INHANCE 5200-01-036Recalls & reported problems
Depuy (Ireland) · INHANCE SHOULDER SYSTEM SHORT STEM MEDIUM diameter 36mm X 74mm · Class 2 · Implant
All models of INHANCE (202)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10603295538202), received between Apr 20, 2022 and May 12, 2026 — classified as 11 injuries, 1 malfunction.
The most frequently reported specific device problem is “Migration” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem6
- Migration2
- Osseointegration Problem2
- Difficult to Insert1
- Loss of or Failure to Bond1
- Use of Device Problem1
Patient problems
- Bone Fracture(s)4
- Insufficient Information3
- Inadequate Osseointegration2
- Pain2
- Joint Dislocation1
- Joint Laxity1
- Unspecified Tissue Injury1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
InjuryMay 12, 2026Migration
IT WAS REPORTED THAT PATIENT HAD REVISION DUE TO STEM VERSION. DOI: ON (B)(6) 2026, DOR: ON (B)(6) 2026, AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Insufficient Information
MAUDE report 25149196 ↗InjuryApr 27, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT NEED MORE STABILITY DOI: (B)(6) 2025. DOR: (B)(6) 2026. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Joint Laxity
MAUDE report 25001639 ↗InjuryApr 21, 2026Use of Device Problem
IT WAS REPORTED THAT MEDIUM SHORT STEM INHANCE, 36+0 POLY AND MEDIUM +0 SHELL REVISING BECAUSE POLY KEPT SLIPPING OUT AND STEM WAS LEFT PROUD. DOI-UNKNOWN DOR-(B)(6) 2026 AFFECTED SIDE-UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 24945442 ↗InjuryDec 30, 2025Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THAT THE DISTAL TIP OF HUMERAL STEM PERFORATED THE HUMERUS. IT WAS REMOVED AND REPLACED WITH A STD STEM. GLENOSPHERE, CUP AND POLY WERE ALSO REPLACED. THERE WAS A LOOSENING OF HUMERAL STEM AT NON-CEMENTED INTERFACE. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Bone Fracture(s); Inadequate Osseointegration
MAUDE report 23924093 ↗InjuryDec 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT WAS REVISED DUE TO CUFF TEAR. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Unspecified Tissue Injury
MAUDE report 23707221 ↗InjuryJul 8, 2025Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO PAIN. THERE WAS LOOSENING OF HUMERAL STEM. THE LOOSENING INTERFACE WAS NON-CEMENTED. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Inadequate Osseointegration; Pain
MAUDE report 22437394 ↗InjuryApr 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO PERIPROSTHETIC FRACTURE. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Bone Fracture(s)
MAUDE report 21902710 ↗InjuryNov 21, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AFTER THE PATIENT HAD THEIR ORIGINAL SHOULDER DONE, THE SHOULDER BECOME DISLOCATED. ACCORDING TO SURGICAL IMPLANT RECORDS, THE PATIENT WAS IMPLANTED WITH A 40+8 GLENOSPHERE, MEDIUM +8 SHELL, AND A 40 +4 RETENTIVE POLY. WITH THAT CONSTRUCT PRETTY MUCH MAXED OUT, THE SURGEON FELT THE PATIENT'S HUMERAL COMPONENTS WOULD NEED TO BE REVISED AS A WHOLE. THE HUMERAL STEM WAS REMOVED AND THE PATIENT WAS THEN PREPPED FOR A BIOMET HUMERAL COMPONENT. WITH THE BASEPLATE WELL FIXED, THE SURGEON OPTED TO LEAVE IT AND PLACE A 40+8 ECCENTRIC GLENOSPHERE. ONCE THE BIOMET PREP WAS COMPLETE, THE SHOULDER WAS REDUCED AND PUT THROUGH A FULL RANGE OF MOTION AND FOUND TO BE STABLE. DOI: ON (B)(6) 2023, DOR: ON (B)(6) 2024, AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 20750854 ↗MalfunctionApr 22, 2024Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE GLENOSPHERE WOULD NOT ENGAGE OR SEAT TO THE BASEPLATE. ADDITIONALLY, THE HUMERAL SHELL AND SHORT STEMS TAPER WOULD NOT PROPERLY ENGAGE AND WAS UNEVEN. NO SURGICAL TIME WAS ADDED. DOE: (B)(6) 2024 AFFECTED SIDE: UNKNOWN SHOULDER.
Patient problems: Insufficient Information
MAUDE report 19151519 ↗InjuryNov 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO BONE FRACTURE IN HUMERUS. DOI: (B)(6) 2023, DOR: (B)(6) 2023, AFFECTED SIDE: UNKNOWN.
Patient problems: Bone Fracture(s)
MAUDE report 18236740 ↗InjuryNov 20, 2023Loss of or Failure to Bond; Migration; Loss of or Failure to Bond; Migration
IT WAS REPORTED THAT PATIENT WAS REVISED INHNACE REVERSE HUMERAL COMPONENTS. POSSIBLE LOOSENING OF STEM-APPEARED TO HAVE FALLEN INTO VARUS. PATIENT COMPLAINS OF PAIN PER MD. NO OTHER INFORMATION KNOWN. DOI: (B)(6) 2023; DOR: (B)(6) 2023; AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Pain
MAUDE report 18172977 ↗InjuryApr 20, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION COMPLETED DUE TO STEM STICKING OUT OF THE POSTERIOR ASPECT OF THE HUMERUS. SURGEON HAD TO REMOVE HEAD, STEM, AND ADAPTER AND WENT WITH A LONGER UNITE STEM AND HEAD IN ORDER TO BRIDGE THE GAP. SURGEON CHOSE TO KEEP THE EXISTING GLENOID, ONLY REVISED HUMERAL COMPONENTS. DOI: UNKNOWN, DOR: (B)(6) 2022, AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Bone Fracture(s)
MAUDE report 14159574 ↗
Other PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED devices
- DJO SURGICAL 509-02-032Encore Medical, L.P. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED364 reports
- DELTA XTEND 130790036Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED171 reports
- DELTA XTEND 130770018Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED159 reports
- TRIGEN 71118200Smith & Nephew, Inc. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED123 reports
- DJO SURGICAL 509-02-432Encore Medical, L.P. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED119 reports
- DELTA XTEND 130738206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED116 reports
- COMPREHENSIVE SHOULDER SYSTEM 113633Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED2 recalls4 reports
- DELTA XTEND 130742206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page