PINNACLE 2274-45-800Recalls & reported problems

Depuy (Ireland) · PINNACLE ONE PC REVISION DOME DRILL 45mm · Class 1

All models of PINNACLE (198)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Pinnacle 45mm revision dome drill bit is used to prepare the hip socket during replacement surgery. The 63 reports mostly describe the bit breaking or snapping during surgery, sometimes in two pieces inside the patient, and bits that would not cut or had dulled, bent or worn with use. In the accounts that give detail, the pieces were retrieved and surgery continued without harm. There are no deaths and no injuries reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
63
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 63 adverse event reports that list this exact device (DI 10603295109228), received between Dec 3, 2017 and Jun 8, 2026 — classified as 63 malfunctions.

The most frequently reported specific device problem is “Break” (39 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break39
  • Failure to Cut21
  • Material Twisted/Bent3
  • Naturally Worn3
  • Device-Device Incompatibility1
  • Dull, Blunt1

Patient problems

  • No Clinical Signs, Symptoms or Conditions38
  • Insufficient Information12
  • No Consequences Or Impact To Patient10
  • No Code Available2
  • Not Applicable2

Recent adverse event reports

The 25 most recent of 63 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 8, 2026Break

    IT WAS REPORTED THAT A DRILL BIT SNAPPED. BOTH PARTS WERE RETRIEVED. AN ALTERNATIVE DRILL BIT WAS USED. NO HARM WAS CAUSED TO THE PATIENT AND THE PROCEDURE WAS CARRIED OUT WITH NO COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25428668 ↗
  • MalfunctionApr 16, 2026Break

    IT WAS REPORTED THAT 45MM DRILL BIT BROKE IN PATIENT. SURGEON WAS ABLE TO RETRIEVE BROKEN PARTS. ALL PARTS RETRIEVED. NO DELAY AS WE HAD AN EXTRA DRILL BIT. NO HARM DONE TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24899826 ↗
  • MalfunctionApr 13, 2026Break

    HIP DRILL INSTRUMENTS TO BE DAMAGED AND REQUIRE REPLACEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24865474 ↗
  • MalfunctionJul 17, 2025Break; Break

    IT WAS REPORTED THAT THERE WAS BREAKAGE AND REPAIR OF CONSIGNED ITEMS IN THE ACCOUNT FROM GENERAL USAGE. NO OTHER INFORMATION WAS PROVIDED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22532164 ↗
  • MalfunctionJan 22, 2025Break; Break

    IT WAS REPORTED THAT THE DRILL BIT BROKE INTO TWO (2) PIECES DURING THE OPERATION. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21208256 ↗
  • MalfunctionMar 25, 2024Break; Break

    IT WAS REPORTED THAT THERE WAS A BROKEN DRILL BIT (45MM). SURGICAL DELAY WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18970983 ↗
  • MalfunctionFeb 7, 2024Break; Break

    IT WAS REPORTED THAT THE TWO DRILL BITS ARE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18660679 ↗
  • MalfunctionJan 24, 2024Break; Break

    IT WAS REPORTED THAT THE DRILL SNAPPED DURING CASE. BOTH ENDS WERE RETRIEVED. AN ALTERNATIVE DRILL WAS USED. PATIENT SAFE. SURGERY WAS NOT DELAYED DUE TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18570889 ↗
  • MalfunctionDec 13, 2023Break; Break

    IT WAS REPORTED THAT THE TIP BROKE IN HALF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18321896 ↗
  • MalfunctionDec 8, 2023Break; Break

    DRILL PIECE WAS SNAPPED INTO TWO PIECES, BY SURGEON DURING THA BEING CONDUCTED. THE PIECES WERE IMMEDIATELY REMOVED AND IDENTIFIED AS WHOLE BY THE SURGEON, SCRUB NURSE AND ASSISTING NURSE. ANOTHER SCREW WAS THEN USED, WHICH WAS READILY AVAILABLE AND THERE WAS NO IMPACT TO THE PATIENT. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18293712 ↗
  • MalfunctionNov 13, 2023Break; Break

    ADDITIONAL INFORMATION WAS RECEIVED: A. DID IT BROKE INTO TWO OR MORE PIECES? 2 PIECES. B. WERE THERE ANY FRAGMENTS GENERATED? IF YES, WERE ALL THE PIECES RETRIEVED? NO FRAGMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18124875 ↗
  • MalfunctionNov 13, 2023Break; Break

    ADDITIONAL INFORMATION WAS RECEIVED: A. DID IT BROKE INTO TWO OR MORE PIECES? 2 PIECES. B. WERE THERE ANY FRAGMENTS GENERATED? IF YES, WERE ALL THE PIECES RETRIEVED? NO FRAGMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18124534 ↗
  • MalfunctionSep 27, 2023Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT BROKE.

    Patient problems: Insufficient Information

    MAUDE report 17830012 ↗
  • MalfunctionSep 27, 2023Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT BROKE DURING HIP SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17829754 ↗
  • MalfunctionAug 25, 2023Break; Break

    T WAS REPORTED THAT THE INSTRUMENT BROKE DURING SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 17624328 ↗
  • MalfunctionAug 22, 2023Break; Break

    IT WAS REPORTED THAT DEVICE WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17597417 ↗
  • MalfunctionJul 6, 2023Break; Break

    IT WAS REPORTED THAT THE DRILL BIT WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17270655 ↗
  • MalfunctionJul 5, 2023Break; Break

    IT WAS REPORTED THAT THE DRILL BIT HAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17257782 ↗
  • MalfunctionMay 23, 2023Break; Break

    IT WAS REPORTED THAT INSTRUMENT BROKE DURING HIP SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16987340 ↗
  • MalfunctionNov 1, 2022Break; Break

    IT WAS REPORTED THAT THE ITEM BROKE DURING HIP SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15707765 ↗
  • MalfunctionSep 29, 2022Break; Break

    IT WAS REPORTED THAT THE DRILL BIT BROKE IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15501903 ↗
  • MalfunctionSep 29, 2022Break; Break

    IT WAS REPORTED THAT DRILL BIT BROKE IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15501727 ↗
  • MalfunctionAug 10, 2022Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT PART OF DRILL BIT JAMMED INTO THE DRILL GUIDE DURING SURGERY. NO SURGICAL DELAY. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15197664 ↗
  • MalfunctionAug 5, 2022Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15178213 ↗
  • MalfunctionMay 9, 2022Break; Break

    INSTRUMENT BROKE DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14338164 ↗