S-ROM 53-1206Recalls & reported problems
Depuy (Ireland) · S-ROM SLIDE HAMMER SHAFT · Class 1
All models of S-ROM (185)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10603295171560), received between Jul 12, 2018 and Sep 5, 2023 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break8
- Naturally Worn4
- Device-Device Incompatibility2
- Material Deformation1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient4
- No Code Available1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 5, 2023Break; Break
ADDITIONAL INFORMATION RECEIVED. A. DID IT BREAK INTO TWO OR MORE PIECES? IF NO, PLEASE SPECIFY WHAT IS THE ALLEGED DEFICIENCY, IS IT BENT, CRACKED, STRIPPED, SCRATCHED, WORN, CROSS THREADED OR ANY DEVICE INTERACTION ISSUES? -BROKE IN TWO PIECES B. WERE ALL PIECES OF THE BROKEN INSTRUMENT RETRIEVED FROM THE PATIENT? -ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17684473 ↗MalfunctionAug 2, 2023Break; Break
IT WAS REPORTED THAT THE DEVICES BROKE AT THREADS. ALL PIECES WERE RECOVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17444250 ↗MalfunctionJan 26, 2023Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
IT WAS REPORTED THAT THE PIN IN T HANDLE OF MILLER FRAME CAME LOOSE. LOOP EXTRACTOR WON'T SCREW INTO SLAP HAMMER HANDLE BECAUSE IT WAS STRIPPED. NO SURGICAL DELAY WAS THERE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16247208 ↗MalfunctionJun 3, 2022Break; Break
IT WAS REPORTED THAT THE SLAP HAMMER BROKE OFF UPON EXTRACTION. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14595078 ↗MalfunctionFeb 22, 2022Break; Break
THE THREADS ON THREADED ADAPTER BROKE OFF IN SLAP HAMMER DURING USE. NO PATIENT HARM OR SURGICAL DELAY OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13579906 ↗MalfunctionFeb 7, 2022Naturally Worn; Naturally Worn
THREADS ON THE INSTRUMENT ARE STRIPPED. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13477143 ↗MalfunctionOct 11, 2021Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE HOOK AND THREADED SHAFT OF THE SROM EXTRACTION ROD IS BELIEVED TO BE CROSS-THREADED AND WILL NOT COME APART. THIS AFFECTS ALL 4 PIECES ON THE SLAP HAMMER EXTRACTION TOOL. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12605293 ↗MalfunctionApr 26, 2021Break; Break
IT WAS REPORTED THAT THE SROM SLAP HAMMER WILL NO LONGER ENGAGE WITH LOOPED CONNECTION, IT APPEARS TO HAVE A BROKEN PIECE OF METAL INSIDE THE THREADS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11726897 ↗MalfunctionAug 25, 2020Device-Device Incompatibility; Device-Device Incompatibility
S-ROM EXTRACTOR HAS A SPLIT ON THE HANDLE. CANNOT DISMANTLE THE HOOK FROM THE EXTRACTOR HANDLE. NO SURGICAL DELAY OR PATIENT CONSEQUENCES REPORTED.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 10449626 ↗MalfunctionMar 27, 2019Break; Material Deformation; Material Twisted/Bent; Naturally Worn; Break; Material Deformation; Material Twisted/Bent; Naturally Worn
IT WAS REPORTED THAT THE FEMALE THREADING STRIPPED. DAMAGED MALE THREADING DURING USE. IT WAS INDICATED BY THE SALES REP THAT NO PIECES CAME OFF THE THREADING AND NO PIECES WERE IN THE PATIENT. THREADS ARE SHEARING OFF. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8455298 ↗MalfunctionOct 12, 2018Break; Break
IT WAS REPORTED THAT THE S-ROM SLAP HAMMER BROKE WHILE EXTRACTING THE STEM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7960843 ↗MalfunctionJul 12, 2018Break; Break
THE SHAFT IS DISSOCIATED OF THE HANDLE. (B)(4) FOR REPLACEMENT. PATIENT CONSEQUENCE? NO. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? YES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7682310 ↗
FDA clearance and classification
The FDA regulates product code HWB (extractor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the S-ROM 53-1206
Has the S-ROM 53-1206 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10603295171560), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the S-ROM 53-1206?
The FDA's MAUDE database lists 12 adverse event reports for this device, received between Jul 12, 2018 and Sep 5, 2023. In those reports, the most common specific device problems are “Break” (8), “Naturally Worn” (4) and “Device-Device Incompatibility” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the S-ROM 53-1206?
None of the reports are classified as deaths, 0 as injuries and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the S-ROM 53-1206 FDA approved?
Devices of this type (product code HWB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).
Who makes the S-ROM 53-1206, and is it still sold?
The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand S-ROM, model 53-1206). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other EXTRACTOR devices
- ATTUNE 2545-00-138Depuy (Ireland) · EXTRACTOR390 reports
- PINNACLE 2217-50-001Depuy (Ireland) · EXTRACTOR242 reports
- RECLAIM 2975-00-635Depuy (Ireland) · EXTRACTOR156 reports
- ATTUNE 2545-00-060Depuy (Ireland) · EXTRACTOR150 reports
- ESC PINNACLE 2218-00-010Depuy (Ireland) · EXTRACTOR150 reports
- NA 2125-00-600Depuy Orthopaedics, Inc. · EXTRACTOR129 reports
- P.F.C. 86-5226Depuy (Ireland) · EXTRACTOR80 reports
- NA 03.010.000Synthes GmbH · EXTRACTOR69 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page