Skip to content

NA 2421-02-000Recalls & reported problems

Depuy Orthopaedics, Inc. · FEMORAL/TIBIAL EXTRACTOR 11mm · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Orthopaedics femoral/tibial extractor is a surgical instrument used to remove knee implant components during surgery. The 34 reports mostly describe the tool breaking or cracking, sometimes snapping in half or chipping, and in a few cases stripped screws or a broken widget leaving pieces that were retrieved; no deaths or injuries were reported. There are no recalls naming this device, and the number of reports is steady (5 in the last 12 months versus 4 in the prior12).

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
33
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 10603295122722), received between Jun 22, 2017 and Apr 14, 2026 — classified as 33 malfunctions.

The most frequently reported specific device problem is “Break” (27 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 3 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break27
  • Crack4
  • Illegible Information2
  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1
  • Material Deformation1
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • No Consequences Or Impact To Patient10
  • No Information5
  • Insufficient Information2
  • No Code Available1

Recent adverse event reports

The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 14, 2026Break

    IT WAS REPORTED THAT PRODUCT WAS CHIPPED AND NEEDS TO BE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24873137 ↗
  • MalfunctionApr 7, 2026Break

    IT WAS REPORTED THAT ON UPON OFFICE IN HOUSE INSPECTION, IT WAS FOUND THAT THE INSTRUMENT WAS BROKEN. NO PREVIOUS CASES WERE AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24811554 ↗
  • MalfunctionJan 20, 2026Break

    IT WAS REPORTED THAT INSTRUMENT WAS BROKEN IN HALF, NEED REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24119606 ↗
  • MalfunctionNov 5, 2025Break; Break

    THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED THAT DURING A KNEE SURGERY, THE SCREWS ON THE FEMUR IMPACTOR WERE STRIPPED AND WOULD NOT THREAD BACK ON TO THE IMPACTOR, RENDERING IT UNUSABLE FOR SURGERY. ALSO, A WIDGET WAS USED TO HIT OFF THE FEMUR AND A PIECE OF THE WIDGET BROKE OFF. ALL PIECES WERE RETRIEVED AND THERE WAS NO DELAY IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476697 ↗
  • MalfunctionOct 28, 2025Crack; Crack

    IT WAS REPORTED THAT THE REPORTED DEVICES WERE CRACKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23401728 ↗
  • MalfunctionMay 15, 2025Break; Break

    IT WAS REPORTED THAT 2 INSTRUMENTS WAS BROKEN WHILE TRYING TO REMOVE THE FEMORAL COMPONENT. ALL PIECES WERE RETRIEVED.

    Patient problems: Insufficient Information

    MAUDE report 22028437 ↗
  • MalfunctionMay 9, 2025Break; Illegible Information; Break; Illegible Information

    IT WAS REPORTED THAT THE INSTRUMENT WAS BROKEN WHILE USING TO EXPLANT FEMUR. THERE WAS NO PATIENT HARM AND NO DELAY TO THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21996755 ↗
  • MalfunctionMay 2, 2025Break; Crack; Break; Crack

    IT WAS REPORTED THAT THE DEVICES WERE CHIPPED AND CRACKED. IT DIDN'T HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21954300 ↗
  • MalfunctionJun 28, 2024Break; Break

    THE INSTRUMENT WAS REPORTED FOR UNKNOWN REASON. DID NOT HAPPEN IN SURGERY. ADDITIONAL INFORMATION: THE TIP OF THE DEVICE CHIPPED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19632409 ↗
  • MalfunctionJun 30, 2023Break; Break

    IT WAS REPORTED THAT DEVICE WAS BROKEN DURING EXTRACTION OF KNEE COMPONENTS. SURGICAL DELAY UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17239187 ↗
  • MalfunctionMar 8, 2023Crack; Crack

    IT WAS REPORTED THAT TIP BROKE OFF EXTRACTOR. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16504627 ↗
  • MalfunctionFeb 24, 2023Break; Break

    THE FOLLOWING DEVICE WAS RECEIVED AS BLIND UNIT. PRODUCT CODE: 242102000. PRODUCT NAME: FEM/TIB EXTRAC 11X7 3/4. QTY: 1. HOWEVER, IT COULDN¿T BE ASSOCIATED WITH A COMPLAINT OR ANY OTHER EXISTING RECORD. THIS COMPLAINT INVOLVES ONE (1) DEVICE.

    Patient problems: Insufficient Information

    MAUDE report 16438484 ↗
  • MalfunctionFeb 20, 2023Break; Break

    THE FOLLOWING DEVICE WAS RECEIVED AS BLIND UNIT PRODUCT CODE: 242102000 PRODUCT NAME: FEM/TIB EXTRAC 11X7 3/4 IN LOT NUMBER: SO2058069 QTY: 1 HOWEVER, IT COULDN¿T BE ASSOCIATED WITH A COMPLAINT OR ANY OTHER EXISTING RECORD. THIS COMPLAINT INVOLVES ONE (1) DEVICE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16403593 ↗
  • MalfunctionAug 25, 2022Break; Break

    IT WAS REPORTED THAT THE HANDLE BROKE IN HALF. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15293619 ↗
  • MalfunctionJul 13, 2022Break; Break

    IT WAS REPORTED THAT WHILE ATTEMPTING TO EXTRACT A TOTAL KNEE, THE SURGEON BROKE THE TIPS AND HANDLE OF THE MORELAND OSTEOTOMES. NEW OSTEOTOMES WERE OPENED AND THE SURGERY WAS SUCCESSFULLY COMPLETED. NO ADVERSE EVENTS OCCURRED. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15000787 ↗
  • MalfunctionJan 5, 2021Break; Break

    IT WAS REPORTED THAT A PIECE OF THE WIDGET BROKE OFF WHILE THE SURGEON WAS TRYING TO IMPACT THE FEMUR IMPLANT OFF. THE PIECE THAT BROKE OFF WAS RETRIEVED FROM THE PATIENT SUCCESSFULLY. NO DELAY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11118481 ↗
  • MalfunctionDec 22, 2020Break; Break

    IT WAS REPORTED THAT THE LITTLE TIP BROKE AT THE END. NOTHING LEFT IN PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11057809 ↗
  • MalfunctionDec 22, 2020Break; Break

    VOLUNTARY REPORT WAS RECEIVED UNDER MW5097975. IT WAS REPORTED THAT WHILE ATTEMPTING TO HAMMER OUT A FEMORAL IMPLANT DURING AN ORTHOPEDIC SURGICAL PROCEDURE, DISTAL END OF THE OSTEOTOME BROKE OFF, FLEW UP AND STRUCK THE OVERHEAD LIGHT AND LANDED BACK ON THE SURGICAL TABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11057574 ↗
  • MalfunctionOct 23, 2020Break; Naturally Worn; Illegible Information; Break; Naturally Worn; Illegible Information

    IT WAS REPORTED THAT THE HANDLE FRACTURED, ALL FRACTURED PIECES RECOVERED. NOTHING LEFT IN PATIENT AND NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10726534 ↗
  • MalfunctionOct 2, 2020Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE INTER-CONDL CHISEL WAS CHIPPED AND THE FEMORAL/TIBIAL EXTRACTOR WAS BENT. THESE DID NOT CAUSE DELAY IN SURGERY.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 10624082 ↗
  • MalfunctionNov 6, 2019Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT WAS BROKEN DURING SURGERY. CHIPPED OFF THE END.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9284285 ↗
  • MalfunctionJun 21, 2019Break; Break

    IT WAS REPORTED THAT THE FEM/TIB EXTRACTOR BROKE DURING CASE. NO SURGICAL DELAY HAPPEN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8723689 ↗
  • MalfunctionJan 10, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE INSTRUMENT WAS REPORTED FOR UNKNOWN REASON.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8236004 ↗
  • MalfunctionAug 14, 2018Crack; Crack

    FEM/TIB EXTRACTOR HANDLE CRACKED AND THE TIP OF THE INSTRUMENT BROKE OFF WHILE REMOVING A FEMORAL COMPONENT. ALL PIECES WERE RETRIEVED AND THERE WAS NO ADDITIONAL TIME ADDED TO THE SURGERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7781320 ↗
  • MalfunctionMar 27, 2018Break; Break

    IT WAS REPORTED THAT THE IMPACTOR HANDLE BROKE. ALL PIECES REMOVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7372991 ↗