Ethicon, LLC · Green Braided Polyester, Nonabsorbable Surgical Suture · Class 2 · Implant

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10705031057692), received between Sep 28, 2021 and Dec 11, 2024 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Material Separation” (6 reports).

8 other recalls cover devices cleared under the same FDA submission (K946173); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 3 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation6
  • Device Contaminated During Manufacture or Shipping1
  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 11, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIOVASCULAR ARCH REPLACEMENT PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING CONTINUOUS SUTURING THE SUTURE WAS DETACHED FROM THE NEEDLE. IT WAS NOT A CONTROL RELEASE NEEDLE. THE SCRUB NURSE GRASPED THE NEEDLE WITH A NEEDLE HOLDER AND HANDED IT TO THE DOCTOR AND THE SUTURE WAS PULLED OUT WHEN THE NEEDLE WAS REMOVED BY GRASPING THE NEEDLE NEAR THE TIP OF THE NEEDLE AND PLACING THE SUTURE OVER THE LARGE ARTERY WITH CANNULATION POUCH SUTURE WITHOUT CHANGING THE NEEDLE. A NEW PACKAGE WAS USED AND THIS SAME EVENT OCCURRED TWICE IN A ROW. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20903869 ↗
  • MalfunctionDec 11, 2024Material Frayed; Material Frayed

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE HAD BEEN FRAYED WHEN THE SUTURE WAS TIED. IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20903617 ↗
  • MalfunctionDec 11, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIOVASCULAR ARCH REPLACEMENT PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING CONTINUOUS SUTURING THE SUTURE WAS DETACHED FROM THE NEEDLE. IT WAS NOT A CONTROL RELEASE NEEDLE. THE SCRUB NURSE GRASPED THE NEEDLE WITH A NEEDLE HOLDER AND HANDED IT TO THE DOCTOR AND THE SUTURE WAS PULLED OUT WHEN THE NEEDLE WAS REMOVED BY GRASPING THE NEEDLE NEAR THE TIP OF THE NEEDLE AND PLACING THE SUTURE OVER THE LARGE ARTERY WITH CANNULATION POUCH SUTURE WITHOUT CHANGING THE NEEDLE. A NEW PACKAGE WAS USED AND THIS SAME EVENT OCCURRED TWICE IN A ROW. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20903340 ↗
  • MalfunctionJul 8, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A MITRAL VALVE REPAIR ON (B)(6)2024 AND SUTURE WAS USED. DURING A MITRAL VALVULOPLASTY: MVP, DURING CONTINUOUS SUTURING FOR THE MITRAL VALVE FORMATION, THE NEEDLE AND SUTURE HAD BEEN DETACHED EASILY. IT OCCURRED THAT 2 NEEDLE-SUTURES IN A ROW. NO PIECES FELL INTO THE PATIENT BECAUSE THE SUTURE WAS DETACHED DURING GRASPING THE DISCONNECTED NEEDLE. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19691779 ↗
  • MalfunctionMar 27, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE SURGERY, THE NEEDLE WAS DETACHED FROM THE SUTURE DURING CONTINUOUS SUTURING. THIS ISSUE OCCURRED 3 TIMES IN A ROW DURING A SINGLE CASE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18985529 ↗
  • MalfunctionMar 27, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE SURGERY, THE NEEDLE WAS DETACHED FROM THE SUTURE DURING CONTINUOUS SUTURING. THIS ISSUE OCCURRED 3 TIMES IN A ROW DURING A SINGLE CASE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18985527 ↗
  • MalfunctionMar 27, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE SURGERY, THE NEEDLE WAS DETACHED FROM THE SUTURE DURING CONTINUOUS SUTURING. THIS ISSUE OCCURRED 3 TIMES IN A ROW DURING A SINGLE CASE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18985526 ↗
  • MalfunctionSep 28, 2021Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS TO BE USED. DURING THE PROCEDURE, A HAIR WAS NOTED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12542519 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K946173. One submission often covers many models, so check the product and lot details in each record.

  • Z-1516-2025 · Mar 7, 2025 · Open, Classified
    ETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH
  • Z-1097-2025 · Dec 20, 2024 · Open, Classified
    PERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovas
  • Z-2486-2020 · Jun 10, 2020 · Open, Classified
    PERMA-HAND/MERSILK SILK BLU and VIRGIN SILK BLU, Product Codes EH7457G EH7983G FV768 FV769 FV7771 U7058 U7059 W1782 W1784 W1819 W1820 W818 W819 W870
  • Z-1842-2019 · Apr 19, 2019 · Terminated
    ETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX Product Code: X865W Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue appr
  • Z-1841-2019 · Apr 19, 2019 · Terminated
    ETHIBOND EXCEL Polyester Suture-ETHIBOND Green 75cm USP1 Single Armed CT-1 Product Code: X425H Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue app
  • Z-0747-2019 · Dec 18, 2018 · Terminated
    PERMA-HAND Silk Suture Size 5, Black 30" (75cm) Double Armed LR Needles, Product Code 485T Product Usage: PERMA-HAND* suture is indicated for use in general soft tissue approxima
  • Z-1839-2015 · May 8, 2015 · Terminated
    Ethicon Fast Absorbing Plain Gut - MultiPass Needles, Sterile, Absorbable Surgical Suture, 5-0 (1.5 metric), 18" (45 cm), Conventional Cutting, Lot #JD6163, Exp: July 2019. Fast ab
  • Z-0492-2015 · Oct 17, 2014 · Terminated
    Ethicon PERMA-HAND Silk siliconized black braided 10 strands per packet Ethicon LLC- Made in USA.

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices →