SIGMA LCS HIGH PERFORMANCERecalls & reported problems across all models
Depuy Orthopaedics, Inc. · 8 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 138
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls
- 0
We found 8 models of SIGMA LCS HIGH PERFORMANCE by Depuy Orthopaedics, Inc. in the FDA's device registry (GUDID). Together they are named in 138 adverse event reports in the FDA's MAUDE database, received between Jun 22, 2017 and Nov 14, 2025 — classified as 2 injuries, 136 malfunctions.
The most frequently reported specific device problem is “Break” (85 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported SIGMA LCS HIGH PERFORMANCE problems
Device problems
- Break85
- Appropriate Device Problem Term/Code Not Available16
- Naturally Worn14
- Device-Device Incompatibility12
- Loose or Intermittent Connection10
- Material Twisted/Bent7
- Failure to Cut6
- Material Deformation4
Patient problems
- No Consequences Or Impact To Patient88
- No Clinical Signs, Symptoms or Conditions34
- No Information7
- No Code Available5
- Not Applicable5
- Insufficient Information2
- Muscle/Tendon Damage1
- Unspecified Infection1
SIGMA LCS HIGH PERFORMANCE models
8 models, most reported first. Each model has its own page with its reports, recalls and device details.
- SIGMA LCS HIGH PERFORMANCE 9505-02-012PUNCH, SURGICAL85 reports
- SIGMA LCS HIGH PERFORMANCE 9505-02-011PUNCH, SURGICAL30 reports
- SIGMA LCS HIGH PERFORMANCE 9505-02-303ORTHOPEDIC MANUAL SURGICAL INSTRUMENT9 reports
- SIGMA LCS HIGH PERFORMANCE 9505-02-015PUNCH, SURGICAL8 reports
- SIGMA LCS HIGH PERFORMANCE 9505-02-014PUNCH, SURGICAL3 reports
- SIGMA LCS HIGH PERFORMANCE 9505-02-022ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1 report
- SIGMA LCS HIGH PERFORMANCE 9505-02-023ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1 report
- SIGMA LCS HIGH PERFORMANCE 9505-02-034INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1 report
Recent SIGMA LCS HIGH PERFORMANCE adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionNov 14, 2025Break; Break
IT WAS REPORTED THAT THE THREADED HEADED PIN SNAPPED INTO TWO PIECES WHEN INSERTING IT UNDER POWER INTO THE FEMORAL SIZER WITH THE PIN DRIVER. THERE WAS NO PATIENT HARM OR ADVERSE EVENT THAT OCCURRED AS A RESULT OF THIS INCIDENT, NOR WAS THERE A SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23559952 ↗InjuryDec 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CLINICAL ADVERSE EVENTS RECEIVED FOR IT BAND TENDINITIS. DEVICE AND PROCEDURE (RELATEDNESS): DEVICE RELATED: DEFINITELY NOT. PROCEDURE RELATED: POSSIBLY. DATE OF EVENT: 26/NOV/2024. DATE OF IMPLANT: NO INFORMATION PROVIDED. DATE OF REVISION: NO INFORMATION PROVIDED. DEVICE LOCATION: LEFT. TREATMENT/IMPACT: IRRIGATION & DEBRIDEMENT. DEPUY SYNTHES PRODUCT USED: CATALOG ID: 151820032, LOT ID: 3802592, COMPONENT TYPE: PATELLAR, DESCRIPTION: ATTUNE MEDIAL DOME PAT 32MM; CATALOG ID: 151820308, LOT ID: JP9042, COMPONENT TYPE: INSERT, DESCRIPTION: ATTUNE LEFT MEDIAL STABILIZED INSERT SIZE 3 8MM; CATALOG ID: 150400103, LOT ID: 8093073, COMPONENT TYPE: FEMORAL, DESCRIPTION: ATTUNE C/RET FEM LT SZ 3 CEMENTED; CATALOG ID: 150670003, LOT ID: 9854210, COMPONENT TYPE: TIBIAL, DESCRIPTION: ATTUNE FB CEMENTED TIBIAL BASE SIZE 3; CATALOG ID: 249095000, LOT ID: M0476R, COMPONENT TYPE: PIN, DESCRIPTION: AMK FIXATION PINS 1/8 DIA X 3; CATALOG ID: 950502303, LOT ID: M1312Z, COMPONENT TYPE: ACC, DESCRIPTION: SIGMA LCS HIGH PERFORMANCE, STERILE THREADED HEADED PINS; CATALOG ID: 3312040, LOT ID: 3807417, COMPONENT TYPE: CEMENT, DESCRIPTION: SMART SET CMW 1 40G.
Patient problems: Muscle/Tendon Damage
MAUDE report 20904074 ↗InjuryMay 17, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
STUDY NO: DOTS CLINICAL ADVERSE EVENT RECEIVED FOR POSSIBLE INFECTION. EVENT IS SERIOUS AND IS CONSIDERED MODERATE. EVENT IS POSSIBLY RELATED TO BOTH DEVICE AND PROCEDURE. DATE OF IMPLANT: (B)(6) 2023 DATE OF EVENT: (B)(6) 2024 (LEFT KNEE) TREATMENT: ASPIRATION.
Patient problems: Unspecified Infection
MAUDE report 19342402 ↗MalfunctionJun 5, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
IT WAS REPORTED THAT SPIKED KEEL PUNCH WAS BROKEN WHERE IT CONNECTS TO THE INSERTER HANDLE. SO IT DID NOT STAY ON THE INSERTER IN ORDER TO IMPACT. THERE WAS NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17060782 ↗MalfunctionMar 31, 2023Break; Device-Device Incompatibility; Material Deformation; Break; Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT THE HEAD OF A THREADED PIN WAS SHEARED OFF WHILE IN THE A CHAMFER CUT BLOCK. UNABLE TO REMOVE PIN FROM BLOCK. NO PIECES BROKE OFF OR WERE LEFT IN THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16654267 ↗MalfunctionDec 9, 2022Break; Break
IT WAS REPORTED THAT YESTERDAY, WHILE COVERING ONE OF THE SURGEON'S SIGMA KNEE REPLACEMENT CASES, HIS PHYSICIAN ASSISTANT ASKED THE REP WHAT IT WOULD TAKE TO REPLACEMENT SOME OF THE KEEL PUNCHES IN THEIR HOSPITAL'S CONSIGNMENT INSTRUMENTATION TRAYS. SHE THEN SHOWED THE REP HOW ONE OR BOTH OF THE SMALL METAL TINES ON THE CONNECTION END OF THE KEEL PUNCH THEY WERE ABOUT TO USE WERE MISSING. SHE EXPLAINED THAT THE PUNCHES CAN STILL BE USED, EVEN WITHOUT THE MISSING TINES, BUT THEY DON'T FIT AS SECURELY ON THE IMPACTION HANDLES WITHOUT THE MISSING TINES. SALES REP EXPLAINED OUR COMPANY'S DEVICE REPORTING SYSTEM AND INSTRUMENTATION REPLACEMENT PROCESS AND THEN PROCEEDED TO INSPECT ALL OF THE KEEL PUNCHES IN THE HOSPITAL'S CONSIGNMENT TRAYS UPON THE COMPLETION OF OUR CASE. A TOTAL OF FIVE PUNCHES WERE FOUND WITH BROKEN TINES. IT IS UNKNOWN WHEN AND WHERE THE TINES BROKE OFF. THE BROKEN PUNCH NEEDED IN THE SURGEON'S CURRENT CASE WAS USED WITHOUT INCIDENCE, SINCE AN IMMEDIATE REPLACEMENT WAS UNAVAILABLE. THE SURGICAL TIME OF THE CASE WAS NOT AFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15951475 ↗MalfunctionDec 9, 2022Break; Break
IT WAS REPORTED THAT YESTERDAY, WHILE COVERING ONE OF THE SURGEON'S SIGMA KNEE REPLACEMENT CASES, HIS PHYSICIAN ASSISTANT ASKED THE REP WHAT IT WOULD TAKE TO REPLACEMENT SOME OF THE KEEL PUNCHES IN THEIR HOSPITAL'S CONSIGNMENT INSTRUMENTATION TRAYS. SHE THEN SHOWED THE REP HOW ONE OR BOTH OF THE SMALL METAL TINES ON THE CONNECTION END OF THE KEEL PUNCH THEY WERE ABOUT TO USE WERE MISSING. SHE EXPLAINED THAT THE PUNCHES CAN STILL BE USED, EVEN WITHOUT THE MISSING TINES, BUT THEY DON'T FIT AS SECURELY ON THE IMPACTION HANDLES WITHOUT THE MISSING TINES. SALES REP EXPLAINED OUR COMPANY'S DEVICE REPORTING SYSTEM AND INSTRUMENTATION REPLACEMENT PROCESS AND THEN PROCEEDED TO INSPECT ALL OF THE KEEL PUNCHES IN THE HOSPITAL'S CONSIGNMENT TRAYS UPON THE COMPLETION OF OUR CASE. A TOTAL OF FIVE PUNCHES WERE FOUND WITH BROKEN TINES. IT IS UNKNOWN WHEN AND WHERE THE TINES BROKE OFF. THE BROKEN PUNCH NEEDED IN THE SURGEON'S CURRENT CASE WAS USED WITHOUT INCIDENCE, SINCE AN IMMEDIATE REPLACEMENT WAS UNAVAILABLE. THE SURGICAL TIME OF THE CASE WAS NOT AFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15951470 ↗MalfunctionDec 9, 2022Break; Break
IT WAS REPORTED THAT YESTERDAY, WHILE COVERING ONE OF THE SURGEON'S SIGMA KNEE REPLACEMENT CASES, HIS PHYSICIAN ASSISTANT ASKED THE REP WHAT IT WOULD TAKE TO REPLACEMENT SOME OF THE KEEL PUNCHES IN THEIR HOSPITAL'S CONSIGNMENT INSTRUMENTATION TRAYS. SHE THEN SHOWED THE REP HOW ONE OR BOTH OF THE SMALL METAL TINES ON THE CONNECTION END OF THE KEEL PUNCH THEY WERE ABOUT TO USE WERE MISSING. SHE EXPLAINED THAT THE PUNCHES CAN STILL BE USED, EVEN WITHOUT THE MISSING TINES, BUT THEY DON'T FIT AS SECURELY ON THE IMPACTION HANDLES WITHOUT THE MISSING TINES. SALES REP EXPLAINED OUR COMPANY'S DEVICE REPORTING SYSTEM AND INSTRUMENTATION REPLACEMENT PROCESS AND THEN PROCEEDED TO INSPECT ALL OF THE KEEL PUNCHES IN THE HOSPITAL'S CONSIGNMENT TRAYS UPON THE COMPLETION OF OUR CASE. A TOTAL OF FIVE PUNCHES WERE FOUND WITH BROKEN TINES. IT IS UNKNOWN WHEN AND WHERE THE TINES BROKE OFF. THE BROKEN PUNCH NEEDED IN THE SURGEON'S CURRENT CASE WAS USED WITHOUT INCIDENCE, SINCE AN IMMEDIATE REPLACEMENT WAS UNAVAILABLE. THE SURGICAL TIME OF THE CASE WAS NOT AFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15951408 ↗MalfunctionDec 9, 2022Break; Break
IT WAS REPORTED THAT YESTERDAY, WHILE COVERING ONE OF THE SURGEON'S SIGMA KNEE REPLACEMENT CASES, HIS PHYSICIAN ASSISTANT ASKED THE REP WHAT IT WOULD TAKE TO REPLACEMENT SOME OF THE KEEL PUNCHES IN THEIR HOSPITAL'S CONSIGNMENT INSTRUMENTATION TRAYS. SHE THEN SHOWED THE REP HOW ONE OR BOTH OF THE SMALL METAL TINES ON THE CONNECTION END OF THE KEEL PUNCH THEY WERE ABOUT TO USE WERE MISSING. SHE EXPLAINED THAT THE PUNCHES CAN STILL BE USED, EVEN WITHOUT THE MISSING TINES, BUT THEY DON'T FIT AS SECURELY ON THE IMPACTION HANDLES WITHOUT THE MISSING TINES. SALES REP EXPLAINED OUR COMPANY'S DEVICE REPORTING SYSTEM AND INSTRUMENTATION REPLACEMENT PROCESS AND THEN PROCEEDED TO INSPECT ALL OF THE KEEL PUNCHES IN THE HOSPITAL'S CONSIGNMENT TRAYS UPON THE COMPLETION OF OUR CASE. A TOTAL OF FIVE PUNCHES WERE FOUND WITH BROKEN TINES. IT IS UNKNOWN WHEN AND WHERE THE TINES BROKE OFF. THE BROKEN PUNCH NEEDED IN THE SURGEON'S CURRENT CASE WAS USED WITHOUT INCIDENCE, SINCE AN IMMEDIATE REPLACEMENT WAS UNAVAILABLE. THE SURGICAL TIME OF THE CASE WAS NOT AFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15951265 ↗MalfunctionDec 9, 2022Break; Break
IT WAS REPORTED THAT YESTERDAY, WHILE COVERING ONE OF THE SURGEON'S SIGMA KNEE REPLACEMENT CASES, HIS PHYSICIAN ASSISTANT ASKED THE REP WHAT IT WOULD TAKE TO REPLACEMENT SOME OF THE KEEL PUNCHES IN THEIR HOSPITAL'S CONSIGNMENT INSTRUMENTATION TRAYS. SHE THEN SHOWED THE REP HOW ONE OR BOTH OF THE SMALL METAL TINES ON THE CONNECTION END OF THE KEEL PUNCH THEY WERE ABOUT TO USE WERE MISSING. SHE EXPLAINED THAT THE PUNCHES CAN STILL BE USED, EVEN WITHOUT THE MISSING TINES, BUT THEY DON'T FIT AS SECURELY ON THE IMPACTION HANDLES WITHOUT THE MISSING TINES. SALES REP EXPLAINED OUR COMPANY'S DEVICE REPORTING SYSTEM AND INSTRUMENTATION REPLACEMENT PROCESS AND THEN PROCEEDED TO INSPECT ALL OF THE KEEL PUNCHES IN THE HOSPITAL'S CONSIGNMENT TRAYS UPON THE COMPLETION OF OUR CASE. A TOTAL OF FIVE PUNCHES WERE FOUND WITH BROKEN TINES. IT IS UNKNOWN WHEN AND WHERE THE TINES BROKE OFF. THE BROKEN PUNCH NEEDED IN THE SURGEON'S CURRENT CASE WAS USED WITHOUT INCIDENCE, SINCE AN IMMEDIATE REPLACEMENT WAS UNAVAILABLE. THE SURGICAL TIME OF THE CASE WAS NOT AFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15951252 ↗MalfunctionOct 3, 2022Break; Break
IT WAS REPORTED THAT THE CONNECTION FOR HANDLE ATTACHMENT IS BROKEN OFF ON BOTH ITEMS. NO LOOSE ITEMS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15527360 ↗MalfunctionSep 28, 2022Break; Break
IT WAS REPORTED THAT THE CONNECTION FOR HANDLE ATTACHMENT IS BROKEN OFF ON BOTH ITEMS. NO LOOSE ITEMS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15496223 ↗MalfunctionSep 28, 2022Break; Break
IT WAS REPORTED THAT THE CONNECTION FOR HANDLE ATTACHMENT IS BROKEN OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15496039 ↗MalfunctionJun 8, 2022Break; Break
IT WAS REPORTED THAT THE HP MBT CEM KEEL PUNCH SZ 4-7 HAS A BROKEN CONNECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14632578 ↗MalfunctionApr 22, 2022Break; Failure to Cut; Break; Failure to Cut
IT WAS REPORTED THAT DEVICE WAS BROKEN AND DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14189261 ↗
Other Depuy Orthopaedics, Inc. products
- PINNACLE ALTRX98 models · 5,631 reports
- BIOLOX DELTA40 models · 4,995 reports · 1 recall
- PINNACLE182 models · 4,587 reports
- QUICKSET7 models · 3,790 reports
- PINNACLE POROCOAT38 models · 2,112 reports
- SUMMIT POROCOAT20 models · 1,646 reports
- PINNACLE GRIPTION57 models · 1,608 reports
- DURALOC93 models · 1,262 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page