SIGMA LCS HIGH PERFORMANCE 9505-02-303Recalls & reported problems

Depuy Orthopaedics, Inc. · SIGMA LCS HIGH PERFORMANCE STERILE THREADED HEADED PINS · Class 1

All models of SIGMA LCS HIGH PERFORMANCE (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10603295228912), received between Jun 22, 2017 and Nov 14, 2025 — classified as 2 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Adverse Event Without Identified Device or Use Problem2
  • Appropriate Device Problem Term/Code Not Available2
  • Bent1
  • Device-Device Incompatibility1
  • Material Deformation1
  • Mechanical Jam1
  • Physical Resistance/Sticking1

Patient problems

  • No Consequences Or Impact To Patient3
  • No Clinical Signs, Symptoms or Conditions2
  • No Information2
  • Muscle/Tendon Damage1
  • Unspecified Infection1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2025Break; Break

    IT WAS REPORTED THAT THE THREADED HEADED PIN SNAPPED INTO TWO PIECES WHEN INSERTING IT UNDER POWER INTO THE FEMORAL SIZER WITH THE PIN DRIVER. THERE WAS NO PATIENT HARM OR ADVERSE EVENT THAT OCCURRED AS A RESULT OF THIS INCIDENT, NOR WAS THERE A SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23559952 ↗
  • InjuryDec 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CLINICAL ADVERSE EVENTS RECEIVED FOR IT BAND TENDINITIS. DEVICE AND PROCEDURE (RELATEDNESS): DEVICE RELATED: DEFINITELY NOT. PROCEDURE RELATED: POSSIBLY. DATE OF EVENT: 26/NOV/2024. DATE OF IMPLANT: NO INFORMATION PROVIDED. DATE OF REVISION: NO INFORMATION PROVIDED. DEVICE LOCATION: LEFT. TREATMENT/IMPACT: IRRIGATION & DEBRIDEMENT. DEPUY SYNTHES PRODUCT USED: CATALOG ID: 151820032, LOT ID: 3802592, COMPONENT TYPE: PATELLAR, DESCRIPTION: ATTUNE MEDIAL DOME PAT 32MM; CATALOG ID: 151820308, LOT ID: JP9042, COMPONENT TYPE: INSERT, DESCRIPTION: ATTUNE LEFT MEDIAL STABILIZED INSERT SIZE 3 8MM; CATALOG ID: 150400103, LOT ID: 8093073, COMPONENT TYPE: FEMORAL, DESCRIPTION: ATTUNE C/RET FEM LT SZ 3 CEMENTED; CATALOG ID: 150670003, LOT ID: 9854210, COMPONENT TYPE: TIBIAL, DESCRIPTION: ATTUNE FB CEMENTED TIBIAL BASE SIZE 3; CATALOG ID: 249095000, LOT ID: M0476R, COMPONENT TYPE: PIN, DESCRIPTION: AMK FIXATION PINS 1/8 DIA X 3; CATALOG ID: 950502303, LOT ID: M1312Z, COMPONENT TYPE: ACC, DESCRIPTION: SIGMA LCS HIGH PERFORMANCE, STERILE THREADED HEADED PINS; CATALOG ID: 3312040, LOT ID: 3807417, COMPONENT TYPE: CEMENT, DESCRIPTION: SMART SET CMW 1 40G.

    Patient problems: Muscle/Tendon Damage

    MAUDE report 20904074 ↗
  • InjuryMay 17, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    STUDY NO: DOTS CLINICAL ADVERSE EVENT RECEIVED FOR POSSIBLE INFECTION. EVENT IS SERIOUS AND IS CONSIDERED MODERATE. EVENT IS POSSIBLY RELATED TO BOTH DEVICE AND PROCEDURE. DATE OF IMPLANT: (B)(6) 2023 DATE OF EVENT: (B)(6) 2024 (LEFT KNEE) TREATMENT: ASPIRATION.

    Patient problems: Unspecified Infection

    MAUDE report 19342402 ↗
  • MalfunctionMar 31, 2023Break; Device-Device Incompatibility; Material Deformation; Break; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE HEAD OF A THREADED PIN WAS SHEARED OFF WHILE IN THE A CHAMFER CUT BLOCK. UNABLE TO REMOVE PIN FROM BLOCK. NO PIECES BROKE OFF OR WERE LEFT IN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16654267 ↗
  • MalfunctionJul 22, 2019Mechanical Jam; Physical Resistance/Sticking; Mechanical Jam; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE HEADED PIN BECAME STUCK IN SIZE 3 REVISION CUT BLOCK. ALL PIECES WERE RETRIEVED AND THE SURGERY WAS NOT DELAYED. RETURNING BLOCK WITH PIN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8811811 ↗
  • MalfunctionDec 20, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT PINS GOT STUCK IN UNI TIBIAL CUTTING GUIDE. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7134106 ↗
  • MalfunctionDec 20, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT PINS GOT STUCK IN UNI TIBIAL CUTTING GUIDE. NO SURGICAL DELAY

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7134046 ↗
  • MalfunctionJul 18, 2017Bent; Bent

    IT WAS REPORTED THAT THE SURGEON TRIED TO REMOVE HEADED PIN FROM THE TIBIA CUTTING GUIDE AND IT GOT STUCK, WHICH IT CAUSED TO BEND.

    Patient problems: No Information

    MAUDE report 6721852 ↗
  • MalfunctionJun 22, 2017Break; Break

    IT WAS REPORTED THAT THE SURGEON ATTEMPTED TO DRILL HEADED PIN THROUGH PIN HOLE ON THE LEFT MEDIAL PROXIMAL TIBIA CUTTING GUIDE. PIN SPUN AND GOT STUCK CAUSING THE TIP TO BREAK OFF INSIDE THE PIN DRIVER. A SECOND PIN DRIVER WAS USED FOR THE REMAINDER OF THE CASE.

    Patient problems: No Information

    MAUDE report 6659598 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →