GLOBAL UNITE 1407-20-101Recalls & reported problems

Depuy (Ireland) · GLOBAL UNITE PLATFORM SHOULDER SYSTEM REVERSE FX EPI CENTER POROCOAT · Class 2 · Implant

All models of GLOBAL UNITE (91)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10603295444749), received between Jul 24, 2019 and Nov 5, 2025 — classified as 5 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (3 reports).

8 other recalls cover devices cleared under the same FDA submission (K170748); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Difficult to Insert3
  • Osseointegration Problem2
  • Loss of Osseointegration1

Patient problems

  • Unspecified Infection4
  • Inadequate Osseointegration3
  • No Clinical Signs, Symptoms or Conditions3
  • No Code Available1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 5, 2025Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT BOTH PRODUCTS COULD NOT BE PROPERLY CONNECTED TO EACH OTHER. THERE WAS A REMAINING GAP OF 2CM. NO APPRECIABLE OP TIME DELAY. NO PATIENT DAMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23478940 ↗
  • InjuryJun 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SUBJECT ID: (B)(6). STUDY NO DOTS. CLINICAL ADVERSE EVENT RECEIVED FOR SUSPICION OF PROSTHETIC INFECTION - LEFT SHOULDER. EVENT IS SERIOUS AND IS CONSIDERED MODERATE. EVENT IS NOT RELATED TO DEVICE OR PROCEDURE. DATE OF IMPLANT: (B)(6) 2024 DATE OF EVENT: (B)(6) 2024 (LEFT SHOULDER) TREATMENT: I&D NO REVISION COMPLETED.

    Patient problems: Unspecified Infection

    MAUDE report 19563226 ↗
  • InjuryJun 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT WAS REVISED DUE TO INFECTED REVERSE TOTAL SHOULDER. ALL IMPLANTS WERE REMOVED AND A SPACER WAS IMPLANTED. THERE WAS NO LOOSENING. DOI: (B)(6) 2023. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT SHOULDER.

    Patient problems: Unspecified Infection

    MAUDE report 19446613 ↗
  • InjuryApr 2, 2024Osseointegration Problem; Osseointegration Problem

    REVISION SHOULDER DELTA MOTION POST HUMERAL FRACTURE DUE TO INFECTION INFECTION AND LOOSENING AT THE STEM HUMERAL. THERE WAS NO SURGICAL DELAY.

    Patient problems: Inadequate Osseointegration; Unspecified Infection

    MAUDE report 19018684 ↗
  • InjuryAug 29, 2023Osseointegration Problem; Osseointegration Problem

    PATIENT WAS REVISED DUE TO LOOSENING OF THE STEM. THE SURGEON REMOVED AND REPLACED HUMERAL STEM AND RETAINED ORIGINAL GLENOID COMPONENTS. DOI: (B)(6) 2020 DOR: (B)(6) 2023 AFFECTED SIDE: RIGHT SHOULDER.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17642526 ↗
  • MalfunctionJan 13, 2023Difficult to Insert; Difficult to Insert

    INTRAOPERATIVELY, DURING INITIAL SHOULDER-TEP IMPLANTATION, IT WAS NOT POSSIBLE TO FIX THE INLAY INTO THE EPIPHYSIS. ANOTHER EPIPHYSIS WAS AVAILABLE AND THE INLAY-EPIPHYSIS-CONSTRUCT WAS IMPLANTED WITHOUT ISSUES. SURGICAL DELAY ABOUT 5 MINUTES. NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16159516 ↗
  • MalfunctionNov 16, 2021Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT LOCKING BOLT IN THE EPI WAS STRIPPED AND SURGEON COULDN¿T FULLY TIGHTEN IT TO THE STEM & REQUESTED A NEW ONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12824161 ↗
  • InjuryAug 20, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO INFECTION. THE HUSBAND OF THE PATIENT THOUGHT THE INFECTION WAS CAUSED BY THE IMPLANT. DOI: (B)(6) 2019; DOI: (B)(6) 2019 (PE CUP); DOR: (B)(6) 2019; RIGHT SHOULDER.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8910562 ↗
  • MalfunctionJul 24, 2019Loss of Osseointegration; Loss of Osseointegration

    SURGEON FROM (B)(6) HOSPITAL HAS REPORTED THAT ANOTHER PATIENT HAS PRESENTED TO HIM WITH A ROTATIONALLY UNSTABLE STEM. PLEASE SEE DETAILS BELOW; ISSUE: ROTATIONALLY UNSTABLE STEM. PRODUCT: GLOBAL REVERSE FRACTURE EPIPHYSIS (1407-20-101), GLOBAL UNITE STEM (TBA), STANDARD DELTA XTEND CUP (TBA) & GLENOID COMPONENTS (TBA) ¿ UNCEMENTED TECHNIQUE USED ¿ NO CEMENT USED DISTALLY. PATIENT: MID TO LATE 50¿S, COMPLIANT, NON-SMOKER. INTRA OPERATIVE: SUCCESSFUL OP, PATIENT PRESENTED AS NON-UNION. WAS FIXED WITH DELTA XTEND USING GLOBAL REVERSE FRACTURE EPI (CENTERED COMPONENT). POST OPERATIVE: COMPLIANT PATIENT, LIGHT REHAB. AFTER 5 MONTHS, REHAB INCREASED. IT THEN SEEMS THE STEM BECAME LOOSE. ACTION: PATIENT AWAITING DATE FOR REVISION SURGERY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 8822166 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K170748. One submission often covers many models, so check the product and lot details in each record.

  • Z-0487-2022 · Dec 7, 2021 · Terminated
    GLOBAL UNITE REV STEM SZ 10 Product Code: 110010600
  • Z-0483-2022 · Dec 7, 2021 · Terminated
    GLOBAL UNITE STD STEM SZ 6 Part Number: 11000610
  • Z-0486-2022 · Dec 7, 2021 · Terminated
    GLOBAL UNITE STD STEM SZ 10 Part Number: 110010100
  • Z-0485-2022 · Dec 7, 2021 · Terminated
    GLOBAL UNITE REV STEM SZ 8 Part Number: 110008600
  • Z-0490-2022 · Dec 7, 2021 · Terminated
    GLOBAL UNITE STD STEM SZ 16 Product Code: 110016100
  • Z-0489-2022 · Dec 7, 2021 · Terminated
    GLOBAL UNITE STD STEM SZ 14 Product Code: 110014100
  • Z-0484-2022 · Dec 7, 2021 · Terminated
    GLOBAL UNITE STD STEM SZ 8 Part Number: 110008100
  • Z-0488-2022 · Dec 7, 2021 · Terminated
    GLOBAL UNITE STD STEM SZ 12 Product Code: 110012100

Other PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED devices

All PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED devices →

Sources and updates