MERSILENE RS22Recalls & reported problems
Ethicon, LLC · White Woven Polyester Fiber, Nonabsorbable Surgical Ligature · Class 2 · Implant
All models of MERSILENE (12)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon MERSILENE RS22 nonabsorbable polyester surgical suture is used for stitching tissue, including cervical cerclage procedures. The 47 reports mostly describe the suture coming apart or pulling away from the needle mid-surgery, sometimes breaking, with surgeons switching to another suture to finish; a few mention infection or tissue injury. There are 6 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 49
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 10705031049802), received between Mar 12, 2018 and Sep 9, 2026 — classified as 6 injuries, 43 malfunctions.
The most frequently reported specific device problem is “Material Separation” (25 reports).
8 other recalls cover devices cleared under the same FDA submission (K946173); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation25
- Break14
- Adverse Event Without Identified Device or Use Problem6
- Detachment Of Device Component2
- Component Misassembled1
- Delivered as Unsterile Product1
- Packaging Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions34
- No Consequences Or Impact To Patient8
- Bacterial Infection2
- Insufficient Information1
- Laceration(s)1
- Menstrual Irregularities1
- Miscarriage1
- Pain1
Recent adverse event reports
The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 9, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50699100 ↗MalfunctionSep 9, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50699097 ↗MalfunctionDec 30, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CERVICAL CERCLAGE PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23922376 ↗MalfunctionDec 24, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CERCLAGE PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23888293 ↗MalfunctionSep 19, 2025Component Misassembled; Component Misassembled
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. PRIOR TO USING THE SUTURE IT WAS NOTED THAT THE NEEDLE TIP WAS INCORRECT NEEDLE SHAPE. ACTION: DISCARDED THE SUTURE AND OPEN ANOTHER SUTURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23094968 ↗MalfunctionSep 4, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CERVICAL CERCLAGE PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE PATIENT UNDERWENT SURGERY AT 14:42 DUE TO PREGNANCY COMPLICATED WITH CERVICAL DYSFUNCTION. DURING THE SURGERY, THE DOCTOR USED A SUTURE TAPE TO SUTURE THE CERVIX, THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED AND COULD NOT CONTINUE SUTURING. THE DOCTOR PASSED THE NEEDLE THROUGH THE SUTURE TAPE, CONTINUED SUTURING, AND COMPLETED THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22963080 ↗MalfunctionAug 21, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CERVICAL CERCLAGE UNDER LAPAROSCOPY ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED WHEN SEWING DURING THE SURGERY. THE NEEDLE DID NOT FALL INTO THE PATIENT. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22860316 ↗MalfunctionApr 17, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A A SURGICAL PROCEDURE TO INSERT A CERCLAGE ON (B)(6) 2025 AND SUTURE WAS USED. THE DESCRIPTION OF THE COMPLAINT WAS THAT THE SUTURE MATERIAL CAME AWAY FROM THE NEEDLE DURING SURGERY. THE EVENT WAS MANAGED WITHOUT ANY CONSEQUENCES TO THE PATIENT AS THE SURGEON WAS ABLE TO USE THE OTHER END OF THE SUTURE TO CLOSE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21862019 ↗MalfunctionMar 7, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21548503 ↗MalfunctionFeb 20, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC UTERINE ISTHMUS CERCLAGE SURGERY ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE PATIENT'S VAGINAL CERVICAL CERCLAGE FAILED, AND SUTURE WAS USED DURING THE LAPAROSCOPIC UTERINE ISTHMUS CERCLAGE SURGERY. DURING THE SURGERY, WHEN THE NEEDLE HAD COMPLETELY PENETRATED THE TISSUE OF THE UTERINE ISTHMUS, THE SUTURE SLIGHTLY EXPOSED THE TISSUE. THE DOCTOR CONTINUED TO PULL THE SUTURE OUT OF THE TISSUE BY PULLING THE NEEDLE BODY, THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED. IMMEDIATELY, A NEEDLE HOLDER WAS USED TO STRAIGHTEN THE NEEDLE BODY AND REMOVE IT. THE SUTURE THAT LEAKED OUT OF THE ORGANIZATION WAS RELATIVELY SHORT, AND THE DOCTOR CONTINUED TO PULL THE EXPOSED SHORT HEAD TO PULL OUT THE SUTURE AND TIE IT. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN THIS CASE IS CONSIDERED FORTUNATE. IF THE SUTURE SOMETIMES DETACHES BEFORE IT PENETRATES THE TISSUE, IT MAY REQUIRE A SECOND PUNCTURE FOR THE PATIENT, RESULTING IN INCREASED BLEEDING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 21427744 ↗MalfunctionJan 17, 2025Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN LAPAROSCOPE PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. WHEN THE OPERATOR WAS USING IT, THE FIXATION BECAME LOOSE DUE TO THE THREAD LOOSENING. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21179312 ↗MalfunctionNov 20, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A GYNECOLOGICAL PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN THE SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE NEEDLE DID NOT FELL INTO THE PATIENT. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20737081 ↗MalfunctionJul 26, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19838746 ↗MalfunctionMay 15, 2024Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A VAGINAL CERCLAGE ON AN UNKNOWN DATE IN 2024 AND A SUTURE TAPE WAS USED. THE SUTURE KEPT PULLING OFF THE SUTURE. THE REST OF THE CASE WAS COMPLETED WITHOUT PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19322535 ↗MalfunctionApr 15, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNKNOWN PROCEDURE THAT THREE NEEDLES POPPED OFF WHILE SUTURING. ANOTHER LIKE SUTURE WAS USED TO CONTINUE WITH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19111325 ↗MalfunctionApr 15, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNKNOWN PROCEDURE THAT THREE NEEDLES POPPED OFF WHILE SUTURING. ANOTHER LIKE SUTURE WAS USED TO CONTINUE WITH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19106834 ↗MalfunctionApr 15, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNKNOWN PROCEDURE THAT THREE NEEDLES POPPED OFF WHILE SUTURING. ANOTHER LIKE SUTURE WAS USED TO CONTINUE WITH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19106830 ↗MalfunctionFeb 16, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CERVICAL CERCLAGE PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. BEFORE USE ON THE PATIENT, IT WAS REPORTED THAT THE NEEDLE HAD BEEN FOUND BROKEN BEFORE USING ON PATIENT . VISUAL INSPECTION AND METALLURGICAL ANALYSIS WERE PERFORMED ON THE RETURNED PRODUCT. A FRACTURE WAS OBSERVED AT THE SUTURE ATTACHMENT OF THE NEEDLE. THE NEEDLE WAS RECEIVED IN TWO PIECES, ONLY ONE OF THE MATING FRACTURE SURFACES WAS EXAMINED FOR THIS EVALUATION. THE NEEDLE WAS NOTED WITH MARKS THAT APPEARS TO BE BY SURGICAL INSTRUMENT. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18722227 ↗MalfunctionDec 19, 2023Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND TAPE WAS USED. THE NEEDLE WAS FOUND BROKEN WHEN IT WAS OPENED TO PREPARE FOR THE SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE LOT IS UNAVAILABLE. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. UPON RECEIPT AND ANALYSIS, IN WAS OBSERVED A NEEDLE WAS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18357013 ↗MalfunctionSep 7, 2023Delivered as Unsterile Product; Packaging Problem; Delivered as Unsterile Product; Packaging Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. BEFORE USE ON THE PATIENT, IT WAS REPORTED THAT THE PRIMARY PACKAGE OF THE SUTURE WAS FOUND DAMAGED PRIOR TO USE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17699047 ↗MalfunctionMay 24, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6)2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE PATIENT UNDERWENT SURGERY DUE TO PREGNANCY COMBINED WITH CERVICAL INCOMPETENCE. DURING THE SURGERY, THE NEEDLE BROKE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16998728 ↗MalfunctionJan 3, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE BROKE WHEN SEWING IN AN UNKNOWN SURGERY. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16092053 ↗MalfunctionJan 3, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE BROKE WHEN SEWING IN AN UNKNOWN SURGERY. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16092052 ↗MalfunctionDec 23, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A WOUND CLOSURE PROCEDURE ON (B)(6) 2022 AND SUTURE TAPE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16042781 ↗MalfunctionDec 23, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A WOUND CLOSURE PROCEDURE ON (B)(6) 2022 AND SUTURE TAPE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16041754 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K946173. One submission often covers many models, so check the product and lot details in each record.
- Z-1516-2025 · Mar 7, 2025 · Open, ClassifiedETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH
- Z-1097-2025 · Dec 20, 2024 · Open, ClassifiedPERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovas
- Z-2486-2020 · Jun 10, 2020 · Open, ClassifiedPERMA-HAND/MERSILK SILK BLU and VIRGIN SILK BLU, Product Codes EH7457G EH7983G FV768 FV769 FV7771 U7058 U7059 W1782 W1784 W1819 W1820 W818 W819 W870
- Z-1842-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX Product Code: X865W Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue appr
- Z-1841-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL Polyester Suture-ETHIBOND Green 75cm USP1 Single Armed CT-1 Product Code: X425H Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue app
- Z-0747-2019 · Dec 18, 2018 · TerminatedPERMA-HAND Silk Suture Size 5, Black 30" (75cm) Double Armed LR Needles, Product Code 485T Product Usage: PERMA-HAND* suture is indicated for use in general soft tissue approxima
- Z-1839-2015 · May 8, 2015 · TerminatedEthicon Fast Absorbing Plain Gut - MultiPass Needles, Sterile, Absorbable Surgical Suture, 5-0 (1.5 metric), 18" (45 cm), Conventional Cutting, Lot #JD6163, Exp: July 2019. Fast ab
- Z-0492-2015 · Oct 17, 2014 · TerminatedEthicon PERMA-HAND Silk siliconized black braided 10 strands per packet Ethicon LLC- Made in USA.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices
- FAST-FIX 72202468Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 death6,626 reports
- FAST-FIX 72201491Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE863 reports
- FAST-FIX 72202469Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE561 reports
- FAST-FIX 72205676Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE439 reports
- Arthrex® AR-4501Arthrex, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE403 reports
- FAST-FIX 72205324Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE383 reports
- LUPINE ORTHOCORD 210712Medos International Sàrl · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE357 reports
- FAST-FIX 72201494Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE249 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page