Depuy (Ireland) · S-ROM STEM EXTRACTOR LOOP · Class 1

Quick summary

This DePuy S-ROM stem extractor loop is a surgical instrument used to pull a hip implant stem out during hip surgery. The 34 reports mostly describe the loop breaking, cracking or bending during use, sometimes snapping at the threads; several mention all pieces were recovered and surgery was not delayed. No deaths and no injuries are reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 10603295171973), received between Sep 13, 2017 and Apr 2, 2026 — classified as 34 malfunctions.

The most frequently reported specific device problem is “Break” (24 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break24
  • Crack7
  • Device-Device Incompatibility3
  • Naturally Worn3
  • Material Deformation2
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions17
  • No Consequences Or Impact To Patient12
  • No Information2
  • Insufficient Information1
  • No Code Available1
  • No Known Impact Or Consequence To Patient1
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 2, 2026Break; Material Deformation

    IT WAS REPORTED THAT THE DEVICE BROKE DURING HIP SURGERY. IT BENT IN HALF WHEN USING IT FOR EXTRACTION. THERE WAS NOTHING LEFT IN THE PATIENT. IT DID NOT CAUSE ANY DELAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24766557 ↗
  • MalfunctionOct 24, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE ARON LOOP WAS NOTICEABLE BENT/WARPED/SPRUNG UPON INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23381848 ↗
  • MalfunctionAug 29, 2025Break; Break

    IT WAS REPORTED THE INSTRUMENT BROKE DURING HIP SURGERY. PART OF IT HAS SNAPPED; NOTHING WAS LEFT INSIDE THE PATIENT AND THERE WERE NO DELAYS TO THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22923245 ↗
  • MalfunctionOct 30, 2023Naturally Worn; Unintended Movement; Naturally Worn; Unintended Movement

    IT WAS REPORTED THAT THE THREADED LOOP ATTACHMENT IS STRIPPED AND KEPT COMING OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18036104 ↗
  • MalfunctionAug 22, 2023Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT BROKE DURING HIP SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17597248 ↗
  • MalfunctionAug 2, 2023Break; Break

    IT WAS REPORTED THAT THE DEVICES BROKE AT THREADS. ALL PIECES WERE RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17444247 ↗
  • MalfunctionJul 28, 2023Crack; Crack

    IT WAS REPORTED THAT THE HORSESHOE LOOKING PIECE OF THE SLAP HAMMER GOT A CRACK THROUGH IT. IT DID NOT HAPPEN DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17413343 ↗
  • MalfunctionJun 20, 2023Break; Break

    IT WAS REPORTED THAT THE THREADED PORTION OF THE LOOP EXTRACTOR WAS BROKEN OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17165151 ↗
  • MalfunctionJun 19, 2023Break; Break

    IT WAS REPORTED THAT DEVICE IS BROKEN, NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17154966 ↗
  • MalfunctionMar 29, 2023Break; Break

    ADDITIONAL INFORMATION RECEIVED CONFIRMED THAT THE EXTRACTED FEMORAL STEM MENTIONED IS NOT A DEPUY PRODUCT. LOT NUMBER WAS ILLEGIBLE IN THE ADAPTERS. IT WAS THE THREADED PART THAT BROKE FROM THE BODY. THERE WERE NO PIECES LEFT IN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16639947 ↗
  • MalfunctionFeb 27, 2023Crack; Crack

    DURING SURGERY YESTERDAY, (B)(6) 2023, THE FOLLOWING SROM INSTRUMENTS WERE DAMAGED DURING THE PROCEDURE WHEN TRYING TO REMOVE THE SIZE 2 MILLER WHICH GOT STUCK IN THE PATIENT¿S FEMUR.

    Patient problems: Insufficient Information

    MAUDE report 16444815 ↗
  • MalfunctionOct 6, 2022Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    IT WAS REPORTED THAT THE SROM LOOP EXTRACTOR WAS CROSS-THREADED IN THE SLAP HAMMER AND WOULD NOT COME APART AT THE END OF THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15552292 ↗
  • MalfunctionJul 5, 2022Break; Break

    DURING REMOVAL OF A STEM THE LOOP BROKE. NO FRAGMENTS. CLEAN BREAK AT TIP. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14914936 ↗
  • MalfunctionMar 15, 2022Break; Crack; Break; Crack

    IT WAS REPORTED THAT THE SROM LOOP EXTRACTOR FOR SLAP HAMMER BROKE DURING STEM EXPLANT. ALL PIECES WERE RETRIEVED AND THERE WAS NO DELAY IN PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13770052 ↗
  • MalfunctionNov 22, 2021Break; Break

    IT WAS REPORTED THAT THE SROM LOOP EXTRACTOR SPLIT IN HALF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12852447 ↗
  • MalfunctionOct 11, 2021Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE HOOK AND THREADED SHAFT OF THE SROM EXTRACTION ROD IS BELIEVED TO BE CROSS-THREADED AND WILL NOT COME APART. THIS AFFECTS ALL 4 PIECES ON THE SLAP HAMMER EXTRACTION TOOL. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12605290 ↗
  • MalfunctionSep 23, 2021Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT THE LOOPED EXTRACTOR BROKE WHILE TRYING TO REMOVE STEM. THREADS FROM EXTRACTOR LOOP STUCK IN SLAP HAMMER HANDLE. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12519106 ↗
  • MalfunctionApr 16, 2021Break; Break

    CONSIGNMENT SROM EXTRACTOR STEM LOOP BROKE WHILE USING IT TO EXTRACT A COMPETITOR STEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11678354 ↗
  • MalfunctionOct 13, 2020Break; Break

    ADDITIONAL INFORMATION RECEIVED STATED THAT THE EVENT HAPPENED DURING THE OPERATION. NO PARTS WERE LEFT BEHIND ON THE STERILE FIELD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10672971 ↗
  • MalfunctionAug 25, 2020Device-Device Incompatibility; Device-Device Incompatibility

    S-ROM EXTRACTOR HAS A SPLIT ON THE HANDLE. CANNOT DISMANTLE THE HOOK FROM THE EXTRACTOR HANDLE. NO SURGICAL DELAY OR PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 10449653 ↗
  • MalfunctionJun 10, 2020Crack; Crack

    ADDITIONAL INFORMATION RECEIVED STATED THAT THE BASE OF THE HOOK WHERE IT ATTACHES TO THE SHAFT WAS CRACKED, BUT STILL IN ONE PIECE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10138450 ↗
  • MalfunctionJan 17, 2020Break; Break

    THE INSTRUMENT HAD BROKEN. THE REPS THAT WERE COVERING WERE UNABLE TO PROVIDE ME WITH ADDITIONAL INFORMATION (DAY IN WHICH THEY BROKE, DURATION OF SURGERY, PATIENT INFORMATION, ETC.).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9601961 ↗
  • MalfunctionDec 16, 2019Break; Break

    ADDITIONAL INFORMATION RECEIVED INDICATING THAT NOTHING WAS DELAYED. THE SLEEVE EXTRACTOR WAS NOT ABLE TO BE REMOVED AND THE LOOP BROKE AT THE LOOP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9475830 ↗
  • MalfunctionAug 22, 2019Break; Break

    IT WAS REPORTED THAT THE SLAP HAMMER HOOK BROKE OPEN DURING NORMAL USE. NO PIECES BROKE OFF OF INSTRUMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8920442 ↗
  • MalfunctionAug 7, 2019Crack; Crack

    IT WAS REPORTED THAT THE LOOP CRACKED / SEPARATED WHEN TRYING TO BACK SLAP A STEM.

    Patient problems: Not Applicable

    MAUDE report 8871080 ↗