SPECIALIST 96-6515Recalls & reported problems
Depuy (Ireland) · SPECIALIST 2 PIN PULLER · Class 1
All models of SPECIALIST (80)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10603295248637), received between Feb 21, 2019 and Aug 6, 2026 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility11
- Contamination /Decontamination Problem2
- Break1
- Corroded1
- Material Twisted/Bent1
- Mechanical Jam1
- Naturally Worn1
- Positioning Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient5
- Insufficient Information2
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2026Corroded; Contamination /Decontamination Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50162881 ↗MalfunctionSep 27, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN IT WAS IDENTIFIED THAT THE DEVICE IS RUSTY. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20323447 ↗MalfunctionSep 12, 2024Device-Device Incompatibility; Device-Device Incompatibility
THE PIN PULLER DOESN¿T GRIP ONTO THE PINS SO NEEDS TO BE REPLACED.
Patient problems: Insufficient Information
MAUDE report 20198920 ↗MalfunctionAug 5, 2024Device-Device Incompatibility; Device-Device Incompatibility
ADDITIONAL INFORMATION RECEIVED STATES THAT THE EVENT OCCURRED FOLLOWING INSTRUMENT STERILIZATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19906622 ↗MalfunctionMay 7, 2024Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
ADDITIONAL INFORMATION WAS RECEIVED: THE SPRING ON THE PIN PULLER HAD WORN OUT, SO IT WOULD NOT RELEASE THE PIN ONCE REMOVED FROM BONE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19264087 ↗MalfunctionApr 21, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE PIN PULLER WOULD NOT GRAB PINS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16788265 ↗MalfunctionMar 28, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE PIN PULLER WON¿T GRAB PIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16629275 ↗MalfunctionAug 22, 2022Device-Device Incompatibility; Device-Device Incompatibility
THE PIN PULLER WAS NO LONGER GRIPPING THE PINS. NO ADVERSE AFFECTS ON THE PATIENT. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15266540 ↗MalfunctionDec 10, 2021Break; Break
SPRING ON INSTRUMENT BROKEN DUE TO WEAR. NEW TRAY OPENED TO GET REPLACEMENT INSTRUMENT. 2 MINUTES DELAY TO SURGERY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 12977697 ↗MalfunctionDec 2, 2020Device-Device Incompatibility; Material Twisted/Bent; Mechanical Jam; Device-Device Incompatibility; Material Twisted/Bent; Mechanical Jam
PIN PULLER HAS BECOME SPRUNG AND PINS ARE ABLE TO BE PULLED BUT THE PIN BECOMES STUCK IN THE INSTRUMENT. NOT DETRIMENTAL TO PATIENT JUST INCONVENIENCE TO CUSTOMERS UTILIZING INSTRUMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10936578 ↗MalfunctionOct 25, 2019Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE INSTRUMENT WON'T LOCK PINS IN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9240808 ↗MalfunctionOct 25, 2019Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE INSTRUMENT WON'T LOCK PINS IN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9240807 ↗MalfunctionSep 18, 2019Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE SPRING WILL NOT RELEASED. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9081744 ↗MalfunctionAug 30, 2019Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE SPECIALIST 2 PIN PULLER WILL NOT GRAB PINS. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8951804 ↗MalfunctionFeb 21, 2019Positioning Problem; Positioning Problem
IT WAS REPORTED THAT THE PIN PULLER WOULD NOT PULL THE PINS OUT. IT KEPT SLIPPING OFF. THERE WAS NO LOT NUMBER. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8359407 ↗
FDA clearance and classification
The FDA regulates product code HWB (extractor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the SPECIALIST 96-6515
Has the SPECIALIST 96-6515 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10603295248637), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the SPECIALIST 96-6515?
The FDA's MAUDE database lists 15 adverse event reports for this device, received between Feb 21, 2019 and Aug 6, 2026. In those reports, the most common specific device problems are “Device-Device Incompatibility” (11), “Contamination /Decontamination Problem” (2) and “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SPECIALIST 96-6515?
None of the reports are classified as deaths, 0 as injuries and 15 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SPECIALIST 96-6515 FDA approved?
Devices of this type (product code HWB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).
Who makes the SPECIALIST 96-6515, and is it still sold?
The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand SPECIALIST, model 96-6515). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other EXTRACTOR devices
- ATTUNE 2545-00-138Depuy (Ireland) · EXTRACTOR390 reports
- PINNACLE 2217-50-001Depuy (Ireland) · EXTRACTOR242 reports
- RECLAIM 2975-00-635Depuy (Ireland) · EXTRACTOR156 reports
- ATTUNE 2545-00-060Depuy (Ireland) · EXTRACTOR150 reports
- ESC PINNACLE 2218-00-010Depuy (Ireland) · EXTRACTOR150 reports
- NA 2125-00-600Depuy Orthopaedics, Inc. · EXTRACTOR129 reports
- P.F.C. 86-5226Depuy (Ireland) · EXTRACTOR80 reports
- NA 03.010.000Synthes GmbH · EXTRACTOR69 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page