Skip to content

HALL MICROFREE PRO8200SBRecalls & reported problems

Conmed Corporation · HALL MicroFree Sagittal Saw · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10653405005735), received between Nov 11, 2025 and Nov 11, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 11, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A DISTRIBUTOR REPORTED ON BEHALF OF A CUSTOMER THAT THE PRO8200SB, HALL MICROFREE SAGITTAL SAW DEVICE WAS USED IN AN UNKNOWN PROCEDURE ON (B)(6) 2025, AND ¿WHEN A SAGITTAL BLADE ¿ WAS ATTACHED TO THE MICROFREE SAGITTAL SAW TO PERFORM BONE CUTTING, THE BLADE EASILY CAME OFF.¿. THERE WAS NO INJURY TO THE PATIENT OR USER. THE PROCEDURE WAS COMPLETED WITHOUT THE USE OF AN ALTERNATE DEVICE AND NO REPORTED DELAY. FURTHER ASSESSMENT FOUND "THE BLADE IS NOT BROKEN, BUT IT HAS COME OFF THE HANDPIECE", FELL INTO THE SURGICAL SITE AND WAS RETRIEVED. THERE WAS NO REPORT OF MEDICAL/SURGICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23525833 ↗

FDA clearance and classification

The FDA regulates product code HAB (saw, powered, and accessories) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the HALL MICROFREE PRO8200SB

Has the HALL MICROFREE PRO8200SB been recalled?

No FDA recall record in our data names this exact device identifier (DI 10653405005735), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the HALL MICROFREE PRO8200SB?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Nov 11, 2025. That report lists device problem “Device Dislodged or Dislocated” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the HALL MICROFREE PRO8200SB?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the HALL MICROFREE PRO8200SB FDA approved?

Devices of this type (product code HAB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Conmed Corporation.

Who makes the HALL MICROFREE PRO8200SB, and is it still sold?

The FDA's device database lists Conmed Corporation as the company responsible for it (brand HALL MICROFREE, model PRO8200SB). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SAW, POWERED, AND ACCESSORIES devices

All SAW, POWERED, AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page