Skip to content

ATTUNE 2545-11-182Recalls & reported problems

Depuy (Ireland) · ATTUNE KNEE SYSTEM FIXED BEARING CENTER TIBIAL DRILL NON-CEMENTED · Class 1

All models of ATTUNE (1475)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295499220), received between May 22, 2023 and Sep 22, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation3
  • Break1
  • Device-Device Incompatibility1
  • Difficult to Advance1
  • Illegible Information1
  • Mechanical Jam1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 22, 2026Material Deformation; Unintended Movement

    It was reported that during a Total Knee Arthroplasty (TKA) the conical drill had a defect on the top connection portion that locks into the hudson adaptor and it wobbled very badly. After the reporter looked at it the reporter noticed it bent and grabbed another tray to complete the prep portion for the tibia on this surgery.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50849734 ↗
  • MalfunctionAug 26, 2026Material Deformation; Illegible Information

    It was reported that Center FB drill was found to be bent and would fit thru the bushing. They just opened a new tray. There was no delays. No patient was harmed drill was being returned and no pieces came off.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50451153 ↗
  • MalfunctionMay 26, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT NON-CEM FB TIBIAL DRILL AND DRILL BUSHING WAS STUCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25278520 ↗
  • MalfunctionAug 22, 2025Break; Break

    IT WAS REPORTED THAT METAL BURRS ON ATTUNE CEMENTLESS FB TIB DRILL PREVENTED IT FROM GOING DOWN THE GUIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22873415 ↗
  • MalfunctionApr 30, 2025Difficult to Advance; Difficult to Advance

    IT WAS REPORTED THAT THE SCREW BACKED OUT ON DRILL AT THE COLLAR. SLEEVE (STOP) WILL NOT SLIDE OVER DRILL.

    Patient problems: Insufficient Information

    MAUDE report 21935321 ↗
  • MalfunctionMay 22, 2023Material Deformation; Mechanical Jam; Material Deformation; Mechanical Jam

    IT WAS REPORTED THAT THE SURGEON INSERTED REAMER INTO BUSHING TO REAM THE TIBIA. SURGEON BEGAN TO REAM AND THE CENTER DRILL GOT STUCK AND GRABBED INSIDE THE BUSHING. SURGEON BELIEVES THE CENTER DRILL IS BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16974663 ↗