DELTA XTEND 1307-62-138Recalls & reported problems
Depuy (Ireland) · DELTA XTEND Lateralized Glenosphere +2mm Standard 038mm · Class 2 · Implant
All models of DELTA XTEND (194)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 10603295502494), received between Aug 27, 2019 and Apr 10, 2026 — classified as 10 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated7
- Difficult to Insert4
- Adverse Event Without Identified Device or Use Problem3
- Device-Device Incompatibility1
Patient problems
- Joint Dislocation7
- Unspecified Infection3
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions1
- Not Applicable1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
InjuryApr 10, 2026Adverse Event Without Identified Device or Use Problem
REMOVAL OF SHOULDER COMPONENTS DUE TO DEEP INFECTION OF THE LEFT SHOULDER JOINT. DEVICES WERE DISCARDED.
Patient problems: Unspecified Infection
MAUDE report 24839395 ↗InjuryOct 16, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT WAS REVISED DUE TO DISLOCATING SHOULDER. DOI: (B)(6) 2025, DOR: (B)(6) 2025, AFFECTED SIDE: UNKNOWN.
Patient problems: Joint Dislocation
MAUDE report 23310558 ↗InjuryJul 24, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO DISLOCATION. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 22590629 ↗InjuryMar 18, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT UNDERWENT REVISION SURGERY DUE TO DISLOCATION - SOME OSTEOPHYTES BUT THE SURGEON COMMENTED HE SHOULD HAVE LATERALISED MORE AS PER MATERIALISE PLAN. SURGEON LATERALISED THE GLENOSPHERE AN EXTRA 2MM (ORIGINALLY 2MM AND LATERALISED TO 4MM) ALSO REPLACED +3/38MM POLY TO +6 RETENTIVE / 38MM. THERE WAS NOTHING WRONG WITH THE PRODUCTS; THE PATIENT'S ANATOMY OF THE SCAPULAR NECK WAS IMPINGING ON THE ORIGINAL GLENOSPHERE SO BY LATERALISING AND UPSIZING THE CUP, THERE WAS NO IMPINGEMENT. THE PATIENT HAD DISLOCATED 3 TIMES, WITH MUA THE FIRST TWO TIMES AND REVISION SURGERY THE THIRD TIME. THE SHOULDER WAS DISLOCATING DUE TO HAVING OSTEPHYTES PRESENT STILL AS WELL AS NOT BEING LATERAL ENOUGH, AND THE IMPLANTS WERE IMPINGING ON THE NECK OF THE HUMERUS. AFFECTED SIDE: LEFT.
Patient problems: Joint Dislocation
MAUDE report 21631852 ↗InjuryMar 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO SUSPECTED LEFT SHOULDER INFECTION. UNKNOWN SURGICAL DELAY. DOI: (B)(6) 2023. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Unspecified Infection
MAUDE report 21607212 ↗InjuryJun 6, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT DISLOCATED AND WAS NOT ABLE TO BE REDUCED IN THE CLINIC. SHE WAS BROUGHT TO THE OR TO OPEN AND REDUCE. ALL COMPONENTS WERE WELL FIXED. THE GLENOSPHERE WAS INCREASED TO A 42+2 STD AND A 42 + 6 RETENTIVE LINER WAS USED. SURGICAL DELAY WAS UNKNOWN. DOI: (B)(6) 2024; DOR: (B)(6) 2024; AFFECTED SIDE: UNKNOWN SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 19479464 ↗InjuryMay 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INFECTION. EVERYTHING WAS EXPLANTED AND A SPACED WAS PUT IN. DOI: (B)(6) 2023. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT KNEE.
Patient problems: Unspecified Infection
MAUDE report 19384328 ↗MalfunctionNov 30, 2023Difficult to Insert; Difficult to Insert
ADDITIONAL INFORMATION WAS RECEIVED AND STATES THAT THE ADVERSE EFFECT WAS UNKNOWN. IT ALSO STATES THAT THERE WERE NO ALLEGATIONS OF SCREWDRIVER.
Patient problems: Insufficient Information
MAUDE report 18236093 ↗MalfunctionNov 30, 2023Difficult to Insert; Difficult to Insert
ADDITIONAL INFORMATION WAS RECEIVED AND STATES THAT THE ADVERSE EFFECT WAS UNKNOWN. IT ALSO STATES THAT THERE WERE NO ALLEGATIONS TO SCREW DRIVER.
Patient problems: Insufficient Information
MAUDE report 18236092 ↗InjuryNov 4, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT WAS REVISED DUE TO SHOULDER DISLOCATION AFTER SURGERY. DOI: (B)(6) 2022. DOR: (B)(6) 2022. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 15734323 ↗MalfunctionSep 8, 2022Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE GLENOSPHERE WOULD NOT SEAT ONTO METAGLENE. THERE WAS NO LOOSENING. THERE WAS SURGICAL DELAY OF 1 HOUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15380132 ↗InjuryJul 6, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE PATIENT WAS REVISED TO ADDRESS A RECENT DISLOCATION. THE PATIENT HAD A PRIOR REVERSE CONSTRUCT FOR FRACTURE. DOI: (B)(6) 2021. DOR: (B)(6) 2021. RIGHT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 12121702 ↗InjuryMar 23, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT WAS REVISED DUE TO DISLOCATED REVERSE SHOULDER IN HOSPITAL BED. INITIAL SURGERY WAS FOR A 4 PART PROXIMAL HUMERAL FRACTURE. DOI: (B)(6) 2021 DOR: (B)(6) 2021 RIGHT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 11553980 ↗MalfunctionAug 27, 2019Difficult to Insert; Device-Device Incompatibility; Difficult to Insert; Device-Device Incompatibility
IT WAS REPORTED THAT THE 38MM +2 LATERALIZED GLENOSPHERE WOULD NOT SEAT IN THE METAGLENE AFTER MULTIPLE INSERTION ATTEMPTS. SUBSEQUENTLY, A STANDARD 38 GLENOSPHERE WAS INSERTED ON THE FIRST ATTEMPT WITHOUT ISSUE. DOE: (B)(6) 2019, LEFT SHOULDER.
Patient problems: Not Applicable
MAUDE report 8934943 ↗
Other PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED devices
- DJO SURGICAL 509-02-032Encore Medical, L.P. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED364 reports
- DELTA XTEND 130790036Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED171 reports
- DELTA XTEND 130770018Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED159 reports
- TRIGEN 71118200Smith & Nephew, Inc. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED123 reports
- DJO SURGICAL 509-02-432Encore Medical, L.P. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED119 reports
- DELTA XTEND 130738206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED116 reports
- COMPREHENSIVE SHOULDER SYSTEM 113633Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED2 recalls4 reports
- DELTA XTEND 130742206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page