Skip to content

PadPro 2001M-CRecalls & reported problems

Conmed Corporation · PADPRO Defibrillation/Pacing/Cardioversion/Monitoring Electrodes, Adult/Child greater than 10 kg Radiotransparent Electrode, Physio-Control QUIK-COMBO (M) 60" (152 cm) · Class 3

All models of PadPro (13)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
1
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10653405042587), received between Jul 2, 2018 and Sep 15, 2025 — classified as 1 death, 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Sparking” (2 reports).

4 other recalls cover devices cleared under the same FDA submission (K014209, P200004); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Sparking2
  • Unable to Obtain Readings2
  • Blocked Connection1
  • Difficult or Delayed Separation1
  • Failure to Conduct1
  • Failure to Deliver Shock/Stimulation1
  • Insufficient Device Problem Information1
  • No Pacing1

Patient problems

  • No Code Available2
  • No Known Impact Or Consequence To Patient2
  • Arrhythmia1
  • Burn(s)1
  • Insufficient Information1
  • No Consequences Or Impact To Patient1
  • Skin Tears1
  • Ventricular Fibrillation1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • DeathSep 15, 2025Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER REPORTED THAT THE DEVICE, 2001M-C, ADULT/CHILD: 10KG. RADIOTRANELECTRODE, PHYSIOCONTROLQUIKCOMBO, WAS BEING USED DURING A RESUSCITATION PROCEDURE ON (B)(6) 2025 WHEN IT WAS REPORTED, ¿DEFIBRILLATOR PADS BURNED/SMOKING. SCALDED PATIENT AND FAILED TO DEFIBRILLATE. REPLACED PADS AND FUNCTIONED CORRECTLY. 2 LOTS USED, UNSURE AS TO WHICH LOT HAD ISSUE. PATIENT STATUS DECEASED.¿. FURTHER ASSESSMENT FOUND THAT THIS WAS A RESUSCITATION EVENT. THE BURN WAS DISCOVERED IMMEDIATELY AFTER PULLING BACK THE GOWN DURING THE EVENT. THE BURN WAS NOT EVALUATED AS THE PATIENT WAS DECEASED. THE BURN OCCURRED ON THE CHEST/UPPER ABDOMEN. THE REPORTER HAS STATED THAT AFTER DEFIBRILLATION "WE HAD RETURN OF SPONTANEOUS CIRCULATION BRIEFLY." AS FOR A CONNECTION BETWEEN DEVICE AND PATIENT DEATH, REPORTER STATED THEY ARE "UNSURE" AND ADDED, "WE WERE UNABLE TO RESUSCITATE THIS PATIENT IN A TIMELY MANNER. THE PATIENT WAS IN A SHOCKABLE ARRHYTHMIA, HENCE WHY DEFIBRILLATION WAS USED. IF WE WOULD HAVE BEEN ABLE TO SHOCK THE PATIENT OUT OF THE RHYTHM SOONER, IT MAY HAVE CHANGED THE OUTCOME OF THE RESUSCITATION EVENT, BUT THAT IS SPECULATIVE, AND I CANNOT SPECIFICALLY SAY THAT THIS WOULD HAVE CHANGED THE

    Patient problems: Arrhythmia; Burn(s)

    MAUDE report 23053740 ↗
  • InjuryJun 14, 2022Difficult or Delayed Separation; Difficult or Delayed Separation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE 2001M-C, PADPRO: AD; RTRANS EL/LW, 12X7, WAS USED IN AN UNKNOWN TYPE OF PROCEDURE ON (B)(6) 2022 AND DURING POST-OP IT WAS REPORTED ¿STAFF FROM CAMPUS CALLED REPORTING THAT THEY HAVE EXPERIENCED A PATIENT INJURY DUE TO REMOVAL OF DEFIBRILLATION PAD 2001M-C. PATIENT'S SKIN WAS REMOVED.¿ THE DISCOVERY WAS MADE POST-OP. THE PAD WAS PLACED ON THE POSTERIOR BACK AND NO PREP OF THE AREA WAS CONDUCTED ON THE PATIENT PRIOR TO APPLICATION. THE DEVICE WAS REMOVED WITH A TWO-HAND SPLINTING TECHNIQUE. A SKIN TEAR WAS VISIBLE AND APPARENT AFTER REMOVAL OF PAD. PRE AND POST SKIN ASSESSMENTS WERE DONE. THERE WAS NO MEDICAL INTERVENTION OR HOSPITALIZATION FOR THE PATIENT. CURRENT STATUS OF THE PATIENT IS "DISCHARGED." THIS REPORT IS BEING RAISED ON THE BASIS OF INJURY DUE TO THE REPORT OF SKIN TEAR WITH REMOVAL OF THE PAD.

    Patient problems: Skin Tears

    MAUDE report 14684466 ↗
  • MalfunctionMar 16, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    THE CUSTOMER REPORTED THAT THE DEVICE, 2001M-C, PADPRO:AD;RTRANS EL/LW, 12X7, WAS BEING USED DURING A CARDIOVERSION ON (B)(6) 2022 WHEN IT WAS REPORTED ¿PATIENT EXPERIENCING REDNESS IN SHAPE AND LOCATION OF THE DEFIB PADS AFTER BEDSIDE CARDIOVERSION. PATIENT STATES THE REDNESS FELT LIKE A SUNBURN.¿. THE REPORTER ADVISED THAT THERE WAS NO FORMAL DIAGNOSIS OF THE ¿BURN¿ AND THERE WAS NO REPORT OF MALFUNCTION OF THE DEVICE. THE PROCEDURE WAS REPORTED AS HAVING BEEN COMPLETED. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. THE PATIENT WAS DISCHARGED AND ADVISED TO CONTACT THE NURSE MANAGER IF THE CONDITION WORSENED OR DID NOT SUBSIDE IN 1-2 DAYS. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: Insufficient Information

    MAUDE report 13789000 ↗
  • MalfunctionOct 14, 2019Unable to Obtain Readings; Unable to Obtain Readings

    THE CONMED REPRESENTATIVE REPORTED ON BEHALF OF THE FACILITY THAT THE 2001M-C, PAD PRO DEFIB PADS WERE USED IN THE CATHERIZATION LAB DURING A PROCEDURE ON (B)(6) 2019. THEY HAD TO USE MULTIPLE PADS FROM A COUPLE OF DIFFERENT LOTS AND EXPERIENCED NO EKG READINGS. THE PADS WERE REPLACED WITH COMPETITOR PADS AND TRACING WAS PERFORMED AS NORMAL. THE CONMEDS REPRESENTATIVE WAS NOTIFIED AT THE END OF THE DAY AND IMMEDIATELY RESPONDED WITH NO RETURN COMMUNICATION. HE STOPPED ONSITE THE NEXT MORNING TO GATHER SAMPLES, AND INTEL OF SITUATION AND WHAT IS GOING TO HAPPEN. THE CURRENT STATUS OF THE PATIENT IS STABLE. THIS REPORT IS BEING RAISED BASED ON DEVICE MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9187311 ↗
  • MalfunctionOct 14, 2019Unable to Obtain Readings; Unable to Obtain Readings

    THE CONMED REPRESENTATIVE REPORTED ON BEHALF OF THE FACILITY THAT THE 2001M-C, PAD PRO DEFIB PADS WERE USED IN THE CATHERIZATION LAB DURING A PROCEDURE ON (B)(6) 2019. THEY HAD TO USE MULTIPLE PADS FROM A COUPLE OF DIFFERENT LOTS AND EXPERIENCED NO EKG READINGS. THE PADS WERE REPLACED WITH COMPETITOR PADS AND TRACING WAS PERFORMED AS NORMAL. THE CONMEDS REPRESENTATIVE WAS NOTIFIED AT THE END OF THE DAY AND IMMEDIATELY RESPONDED WITH NO RETURN COMMUNICATION. HE STOPPED ONSITE THE NEXT MORNING TO GATHER SAMPLES, AND INTEL OF SITUATION AND WHAT IS GOING TO HAPPEN. THE CURRENT STATUS OF THE PATIENT IS STABLE. THIS REPORT IS BEING RAISED BASED ON DEVICE MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9187270 ↗
  • MalfunctionSep 30, 2019Therapeutic or Diagnostic Output Failure; No Pacing; No Visual Prompts/Feedback; Therapeutic or Diagnostic Output Failure; No Pacing; No Visual Prompts/Feedback

    THE CUSTOMER REPORTED THE 2001M-C, PAD PRO WAS DEFECTIVE AND NOT WORKING DURING USE. FURTHER CLARIFICATION WAS RECEIVED. IT WAS STATED THAT DURING A CODE, THE STAFF HOOKED UP THE PADS AND MACHINE. THEY GOT NO READING. THEY TRIED ANOTHER MACHINE, STILL HAD NO READINGS. THEY CHANGED THE PADS USING ANOTHER OF THE SAME PAD BUT DIFFERENT LOT NUMBER AND IT WORKED. THERE WAS NO PATIENT INJURY OR IMPACT REPORTED. THIS REPORT IS BEING RAISED BASED ON DEVICE MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9136513 ↗
  • InjurySep 5, 2018Failure to Conduct; Blocked Connection; Failure to Conduct; Blocked Connection

    THE CUSTOMER REPORTED AN ISSUE OCCURRING WITH 2001M-C, PADPRO, P/C DC, TRANSPARENT-L/WS DURING A REDO MITRAL VALVE REPLACEMENT ON (B)(6) 2018. THE PROCEDURE INVOLVED A (B)(6) MALE WITH SEVERE MITRAL VALVE STENOSIS WHO WAS STATUS POST AORTIC VALVE REPLACEMENT IN 2005. IN PREPARATION FOR THE START OF THE PROCEDURE, THE ABOVE NOTED DEFIBRILLATION ELECTRODES WERE PLACED ON THE PATIENT. NURSING STAFF RESPONSIBLE FOR THE PLACEMENT OF THESE PADS INDICATED THAT THERE WAS INITIALLY GOOD ADHERENCE OF THE PAD TO THE PATIENT'S SKIN. DURING REWARMING, THE PATIENT WAS NOTED TO GO INTO VENTRICULAR FIBRILLATION. TWO ATTEMPTS WERE MADE TO DEFIBRILLATE THE PATIENT. HOWEVER, THERE WAS NO CARDIOVERSION NOTED. AN ERROR MESSAGE OF "NO CONNECTION" (OR SIMILAR) WAS NOTED ON THE MONITOR. THE INITIAL SET OF PADS WERE EXCHANGED FOR A SECOND SET OF THE SAME BRAND AND SAME LOT NUMBER. THE CABLE BETWEEN THE DEFIBRILLATOR AND THE PADS WAS EXCHANGED BUT THE ERROR MESSAGE PERSISTED. A SECOND DEFIBRILLATOR WAS BROUGHT INTO THE ROOM AND CONNECTED TO THE CABLES, HOWEVER THE ERROR MESSAGE PERSISTED AND THE STAFF ARE UNABLE TO CARDIOVERT THE PATIENT. ANOTHER STAFF MEMBER WAS ABLE TO LOCATE A THIRD SET OF DEFIBRILLATOR

    Patient problems: Ventricular Fibrillation

    MAUDE report 7847971 ↗
  • MalfunctionJul 2, 2018Sparking; Sparking

    THE CONMED REPRESENTATIVE REPORTED ON BEHALF OF THE USER FACILITY THAT DURING A CARDIOVERSION WITH 2001M-C PAD, A SPARK WAS WITNESSED WHERE THE WIRE ENTERED THE PAD. THE PAD INVOLVED WAS THE ANTERIOR PAD PLACED ON THE PATIENT'S MIDSTERNUM. THE CARDIOVERSION WAS SUCCESSFULLY COMPLETED WITH NO ADDITIONAL INTERVENTIONS REQUIRED. NO PATIENT INJURY OR SURGICAL DELAYS WAS REPORTED. THIS REPORT IS BEING RAISED ON THE BASIS OF A REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY WITH RECURRENCE.

    Patient problems: No Code Available

    MAUDE report 7655064 ↗
  • MalfunctionJul 2, 2018Sparking; Sparking

    THE CONMED REPRESENTATIVE REPORTED ON BEHALF OF THE USER FACILITY THAT DURING A CARDIOVERSION WITH 2001M-C PAD, A SPARK WAS WITNESSED WHERE THE WIRE ENTERED THE PAD. THE PAD INVOLVED WAS THE ANTERIOR PAD PLACED ON THE PATIENT'S MIDSTERNUM. UPON REMOVING THE PAD, A SMALL PINK SPOT WAS OBSERVED. THE CARDIOVERSION WAS SUCCESSFULLY COMPLETED WITH NO ADDITIONAL INTERVENTIONS REQUIRED. NO PATIENT INJURY OR SURGICAL DELAYS WAS REPORTED. THIS REPORT IS BEING RAISED ON THE BASIS OF A REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY WITH RECURRENCE.

    Patient problems: No Code Available

    MAUDE report 7654867 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P200004 on Sep 26, 2021. Submitted by Conmed Corporation as “ConMed PadPro Multifunction Electrodes, ConMed PadPro Multifunction Electrode Adapters”. The FDA review took 604 days. The FDA has since approved 1 supplement (changes to the device or its labeling), most recently on Jun 16, 2022. FDA record ↗

Cleared through Special 510(k) K014209 on Jan 18, 2002 (decision: Substantially Equivalent). Submitted by Padpro, LLC as “Padpro Model 2001, 2001-S, 2001-C, 2001-Eps”. The FDA review took 28 days. FDA record ↗

The FDA regulates product code MKJ (automated external defibrillators (non-wearable)) under 21 CFR 870.5310 as a Class III device. 21 CFR 870.5310 ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

FDA definition of this device type: This device is a non-wearable prescription use only aed. These are devices that include automated external defibrillation. Automated external defibrillators use external pad-type electrodes to sense, detect, classify and treat (with an electrical shock) ventricular fibrillation. These devices are intended to be used on suspected victims of sudden cardiac arrest. A person in cardiac arrest is unresponsive and is not breathing normally. The device can be sold with prescription only.

Recalls under the same FDA submission

These recalls involve devices cleared under K014209, P200004. One submission often covers many models, so check the product and lot details in each record.

  • Z-0454-2015 · Nov 6, 2014 · Terminated
    ADULT Radiotransparent Electrode, REF/Catalog Number 2001H, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELECTRODES PLUG STYLE CONNECTOR Intended for use during defibrillat
  • Z-0456-2015 · Nov 6, 2014 · Terminated
    ADULT Radiotransparent Electrode, REF/Catalog Number 2001H-PC, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELECTRODES PLUG STYLE CONNECTOR
  • Z-0455-2015 · Nov 6, 2014 · Terminated
    ADULT Radiotransparent Electrode, REF/Catalog Number 2001H-C, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELECTRODES PLUG STYLE CONNECTOR
  • Z-1317-2013 · May 3, 2013 · Terminated
    PadPro(R) Adult Radiotransparent Multifunction Electrodes, REF/Catalog No. 2001M-C. The PadPro(R) Adult Radiotransparent Multifunction Electrodes with Carbon Leadwires and Medtro

Common questions about the PadPro 2001M-C

Has the PadPro 2001M-C been recalled?

No FDA recall record in our data names this exact device identifier (DI 10653405042587), based on the FDA recall file dated Nov 27, 2026. However, 4 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the PadPro 2001M-C?

The FDA's MAUDE database lists 9 adverse event reports for this device, received between Jul 2, 2018 and Sep 15, 2025. In those reports, the most common specific device problems are “Sparking” (2), “Unable to Obtain Readings” (2) and “Blocked Connection” (1); the most common patient problems are “Arrhythmia” (1), “Burn(s)” (1) and “Skin Tears” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PadPro 2001M-C?

1 report is classified as a death, 2 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PadPro 2001M-C FDA approved?

It was cleared through Special 510(k) K014209 on Jan 18, 2002; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the PadPro 2001M-C, and is it still sold?

The FDA's device database lists Conmed Corporation as the company responsible for it (brand PadPro, model 2001M-C). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Automated External Defibrillators (Non-Wearable) devices

All Automated External Defibrillators (Non-Wearable) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page