S-ROM 53-4400Recalls & reported problems
Depuy (Ireland) · S-ROM STEM EXTRACTOR LOOP · Class 1
All models of S-ROM (185)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy S-ROM stem extractor loop is a surgical instrument used to pull a hip implant stem out during hip surgery. The 34 reports mostly describe the loop breaking, cracking or bending during use, sometimes snapping at the threads; several mention all pieces were recovered and surgery was not delayed. No deaths and no injuries are reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 34
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 10603295171973), received between Sep 13, 2017 and Apr 2, 2026 — classified as 34 malfunctions.
The most frequently reported specific device problem is “Break” (24 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break24
- Crack7
- Device-Device Incompatibility3
- Naturally Worn3
- Material Deformation2
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions17
- No Consequences Or Impact To Patient12
- No Information2
- Insufficient Information1
- No Code Available1
- No Known Impact Or Consequence To Patient1
- Not Applicable1
Recent adverse event reports
The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 2, 2026Break; Material Deformation
IT WAS REPORTED THAT THE DEVICE BROKE DURING HIP SURGERY. IT BENT IN HALF WHEN USING IT FOR EXTRACTION. THERE WAS NOTHING LEFT IN THE PATIENT. IT DID NOT CAUSE ANY DELAYS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24766557 ↗MalfunctionOct 24, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT THE ARON LOOP WAS NOTICEABLE BENT/WARPED/SPRUNG UPON INSPECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23381848 ↗MalfunctionAug 29, 2025Break; Break
IT WAS REPORTED THE INSTRUMENT BROKE DURING HIP SURGERY. PART OF IT HAS SNAPPED; NOTHING WAS LEFT INSIDE THE PATIENT AND THERE WERE NO DELAYS TO THE SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22923245 ↗MalfunctionOct 30, 2023Naturally Worn; Unintended Movement; Naturally Worn; Unintended Movement
IT WAS REPORTED THAT THE THREADED LOOP ATTACHMENT IS STRIPPED AND KEPT COMING OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18036104 ↗MalfunctionAug 22, 2023Break; Break
IT WAS REPORTED THAT THE INSTRUMENT BROKE DURING HIP SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17597248 ↗MalfunctionAug 2, 2023Break; Break
IT WAS REPORTED THAT THE DEVICES BROKE AT THREADS. ALL PIECES WERE RECOVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17444247 ↗MalfunctionJul 28, 2023Crack; Crack
IT WAS REPORTED THAT THE HORSESHOE LOOKING PIECE OF THE SLAP HAMMER GOT A CRACK THROUGH IT. IT DID NOT HAPPEN DURING SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17413343 ↗MalfunctionJun 20, 2023Break; Break
IT WAS REPORTED THAT THE THREADED PORTION OF THE LOOP EXTRACTOR WAS BROKEN OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17165151 ↗MalfunctionJun 19, 2023Break; Break
IT WAS REPORTED THAT DEVICE IS BROKEN, NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17154966 ↗MalfunctionMar 29, 2023Break; Break
ADDITIONAL INFORMATION RECEIVED CONFIRMED THAT THE EXTRACTED FEMORAL STEM MENTIONED IS NOT A DEPUY PRODUCT. LOT NUMBER WAS ILLEGIBLE IN THE ADAPTERS. IT WAS THE THREADED PART THAT BROKE FROM THE BODY. THERE WERE NO PIECES LEFT IN THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16639947 ↗MalfunctionFeb 27, 2023Crack; Crack
DURING SURGERY YESTERDAY, (B)(6) 2023, THE FOLLOWING SROM INSTRUMENTS WERE DAMAGED DURING THE PROCEDURE WHEN TRYING TO REMOVE THE SIZE 2 MILLER WHICH GOT STUCK IN THE PATIENT¿S FEMUR.
Patient problems: Insufficient Information
MAUDE report 16444815 ↗MalfunctionOct 6, 2022Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
IT WAS REPORTED THAT THE SROM LOOP EXTRACTOR WAS CROSS-THREADED IN THE SLAP HAMMER AND WOULD NOT COME APART AT THE END OF THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15552292 ↗MalfunctionJul 5, 2022Break; Break
DURING REMOVAL OF A STEM THE LOOP BROKE. NO FRAGMENTS. CLEAN BREAK AT TIP. THERE WAS NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14914936 ↗MalfunctionMar 15, 2022Break; Crack; Break; Crack
IT WAS REPORTED THAT THE SROM LOOP EXTRACTOR FOR SLAP HAMMER BROKE DURING STEM EXPLANT. ALL PIECES WERE RETRIEVED AND THERE WAS NO DELAY IN PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13770052 ↗MalfunctionNov 22, 2021Break; Break
IT WAS REPORTED THAT THE SROM LOOP EXTRACTOR SPLIT IN HALF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12852447 ↗MalfunctionOct 11, 2021Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE HOOK AND THREADED SHAFT OF THE SROM EXTRACTION ROD IS BELIEVED TO BE CROSS-THREADED AND WILL NOT COME APART. THIS AFFECTS ALL 4 PIECES ON THE SLAP HAMMER EXTRACTION TOOL. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12605290 ↗MalfunctionSep 23, 2021Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
IT WAS REPORTED THAT THE LOOPED EXTRACTOR BROKE WHILE TRYING TO REMOVE STEM. THREADS FROM EXTRACTOR LOOP STUCK IN SLAP HAMMER HANDLE. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12519106 ↗MalfunctionApr 16, 2021Break; Break
CONSIGNMENT SROM EXTRACTOR STEM LOOP BROKE WHILE USING IT TO EXTRACT A COMPETITOR STEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11678354 ↗MalfunctionOct 13, 2020Break; Break
ADDITIONAL INFORMATION RECEIVED STATED THAT THE EVENT HAPPENED DURING THE OPERATION. NO PARTS WERE LEFT BEHIND ON THE STERILE FIELD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10672971 ↗MalfunctionAug 25, 2020Device-Device Incompatibility; Device-Device Incompatibility
S-ROM EXTRACTOR HAS A SPLIT ON THE HANDLE. CANNOT DISMANTLE THE HOOK FROM THE EXTRACTOR HANDLE. NO SURGICAL DELAY OR PATIENT CONSEQUENCES REPORTED.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 10449653 ↗MalfunctionJun 10, 2020Crack; Crack
ADDITIONAL INFORMATION RECEIVED STATED THAT THE BASE OF THE HOOK WHERE IT ATTACHES TO THE SHAFT WAS CRACKED, BUT STILL IN ONE PIECE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10138450 ↗MalfunctionJan 17, 2020Break; Break
THE INSTRUMENT HAD BROKEN. THE REPS THAT WERE COVERING WERE UNABLE TO PROVIDE ME WITH ADDITIONAL INFORMATION (DAY IN WHICH THEY BROKE, DURATION OF SURGERY, PATIENT INFORMATION, ETC.).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9601961 ↗MalfunctionDec 16, 2019Break; Break
ADDITIONAL INFORMATION RECEIVED INDICATING THAT NOTHING WAS DELAYED. THE SLEEVE EXTRACTOR WAS NOT ABLE TO BE REMOVED AND THE LOOP BROKE AT THE LOOP.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9475830 ↗MalfunctionAug 22, 2019Break; Break
IT WAS REPORTED THAT THE SLAP HAMMER HOOK BROKE OPEN DURING NORMAL USE. NO PIECES BROKE OFF OF INSTRUMENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8920442 ↗MalfunctionAug 7, 2019Crack; Crack
IT WAS REPORTED THAT THE LOOP CRACKED / SEPARATED WHEN TRYING TO BACK SLAP A STEM.
Patient problems: Not Applicable
MAUDE report 8871080 ↗
FDA clearance and classification
The FDA regulates product code HWB (extractor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the S-ROM 53-4400
Has the S-ROM 53-4400 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10603295171973), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the S-ROM 53-4400?
The FDA's MAUDE database lists 34 adverse event reports for this device, received between Sep 13, 2017 and Apr 2, 2026. In those reports, the most common specific device problems are “Break” (24), “Crack” (7) and “Device-Device Incompatibility” (3); the most common patient problem is “Not Applicable” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the S-ROM 53-4400?
None of the reports are classified as deaths, 0 as injuries and 34 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the S-ROM 53-4400 FDA approved?
Devices of this type (product code HWB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).
Who makes the S-ROM 53-4400, and is it still sold?
The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand S-ROM, model 53-4400). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other EXTRACTOR devices
- ATTUNE 2545-00-138Depuy (Ireland) · EXTRACTOR390 reports
- PINNACLE 2217-50-001Depuy (Ireland) · EXTRACTOR242 reports
- RECLAIM 2975-00-635Depuy (Ireland) · EXTRACTOR156 reports
- ATTUNE 2545-00-060Depuy (Ireland) · EXTRACTOR150 reports
- ESC PINNACLE 2218-00-010Depuy (Ireland) · EXTRACTOR150 reports
- NA 2125-00-600Depuy Orthopaedics, Inc. · EXTRACTOR129 reports
- P.F.C. 86-5226Depuy (Ireland) · EXTRACTOR80 reports
- NA 03.010.000Synthes GmbH · EXTRACTOR69 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page