DELTA XTEND 2307-89-000Recalls & reported problems

Depuy (Ireland) · DELTA Xtend GLENOID CANNULATED STOP DRILL DIA 7.5MM · Class 1

All models of DELTA XTEND (194)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Delta Xtend cannulated stop drill is used in shoulder surgery to prepare a hole for a joint implant. The 68 reports mostly describe the drill being dull or failing to cut, sometimes leaving it stuck on the surgical wire or unable to detach from the power tool, with a few cases of bending or breaking; most procedures were completed without harm to the patient. There are no deaths and no injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
68
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 68 adverse event reports that list this exact device (DI 10603295116509), received between Dec 20, 2017 and May 7, 2026 — classified as 68 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (38 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Cut38
  • Dull, Blunt19
  • Device-Device Incompatibility6
  • Break2
  • Material Twisted/Bent2
  • Naturally Worn2
  • Appropriate Device Problem Term/Code Not Available1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions40
  • No Consequences Or Impact To Patient23
  • Insufficient Information5

Recent adverse event reports

The 25 most recent of 68 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 7, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE DELTA XTEND CENTRAL STOP DRILL IS BLUNT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25111848 ↗
  • MalfunctionMay 19, 2023Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE INSTRUMENT WAS NO LONGER SHARP.

    Patient problems: Insufficient Information

    MAUDE report 16962485 ↗
  • MalfunctionApr 10, 2023Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE CANNULATED STOP DRILL WAS BLUNT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16706679 ↗
  • MalfunctionFeb 8, 2023Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE DRILL IS DULL AND WILL NOT CUT. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16321794 ↗
  • MalfunctionFeb 1, 2023Device-Device Incompatibility; Device-Device Incompatibility

    THIS PC IS RELATED TO (B)(4), WHICH REPORTS ABOUT THE POWER TOOL (CHUCK) USED IN THE SURGERY. THIS PC REPORTS ABOUT THE JOINT/JRN PRODUCT (DRILL) USED IN THE SURGERY. IT WAS REPORTED THAT THIS WAS AN RSA TREATING SHOULDER PERFORMED ON (B)(6) 2023. THE GLENOID CANNULATED STOP DRILL GOT UNABLE TO BE DETACHED FROM THE CHUCK BECAUSE THE GLENOID CANNULATED STOP DRILL WAS FORCIBLY ATTACHED TO THE CHUCK THAT DOES NOT MATCH TO THE GLENOID CANNULATED STOP DRILL. THE PROCEDURE WAS DONE BY USING THE GLENOID CANNULATED STOP DRILL WITH NO PROBLEMS. THERE WAS NO SURGICAL DELAY AND NO HARM TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16277867 ↗
  • MalfunctionDec 9, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE DAMAGED PEG DRILL COIL NOT FIT OVER METAGLENE GUIDE PIN. THERE WAS A FIVE MINUTE SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15948958 ↗
  • MalfunctionDec 7, 2022Device-Device Incompatibility; Device-Device Incompatibility

    INTRAOPERATIVELY, THE CENTRAL DRILL WAS INTRODUCED WITH THE PEG DRILL (7.5MM) OVER THE ALREADY LYING K-WIRE AND THE CENTRAL DRILL CHANNEL WAS PREPARED. WHEN THE DRILL WAS REMOVED, THE CENTRAL K-WIRE JAMMED IN THE DRILL AND COULD NO LONGER BE REMOVED. THE DRILL AND K-WIRE WERE REMOVED FROM THE PATIENT WITHOUT LEAVING ANY REMNANTS. THE REST OF THE SURGICAL PROCEDURE PROCEEDED AS PLANNED. NO PROLONGATION OF THE SURGERY TIME, NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15933613 ↗
  • MalfunctionNov 21, 2022Failure to Cut; Material Integrity Problem; Naturally Worn; Failure to Cut; Material Integrity Problem; Naturally Worn

    IT WAS REPORTED THAT THE DRILL IS NOT SHARP, IS DAMAGED. SO THE DEVICE IS NOT PROPERLY IN GOOD CONDITIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15833835 ↗
  • MalfunctionSep 29, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE PEG DRILL IS GETTING STUCK IN DRILL ATTACHMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15508147 ↗
  • MalfunctionJul 5, 2022Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT THE INSTRUMENT WAS REPORTED TO BE BROKEN/WORN OUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14914641 ↗
  • MalfunctionJun 14, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMER IS DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14686262 ↗
  • MalfunctionMay 10, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE INSTRUMENT WAS FOUND TO BE USEABLE BUT SLIGHTLY BLUNT NO COMPLICATIONS. NO DELAY TO PROCESS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14346795 ↗
  • MalfunctionApr 20, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE DRILLS WERE DULL. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14160199 ↗
  • MalfunctionApr 20, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE DRILLS WERE DULL. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14160196 ↗
  • MalfunctionApr 20, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE DRILLS WERE DULL. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14160193 ↗
  • MalfunctionApr 20, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE DRILLS WERE DULL. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14160190 ↗
  • MalfunctionApr 19, 2022Failure to Cut; Failure to Cut

    GLENOID CANNULATED STOP DRILL AS IT WAS BLUNT. HUMERAL IMPLANT DRIVER. THE CONNECTION BETWEEN THE DRIVER AND THE TRIAL STEM/PROTHESIS WAS LOOSE DUE TO WEAR AND TEAR. WHEN THE SURGEON REATTACHED THE IMPLANT DRIVER TO THE TRIAL STEM (IN SITU IN THE PATIENTS HUMERUS) IT KEPT PULLING OFF THE TRIAL. SURGICAL DELAY OF 2 MINUTES OCCURRED. NO ADVERSE AFFECTS ON THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 14140934 ↗
  • MalfunctionMar 23, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GLENOID STEP DRILL AND GLENOID REAMER BLUNT OUT OF CONSIGNMENT SET- NOTED IN CASE TODAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13869293 ↗
  • MalfunctionMar 15, 2022Failure to Cut; Failure to Cut

    ADDITIONAL INFORMATION RECEIVED INDICATED THAT THERE WAS A 5-MINUTE SURGICAL DELAY. THE SURGEON USED THE 10MM DRILL BUT DID NOT GO TO THE DEPTH STOP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13771408 ↗
  • MalfunctionFeb 25, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE DELTA GLENOID CANNULATED STOP DRILL 2307-89-000 WAS DULL EVEN THOUGH IT¿S ONLY BEEN IN THE SET SINCE 2010. A REPLACEMENT IS NEEDED. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13600900 ↗
  • MalfunctionDec 22, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THIS WAS AN UNKNOWN PROCEDURE PERFORMED ON (B)(6) 2021. THE PATIENT HAD THE INFERIOR BONE QUALITY. THE DRILL WAS TOO DULL TO CUT THE LESION. THE PROCEDURE WAS COMPLETED WITH A REPLACING +10MM DRILL WITHOUT A FURTHER ISSUE. THERE WAS NO SURGICAL DELAY AND NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13061286 ↗
  • MalfunctionDec 14, 2021Failure to Cut; Device-Device Incompatibility; Failure to Cut; Device-Device Incompatibility

    THE METAGLENE GUIDE PIN GOT STUCK IN THE METAGLENE POSITIONING PLATE. IT SNAPPED OFF IN THE PLATE. THE STAFF WERE UNABLE TO REMOVE THE BROKEN SHARPE END FROM THE PLATE. ITS GONE OFF TO THEIR SSD DEPT AND IF THEY CANT REMOVE IT BE CAREFUL AS IT WILL COME BACK LIKE THAT TO US. THE 2ND METAGLENE GUIDE PIN GOT STUCK IN THE STOP DRILL AND ITS STILL IN THERE AS WELL SO BE CAREFUL WHEN IT COMES BACK. THE STOP DRILL WAS ALSO REPORTED AS BEING BLUNT. THE SURGEON WAS ABLE TO CONTINUE WITH THE CASE AND THANKFULLY WAS HAPPY WITH THE END RESULT OF THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13005545 ↗
  • MalfunctionDec 3, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT DURING A DELTA REVERSE TSA, THE SURGEON COMPLAINED THAT THE CENTRAL DRILL WAS DULL. A SECOND SET WAS OPENED AND HE FELT IT WAS ALSO DULL. DURING THE SAME CASE IT WAS NOTED THAT THE DRIVER SHAFT WAS BENT AND WOBBLED. SURGICAL DELAY OF 5 MINUTES.

    Patient problems: Insufficient Information

    MAUDE report 12923744 ↗
  • MalfunctionDec 3, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT DURING A DELTA REVERSE TSA, THE SURGEON COMPLAINED THAT THE CENTRAL DRILL WAS DULL. A SECOND SET WAS OPENED AND HE FELT IT WAS ALSO DULL. DURING THE SAME CASE IT WAS NOTED THAT THE DRIVER SHAFT WAS BENT AND WOBBLED. SURGICAL DELAY OF 5 MINUTES.

    Patient problems: Insufficient Information

    MAUDE report 12923705 ↗
  • MalfunctionDec 2, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE CENTRAL METAGLENE DRILL BIT WAS LOOKING BLUNT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12915567 ↗

Other BIT, DRILL devices

All BIT, DRILL devices →