This DePuy Attune tibial drill is used to prepare the knee bone for implant surgery during total knee replacement. The 70 reports mostly describe instruments binding, breaking, or wearing out mid-operation, sometimes leaving metal filings or loose fragments in the bone, and in one case a color ring coming loose. There are no deaths and no injuries. No recalls name this device, and the number of reports is steady at 5 in each of the last two 12-month periods.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
71
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 71 adverse event reports that list this exact device (DI 10603295445647), received between Jan 10, 2018 and Sep 29, 2026 — classified as 71 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (36 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility36
  • Break18
  • Unintended Movement7
  • Naturally Worn5
  • Difficult to Advance3
  • Material Twisted/Bent3
  • Mechanical Jam3
  • Scratched Material3

Patient problems

  • No Clinical Signs, Symptoms or Conditions48
  • No Consequences Or Impact To Patient15
  • Not Applicable5
  • Insufficient Information3
  • No Code Available2

Recent adverse event reports

The 25 most recent of 71 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Device-Device Incompatibility; Unintended Movement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50930584 ↗
  • MalfunctionJul 28, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50056741 ↗
  • MalfunctionJul 7, 2026Device-Device Incompatibility; Scratched Material

    IT WAS REPORTED THAT THE ATTUNE CEMENTED DRILL AND DRILL TOWER BEGAN TO BIND DURING TIBIAL PREP. THE INSTRUMENTS WERE IMMEDIATELY REMOVED AND A NEW TRAY WAS OPENED. INSPECTION REVEALED SCORING ON BOTH INSTRUMENTS. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25737165 ↗
  • MalfunctionApr 3, 2026Break; Device-Device Incompatibility

    IT WAS REPORTED THAT DRILL TOWER AND TIBIAL DRILL DIDN¿T MEET WELL AND CREATED METAL FILINGS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24776143 ↗
  • MalfunctionMar 30, 2026Device-Device Incompatibility

    THE INSTRUMENT IS SOMEHOW BROKEN/DAMAGED FROM THE BEGINNING (YOU HAVE TO FORCE THE DRILL INTO THE DRILL TOWER ¿ VERY STIFF AND DOESN¿T FIT) THIS IS THE INSTRUMENT SET THAT THEY STARTED TO USE LATER BECAUSE OF THE MISSING 4L FEMUR PART.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24717246 ↗
  • MalfunctionDec 12, 2025Break; Break

    IT WAS REPORTED THAT THE COLOR RING OF THE ATTUNE LTL CEM TIBIAL DRILL HAS COME LOOSE AND COMPROMISING STERILITY AND RAISING CONCERNS THAT FOREIGN MATERIAL COULD REMAIN IN THE PATIENT. THE PRODUCT WAS REMOVED FROM CIRCULATION AND PLACED IN QUARANTINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23793118 ↗
  • MalfunctionSep 5, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT THE BEARINGS ON PROXIMAL PART OF DRILL ARE BROKEN. THE RED DRILL STOP WOULD NOT "CLICK" ON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22982509 ↗
  • MalfunctionSep 5, 2025Break; Break

    IT WAS REPORTED THAT METAL SHAVINGS WERE COMING OFF THE TIBIAL REAMER WHEN USED TO REAM THE TIBIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22979777 ↗
  • MalfunctionJul 16, 2025Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE TIBIA PREPARATION DRILL BLOCKER PIECE WAS FALLING OFF AND DID NOT GRAB THE DRILL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22518905 ↗
  • MalfunctionFeb 20, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING TIBIAL PREPARATION, THE TIBIAL TOWER AND DRILL WAS USED, AND THE TOWER'S OPENING HAD SHRUNK/REDUCED RESULTING IN BURRING FROM THE DRILL UPON ACTION THROUGH THE TOWER OPENING TO THE TIBIA. THIS CAUSED METAL TO BE STRIPPED FROM THE DRILL, DAMAGING IT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO DELAY. THE FRAGMENTS THAT WERE GENERATED WERE EASILY REMOVED WITHOUT REQUIRING ADDITIONAL INTERVENTION. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21420391 ↗
  • MalfunctionFeb 6, 2025Break; Material Discolored; Device-Device Incompatibility; Material Deformation; Break; Material Discolored; Device-Device Incompatibility; Material Deformation

    DURING DRILLING OF THE FIXED BEARING TIBIA THE DRILL WAS NOT ABLE TO PASS THROUGH THE DRILL TOWER. THE SURGEON APPLIED PRESSURE BUT THE DRILL WOULD STILL NOT MAKE CONTACT WITH THE TIBIA. WHEN DRILL WAS REMOVED IT WAS REVEALED A BURR WITHIN THE TOWER WAS PREVENTING THE DRILL FROM PASSING THROUGH AND HAD DAMAGED THE DRILL. AN ALTERNATE SET WAS OPENED AND USED. SURGERY WAS DELAYED 2 MINUTES WHILE AN ALTERNATE SET WAS OPENED. PROCEDURE SUCCESSFULLY COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21326841 ↗
  • MalfunctionOct 21, 2024Difficult to Advance; Difficult to Advance

    IT WAS REPORTED THAT PEG BROKE, NO EFFECT ON PATIENT. PEG RETRIEVED. 3/5 AND 6/8 STOPS WERE NOT STAYING IN PLACE. THERE WAS NO SURGICAL DELAY AND NO EFFECT ON SURGERY AND PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20493608 ↗
  • MalfunctionOct 3, 2024Break; Break

    IT WAS REPORTED THAT METAL IS CHIPPING.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20360907 ↗
  • MalfunctionSep 30, 2024Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT BROKE. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20334708 ↗
  • MalfunctionJul 11, 2024Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT ONE OF THE BALL BEARINGS HAS PUSHED INSIDE THE DRILL ITSELF, THIS MEANS THAT IT NO LONGER HOLD THE DRILL COLLAR IN PLACE. THIS DID NOT CAUSE AN ISSUE DURING SURGERY AND NO EFFECT TO THE PATIENT. THE INCIDENT DATE IS UNKNOWN

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19720317 ↗
  • MalfunctionJun 18, 2024Device-Device Incompatibility; Mechanical Jam; Device-Device Incompatibility; Mechanical Jam

    THE BALL BEARING OF THE TIBIAL DRILL SEEM TO HAVE DROPPED INTO THE DRILL ITSELF, THEREFORE IT DOES NOT GRIP THE DRILL STOP PROPERLY. IT CAUSED LESS THAN 1 MINUTE DELAY TO SURGERY AND NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19562584 ↗
  • MalfunctionNov 2, 2023Difficult to Advance; Difficult to Advance

    IT WAS REPORTED THAT THE RED STOPPER NO LONGER CLIPS INTO THE ATTUNE TIBIA DRILL AND NEEDS REPLACEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18056190 ↗
  • MalfunctionMay 31, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE DRILL NO LONGER CONNECTS TO DRILL QUICK CONNECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17032751 ↗
  • MalfunctionMar 29, 2023Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE LOCKING SPLIT IN THE PROXIMAL PART OF THE DRILL IS MISSING. THIS WAS DISCOVERED DURING INSTRUMENT CLEANING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16636430 ↗
  • MalfunctionMar 10, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE DRILL CREATED BURR MARK ON THE DRILL TOWER. DRILL WOULD NOT PASS SMOOTHLY THROUGH DRILL TOWER. NO INJURY OR DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16521069 ↗
  • MalfunctionFeb 8, 2023Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE DRILL AND TOWER NO LONGER FIT PROPERLY TOGETHER. THERE IS A BURR ON THE DRILL THAT PREVENTS IT FROM SEATING. ADDITIONALLY, THE TOWER HAS BECOME TOO NARROW FOR THE DRILL. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16328417 ↗
  • MalfunctionJan 10, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE RED DEPTH STOPPERS DO NOT STICK TO THE TOP OF THE DRILL. WHEN SOMEONE SLIDES THEM UP THE DRILL, THEY JUST FALL RIGHT OFF THE REAMER. SURGICAL DELAY UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16131885 ↗
  • MalfunctionJan 5, 2023Naturally Worn; Naturally Worn

    IT WAS REPORTED THE ATTUNE INSTRUMENTS IS DAMAGED AFTER CLEANING. OCCURRED AFTER CLEANING. NO ADVERSE AFFECTS ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16102046 ↗
  • MalfunctionJan 5, 2023Naturally Worn; Naturally Worn

    IT WAS REPORTED THE ATTUNE INSTRUMENTS IS DAMAGED AFTER CLEANING. OCCURRED AFTER CLEANING. NO ADVERSE AFFECTS ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16102043 ↗
  • MalfunctionDec 15, 2022Scratched Material; Scratched Material

    IT WAS REPORTED THAT THE TIBIAL DRILL CONNECTION DENTED AND UNUSABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15993118 ↗