Skip to content

PINNACLE 2274-81-000Recalls & reported problems

Depuy (Ireland) · PINNACLE ONE PC REVISION PERIPHERAL DRILL · Class 1

All models of PINNACLE (198)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Pinnacle peripheral drill bit is used to drill screw holes during hip replacement surgery. The 30 adverse event reports mostly describe the drill bit breaking during use, sometimes into pieces that had to be removed from the patient, along with bits going dull and failing to cut. No deaths or injuries were reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 10603295109433), received between Dec 7, 2017 and May 13, 2026 — classified as 29 malfunctions.

The most frequently reported specific device problem is “Break” (26 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break26
  • Entrapment of Device2
  • Device Contamination with Chemical or Other Material1
  • Dull, Blunt1
  • Failure to Cut1
  • Fracture1
  • Material Deformation1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14
  • No Consequences Or Impact To Patient11
  • Foreign Body In Patient2
  • Insufficient Information2
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2026Break

    IT WAS REPORTED THAT THE DRILL BIT BROKE WHILE DRILLING FOR THE SCREW. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25163960 ↗
  • MalfunctionFeb 4, 2026Break

    IT WAS REPORTED THAT DURING A TOTAL HIP REPLACEMENT PROCEDURE, THE DRILL BIT WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24257518 ↗
  • MalfunctionOct 31, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT'S TIP BROKE WHILE DRILLING. ALL PARTS WERE RETRIEVED FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23434946 ↗
  • MalfunctionOct 21, 2025Break; Break

    IT WAS REPORTED THAT THE DRILL BIT WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23351570 ↗
  • MalfunctionDec 4, 2024Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT A DRILL BIT BROKE POSTOPERATIVELY. X-RAYS REVEALED THAT THE DRILL BIT FRAGMENT HAD NOT BEEN RETRIEVED. UNKNOWN SURGICAL DELAY.

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 20844151 ↗
  • MalfunctionOct 28, 2024Break; Break

    IT WAS REPORTED THAT THE DRILL BIT BROKE DURING USE. ALL PIECES WERE RETRIEVED. THERE WAS NO SURGICAL DELAY. THERE WERE NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20544048 ↗
  • MalfunctionSep 17, 2024Break; Break

    ADDITIONAL INFORMATION WAS RECEIVED: IN REGARD TO REAMERS NO NORMAL WEAR AND TEAR AND IN REGARD TO DRILL BITS NO ALL PIECES RECOVERED QUICKLY. IN REGARD TO PERIPERAL REVISION DRILL BIT, IT WAS CONFIRMED THAT ADVERSE EFFECTS OCCURRED ON THE PATIENT. THE REVISION SURGEON WAS NOT ABLE TO PUT PERIPHERAL SCREWS IN AS HE WOULD HAVE LIKED. THE PERIPHERAL DRILL BIT BROKE IN 2X PIECES AND WERE ALL RECOVERED. THE DOME HOLE DRILL BIT WAS DULL. WE JUST WENT UP A SIZE IN DRILL BIT (NO NOTICEABLE DELAY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20235170 ↗
  • MalfunctionAug 5, 2024Break; Break

    IT WAS REPORTED THAT THE DRILL IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19905753 ↗
  • MalfunctionJul 25, 2024Break; Break

    IT WAS REPORTED THAT THE DRILL BIT BROKE AT TIP, ALL PARTS RETRIEVED.

    Patient problems: Insufficient Information

    MAUDE report 19825921 ↗
  • MalfunctionDec 6, 2023Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT THE DRILL BIT IS BENT AND BROKEN IN TWO PIECES. IT IS UNKNOWN IF THE SURGERY WAS DELAYED DUE TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18274098 ↗
  • MalfunctionNov 28, 2023Break; Break

    ADDITIONAL INFORMATION WAS RECEIVED AND STATED. A. PLEASE CONFIRM, ABOUT SURGICAL DELAY? THERE WAS NO DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18218397 ↗
  • MalfunctionJul 19, 2023Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT DURING A REVISION SURGERY IN WHICH THE COMPONENTS OF THE ACETABULUM AND THE HEAD OF THE SAID IMPLANT WERE REPLACED, A PIN REV PERIPH DRILL (REUSABLE ITEM) WAS USED FOR DRILLING, AMONG OTHER THINGS. DURING THE USE OF THE DRILL IT BROKE IN THE PATIENT'S BODY. THE SURGEON DECIDED NOT TO REMOVE THE BROKEN PART OF THE DRILL FROM THE PATIENT'S BODY AND COMPLETED THE OPERATION WITH ANOTHER DRILL OF A MATCHING DIAMETER FOR THE IMPLANT FIXATION. ACCORDING TO THE SURGEON, NO FURTHER MEDICAL INTERVENTION IS PLANNED AT THIS STAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17353236 ↗
  • MalfunctionOct 31, 2022Break; Break

    ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE TIP OF THE DRILL BIT BROKE INTO TWO PIECES. IT BROKE BECAUSE THE SURGEON OVER TORQUED THE DRILL BIT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15699237 ↗
  • MalfunctionAug 22, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE PINNACLE REVISION PERIPHERAL SCREW DRILL BIT IS DULL. DRILL BIT BROKE WHILE ATTEMPTING TO DRILL FOR A PERIPHERAL SCREW. A SECONDARY DRILL BIT WAS USED TO COMPLETE THE PROCEDURE. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15266005 ↗
  • MalfunctionApr 27, 2022Break; Break

    IT WAS REPORTED THAT THE TWO REVISION SCREW DRIVERS WERE BROKEN WHEN PREPPING FOR BUTTRESS. THERE WAS NO DELAY IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14219784 ↗
  • MalfunctionApr 27, 2022Break; Break

    IT WAS REPORTED THAT THE TWO REVISION SCREW DRIVERS WERE BROKEN WHEN PREPPING FOR BUTTRESS. THERE WAS NO DELAY IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14219781 ↗
  • MalfunctionOct 23, 2020Break; Break

    IT WAS REPORTED THAT THE PERIPHERAL DRILL BIT BROKE WHILE PLACING SCREWS. IT BROKE IN THE MIDDLE OF THE FLUTED TIP. ALL PIECES WERE REMOVED. ALSO, THE RIGID HEX DRIVER WAS STRIPPED DURING TRIALING. NO PIECES WERE BROKEN OFF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10730750 ↗
  • MalfunctionApr 27, 2020Break; Break

    THE INSTRUMENTS WERE REPORTED FOR UNKNOWN REASON. FOUND IN OFFICE SETS AND NO SURGERY WAS EFFECTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10002439 ↗
  • MalfunctionApr 3, 2020Break; Break

    IT WAS REPORTED THAT THE TOP OF DRILL BIT BROKE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9924191 ↗
  • MalfunctionJun 18, 2019Break; Break

    IT WAS REPORTED THAT THE PERIPHERAL DRILL BIT BROKE, ALL PIECES WERE REMOVED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8708554 ↗
  • MalfunctionMay 17, 2019Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT DURING A PINNACLE REVISION, A DRILL BIT WAS BROKEN WITH A REVISION CUP IN AND MULTIPLE SCREWS IN. THE SURGEON DETERMINED IT IS BEST TO LEAVE THE PART OF THE BIT INSTEAD OF HARMING THE PELVIS MORE BY TAKING THE CUP OUT AND DIGGING FOR IT. NO FURTHER PATIENT INFORMATION IS AVAILABLE AT THIS TIME. DRILL BIT WILL NOT BE RETURNED AS IT IS BEING SENT TO RISK MANAGEMENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 8619422 ↗
  • MalfunctionMar 29, 2019Material Twisted/Bent; Material Twisted/Bent

    DRILL BIT BENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8465388 ↗
  • MalfunctionMar 1, 2019Break; Device Contamination with Chemical or Other Material; Break; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT WHILE DRILLING FOR A PINNACLE REVISION PERIPHERAL SCREW PART OF THE DRILL BIT WAS BROKE OFF INSIDE THE PATIENT. THE SURGEON CHOSE TO LEAVE IT IN THE PATIENT. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8382657 ↗
  • MalfunctionJan 2, 2019Break; Break

    IT WAS REPORTED THAT THE INCREDIBLE BONE QUALITY BROKE MULTIPLE DRILL BITS WHILE TRYING TO DRILL THROUGH THIS PATIENTS BONE FOR SCREWS IN THE ACETABULAR SHELL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8210448 ↗
  • MalfunctionNov 15, 2018Break; Break

    THE DRILL BIT BROKE

    Patient problems: Not Applicable

    MAUDE report 8075631 ↗

Other BIT, DRILL devices

All BIT, DRILL devices →

Sources and updates