many reports and the surgery completed with another drill. There were no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
103
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 103 adverse event reports that list this exact device (DI 10603295109181), received between Aug 25, 2017 and Jun 9, 2026 — classified as 1 injury, 102 malfunctions.

The most frequently reported specific device problem is “Break” (64 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 3 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 3 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break64
  • Failure to Cut12
  • Material Deformation12
  • Material Twisted/Bent8
  • Bent4
  • Entrapment of Device4
  • Appropriate Device Problem Term/Code Not Available2
  • Crack2

Patient problems

  • No Clinical Signs, Symptoms or Conditions71
  • No Consequences Or Impact To Patient18
  • Not Applicable5
  • Foreign Body In Patient3
  • Insufficient Information3
  • No Information3
  • No Code Available1

Recent adverse event reports

The 25 most recent of 103 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 9, 2026Break

    IT WAS REPORTED THAT 35MM FLEXIBLE DRILL BIT BROKE IN THE MIDDLE OF THE FLUTES WHILE DRILLING FOR A CANCELLOUS BONE SCREW. BOTH PIECES WERE RETRIEVED. SURGICAL TIME WAS NOT EXTENDED AND NO ADVERSE CONSEQUENCES FOR THE PATIENT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25443522 ↗
  • MalfunctionApr 20, 2026Break

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE DRILL BIT BROKE/SNAPPED INTO 2 PIECES DURING NORMAL USE. IT WAS REPORTED THAT THERE WAS NO SURGICAL DELAY, AND NO ADVERSE PATIENT CONSEQUENCES. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24925151 ↗
  • MalfunctionApr 6, 2026Break

    WHILE PERFORMING A HIP REPLACEMENT, TWO DRILL BITS WERE SNAPPED INTO TWO PIECES. ALL THE FRAGMENTS WERE RETRIEVED; HOWEVER, BOTH DRILL BITS WERE DISPOSED OF. NO PATIENT HARM, NO X-RAYS, NO DELAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24793396 ↗
  • MalfunctionMar 20, 2026Break

    IT WAS REPORTED THAT THE INSTRUMENT SNAPPED IN HALF. THERE WERE NO FRAGMENTS LEFT IN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24647669 ↗
  • MalfunctionJan 20, 2026Break

    IT WAS REPORTED THAT THE DRILL END HAS BROKEN IN TWO. BOTH BITS WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24117186 ↗
  • MalfunctionJan 16, 2026Break

    IT WAS REPORTED THAT THE DRILL BROKE DURING A PROCEDURE. THE BROKEN PIECE WAS RETRIEVED WITH NO ISSUE AND A DIFFERENT DRILL FROM THE SET WAS USED TO COMPLETE THE SURGERY. THERE WAS NO SURGICAL DELAY AND THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24103333 ↗
  • MalfunctionJan 16, 2026Break

    DURING A PROCEDURE, THE DRILL BROKE DURING USE. THE BROKEN PIECE WAS RETRIEVED ON BOTH OCCASIONS WITH NO ISSUE AND A DIFFERENT DRILL FROM THE SET WAS USED TO COMPLETE THE SURGERY. THERE WAS NO PATIENT CONSEQUENCE. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24094586 ↗
  • MalfunctionDec 8, 2025Break; Break

    IT WAS REPORTED THAT DRILL BIT BROKE DURING THE CASE AND WAS THROWN OUT BY SCRUB TECH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23749265 ↗
  • MalfunctionAug 19, 2025Break; Break

    IT WAS REPORTED THAT DURING A PROCEDURE, THE DRILL BROKE INTO TWO PIECES. PROCEDURE WAS COMPLETED WITH SAME LIKE PRODUCT. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22842346 ↗
  • MalfunctionJul 31, 2025Break; Break

    IT WAS REPORTED THAT THE ITEM SNAPPED INTO TWO PIECES WHILE IT WAS IN USE. BOTH PIECES WERE RETRIEVED. THERE WAS NO PATIENT HARM, AND NO DELAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22675220 ↗
  • MalfunctionJul 7, 2025Break; Break

    IT WAS REPORTED THAT MULTIPLE DRILL BITS WERE BROKEN DURING SURGERY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22410795 ↗
  • MalfunctionJun 26, 2025Break; Break

    IT WAS REPORTED THAT DURING SURGERY, THE SURGEON WAS DRILLING A SCREW HOLE IN A PINNACLE CUP AND THE DRILL BROKE. THE BROKEN PIECE WAS RETRIEVED AND A DIFFERENT DRILL WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT CONSEQUENCE, AND THE SURGERY WAS NOT PROLONGED DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22335633 ↗
  • MalfunctionJun 19, 2025Break; Break

    IT WAS REPORTED THAT A 35MM DRILL BIT BROKE WHILE ATTEMPTING TO DRILL ACETABULUM. NOTHING WAS LEFT IN PATIENT. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22293753 ↗
  • MalfunctionJun 5, 2025Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL HIP REPLACEMENT ON (B)(6) 2025. THE DRILL BIT BROKE OFF IN THE PATIENT WHILE DRILLING A HOLE FOR SCREWS. A FRAGMENT WAS LEFT IN THE LEFT SIDE OF THE PELVIS, SUPERIOR TO THE CUP. SURGEON DEEMED IT UNRETRIEVABLE WITHOUT RISK FAR GREATER THAN LEAVING IT IN FOR NOW. PROCEDURE WAS COMPLETED WITH A THIRTY MINUTE DELAY.

    Patient problems: Foreign Body In Patient

    MAUDE report 22149524 ↗
  • MalfunctionApr 3, 2025Break; Break

    IT WAS REPORTED THAT DURING A PROCEDURE, A DRILL PIECE SNAPPED IN TWO. BOTH PARTS WERE RETRIEVED. THERE WAS NO PATIENT CONSEQUENCE. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21766781 ↗
  • MalfunctionMar 31, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT DURING PROCEDURE THE DRILL WAS FOUND TO BE BLUNT DURING USE SO A DIFFERENT DRILL FROM THE SAME SET WAS USED. THERE WAS NO PATIENT CONSEQUENCE AND NO SURGICAL DELAY. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED DEVICE IT WAS REVEALED THAT THE TIP BROKEN FROM PIN REV DOME DRILL 35MM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21732999 ↗
  • MalfunctionFeb 19, 2025Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE DRILL BIT BROKE WHEN DRILLING THE ACETABULUM. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21411944 ↗
  • MalfunctionJan 15, 2025Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT DURING A TOTAL HIP REPLACEMENT SURGERY, THE DRILL BIT BROKE WHILE IN USE. UNABLE TO LOCATE BROKEN PIECE INSIDE THE PATIENT DESPITE ATTEMPT TO REMOVE CUP TO TRY AND LOCATE AND RETRIEVE BROKEN INSTRUMENT. SURGEON WAS UNABLE TO LOCATE AND SO DECIDED TO CONTINUE IMPLANTING CUP, LEAVING DRILL BIT FRAGMENT INSIDE THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 21158956 ↗
  • MalfunctionDec 13, 2024Break; Break

    IT WAS REPORTED THAT THE REVISION DOME DRILL BIT WAS BROKEN. THERE WERE NO PATIENT CONSEQUENCES. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20925868 ↗
  • MalfunctionOct 2, 2024Break; Break

    IT WAS REPORTED THAT THE BORIC WAS BROKEN INTO TWO PIECES INTRAOPERATIVELY. SIMILAR PRODUCT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20353084 ↗
  • MalfunctionAug 26, 2024Break; Break

    IT WAS REPORTED THAT THE DRILL BITS BROKE DURING SURGERY, HOWEVER THERE WAS NO HARM TO PATIENT AS A RESULT OF THIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20065941 ↗
  • MalfunctionJun 4, 2024Break; Break

    IT WAS REPORTED THAT THE DRILL BROKE OF INTO TWO PIECES. PATIENT WAS NOT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19457729 ↗
  • MalfunctionJun 3, 2024Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    THE DRILL BIT BROKE OFF AND GOT STUCK IN THE FLEXI DRILL SHAFT MAKING IT UNUSABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19447573 ↗
  • MalfunctionMay 8, 2024Break; Break

    IT WAS REPORTED THAT THE FLEXI DRILL SNAPPED IN 2 PIECES. BOTH PIECES WERE RETRIEVED AND AN ALTERNATIVE WAS USED. SURGERY WAS NOT DELAYED DUE TO THE REPORTED EVENT.

    Patient problems: Insufficient Information

    MAUDE report 19267990 ↗
  • MalfunctionApr 10, 2024Break; Failure to Cut; Break; Failure to Cut

    ADDITIONAL INFORMATION WAS RECEIVED: A. WHAT IS THE EVENT DATE? - AS I SAID IN MY ORIGINAL COMPLAINT: ¿THERE HAVE RECENTLY BEEN A NUMBER OF THEIR CONSIGNED INSTRUMENT TRAYS TAKEN OUT OF CIRCULATION DUE TO BROKEN OR DAMAGED INSTRUMENTS. THE BROKEN OR DAMAGED INSTRUMENTS HAVE BEEN DISPOSED OF EITHER IN THEATRES OR IN CSSD AND THE TRAYS SET ASIDE, SO I DON¿T HAVE ANY DETAILS ON WHAT HAPPENED WITH ANY OF THEM, OTHER THAN THEY WERE DAMAGED OR BROKEN IN THE COURSE OF NORMAL USE.¿ THESE TRAYS WERE TAKEN OUT OF CIRCULATION BY THE HOSPITAL, NOT BY ME, SO I DON¿T HAVE ANY INFORMATION AS TO WHEN THEY DID SO. THERE ARE MULTIPLE TRAYS AND THEREFORE PROBABLY MULTIPLE EVENT DATES, BUT NONE OF THE DATES HAVE BEEN RECORDED BY THE HOSPITAL. B. PLEASE CLARIFY WHAT YOU MEAN BY DAMAGED? IS IT BROKEN INTO TWO PIECES, DULL, BENT, CRACKED, STRIPPED, WORN, SCRATCHED, CROSS THREADED OR ANY DEVICE INTERACTION ISSUES? -AGAIN, I DON¿T HAVE ANY INFORMATION ON THIS BUT I CAN SPECULATE THAT THE DRILLS PROBABLY GOT BLUNT OR BROKE, THE TIBIAL IMPACTORS PROBABLY BROKE INTO 2 PIECES (WE¿VE HAD MANY OF THESE BREAK IN HALF WHEN THEY GET OLD), THE POLY INSERT IMPACTOR PROBABLY CRACKED AS IT IS PLASTIC, AND THE 2 DRIVERS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19080214 ↗

FDA clearance and classification

The FDA regulates product code HTW (bit, drill) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the PINNACLE 2274-35-800

Has the PINNACLE 2274-35-800 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10603295109181), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the PINNACLE 2274-35-800?

The FDA's MAUDE database lists 103 adverse event reports for this device, received between Aug 25, 2017 and Jun 9, 2026. In those reports, the most common specific device problems are “Break” (64), “Failure to Cut” (12) and “Material Deformation” (12); the most common patient problems are “Not Applicable” (5) and “Foreign Body In Patient” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PINNACLE 2274-35-800?

None of the reports are classified as deaths, 1 as an injury and 102 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PINNACLE 2274-35-800 FDA approved?

Devices of this type (product code HTW) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).

Who makes the PINNACLE 2274-35-800, and is it still sold?

The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand PINNACLE, model 2274-35-800). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.