PDS Z493GRecalls & reported problems

Ethicon, LLC · Undyed Monofilament (Polydioxanone), Absorbable Synthetic Surgical Suture · Class 2 · Implant

All models of PDS (224)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Ethicon PDS Z493G is an absorbable surgical suture (polydioxanone monofilament) used to close tissue during surgery. The 54 reports mostly describe the thread breaking or separating during use, along with twisted or bent material and unreadable package expiration dates. Reports mention no deaths and 2 injuries. One open recall began in August 2023 after testing of returned product found some sutures in that lot did not meet the tensile strength requirement.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
56
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 56 adverse event reports that list this exact device (DI 10705031061156), received between Nov 29, 2017 and Sep 30, 2026 — classified as 2 injuries, 54 malfunctions.

The most frequently reported specific device problem is “Break” (41 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 5 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 10, 2023Open, Classified

    Internal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength requirement.

    Z-2617-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break41
  • Material Separation11
  • Adverse Event Without Identified Device or Use Problem2
  • Device Markings/Labelling Problem2
  • Material Twisted/Bent2
  • Misassembled1

Patient problems

  • No Clinical Signs, Symptoms or Conditions34
  • No Consequences Or Impact To Patient20
  • Anaphylactic Shock1
  • Insufficient Information1
  • Local Reaction1

Recent adverse event reports

The 25 most recent of 56 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947876 ↗
  • MalfunctionAug 21, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50374459 ↗
  • MalfunctionMay 5, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, THE SUTURE WAS BROKEN WHEN TIRED TO USE IT. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. WHEN WE SEND THE SAMPLE TO YOU, WE WILL LET YOU KNOW ITS RETURN DATE AND TRACKING NUMBER. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25085953 ↗
  • MalfunctionJun 27, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE, AND SUTURE WAS USED. THE STERILIZATION EXPIRATION DATE OF THE PACKAGE WAS CROPPED OUT. THE STERILIZATION EXPIRATION DATE HAS BEEN PRINTED OUT OF THE DESIGNATED FIELD AND THE CORRECT INFORMATION COULD NOT BE READ. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19624443 ↗
  • MalfunctionJun 27, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE, AND SUTURE WAS USED. THE STERILIZATION EXPIRATION DATE OF THE PACKAGE WAS CROPPED OUT. THE STERILIZATION EXPIRATION DATE HAS BEEN PRINTED OUT OF THE DESIGNATED FIELD AND THE CORRECT INFORMATION COULD NOT BE READ. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19624442 ↗
  • MalfunctionDec 26, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, 2 DEVICES WITH THE SAME ISSUE. NEEDLE SEPARATED FROM SUTURE DURING KNOTTING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18392420 ↗
  • MalfunctionDec 26, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, 2 DEVICES WITH THE SAME ISSUE. NEEDLE SEPARATED FROM SUTURE DURING KNOTTING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18392416 ↗
  • MalfunctionDec 19, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PLASTIC SURGERY ON 21/11/2023 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT THE SUTURE DETERIORATED OR BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18356195 ↗
  • MalfunctionDec 19, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PLASTIC SURGERY ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT THE SUTURE DETERIORATED OR BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18356190 ↗
  • MalfunctionDec 19, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PLASTIC SURGERY ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT THE SUTURE DETERIORATED OR BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18356186 ↗
  • MalfunctionNov 15, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A UNKNOWN DERMATOLOGIC PROCEDURE ON (B)(6)2023 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, WHEN THE SURGEON PULLED THE SUTURE OF PDS5-0 A LITTLE BIT DURING SUTURE, THE SUTURE WAS BROKEN. IT IS UNKNOWN IF THE PRODUCT NUMBER - Z493G ¿ IS THE CORRECT PRODUCT NUMBER. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18142566 ↗
  • MalfunctionSep 12, 2023Break; Break

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PLASTIC SURGERY ON (B)(6) 2023 AND SUTURE WAS USED. THE NEEDLE WAS BROKEN AT THE SWAGE PART WHEN THE SURGEON APPLIED THE FIRST STITCH ON THE SKIN. THE NEEDLE WAS COLLECTED IMMEDIATELY. NO ADDITIONAL TREATMENT WAS PERFORMED. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. . THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17733303 ↗
  • InjuryJun 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6)2022 AND SUTURE WAS USED. POST-OP FROM (B)(6)2022 TO (B)(6)2023, THE PATIENT CAME BACK FOUR TIMES. ON (B)(6)2023, THE PATIENT'S INCISION WAS DEBRIDED AND A DIFFERENT SUTURE WAS USED AND PATIENT HEALED PROPERLY. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 17171660 ↗
  • MalfunctionMay 26, 2023Break; Material Twisted/Bent; Break; Material Twisted/Bent

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. THE NEEDLE TIP WAS EASILY BENT AND THEN BROKEN DURING INTERRUPTED SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17009502 ↗
  • MalfunctionMay 26, 2023Break; Material Twisted/Bent; Break; Material Twisted/Bent

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. THE NEEDLE TIP WAS EASILY BENT AND THEN BROKEN DURING INTERRUPTED SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17009444 ↗
  • MalfunctionDec 7, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE THREAD BROKE WHILE THE SURGEON SUTURED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15935963 ↗
  • MalfunctionOct 4, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE DURING SUTURING. IT WAS POINTED OUT THAT THERE MAY BE NO NORMAL STRENGTH IN THE SUTURE. THE PART WHERE THE SUTURE BROKE WAS UNKNOWN. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15535565 ↗
  • MalfunctionSep 15, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. AFTER MAKING THE FIRST STITCH, WHEN THE SUTURE WAS PULLED, IT BROKE. IT OCCURRED WITHOUT GRASPING WITH FORCEPS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15427743 ↗
  • MalfunctionAug 31, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. WHEN OPENING THE PACKAGE, THERE WAS A SUTURE WITHOUT NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15326373 ↗
  • MalfunctionApr 29, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. THE SUTURE BROKE DURING USE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14254120 ↗
  • MalfunctionApr 29, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. THE SUTURE BROKE DURING USE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14253995 ↗
  • MalfunctionMar 21, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN TYING A KNOT WITH AN INSTRUMENT DURING INTERRUPTED SUTURING, THE SUTURE WAS BROKEN AT THE 1ST LIGATION. AFTER THAT, THREE SUTURES WERE USED, BUT THE SAME ISSUE OCCURRED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13841452 ↗
  • MalfunctionMar 21, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN TYING A KNOT WITH AN INSTRUMENT DURING INTERRUPTED SUTURING, THE SUTURE WAS BROKEN AT THE 1ST LIGATION. AFTER THAT, THREE SUTURES WERE USED, BUT THE SAME ISSUE OCCURRED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13841451 ↗
  • MalfunctionMar 21, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN TYING A KNOT WITH AN INSTRUMENT DURING INTERRUPTED SUTURING, THE SUTURE WAS BROKEN AT THE 1ST LIGATION. AFTER THAT, THREE SUTURES WERE USED, BUT THE SAME ISSUE OCCURRED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13841449 ↗
  • InjuryMar 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY THE PATIENT THAT SHE UNDERWENT A HAND PROCEDURE IN (B)(6) 2020 AND SUTURE WAS USED FOR CLOSURE. SEVERAL WEEKS AFTER THE PROCEDURE, THE PATIENT STARTED BREAKING OUT AND HAS HAD FREQUENT DEGRANULATION'S "FLARE UPS" OF HER MAST CELL DISEASE SINCE THEN WHICH PUT HER INTO ANAPHYLACTIC SHOCK. WITH THE SUTURE, STARTED BREAKING OUT AROUND THE SITE WITH LOCALIZED MAST CELL REACTION AND PROBLEMS CASCADING FROM THAT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Anaphylactic Shock; Local Reaction

    MAUDE report 13636820 ↗

Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices

All SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices →

Sources and updates