PDS Z507GRecalls & reported problems
Ethicon, LLC · Undyed Monofilament (Polydioxanone), Absorbable Synthetic Surgical Suture · Class 2 · Implant
All models of PDS (222)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10705031061293), received between Oct 29, 2018 and Nov 12, 2024 — classified as 1 injury, 16 malfunctions.
The most frequently reported specific device problem is “Break” (11 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 10, 2023Class IIOpen, Classified
Internal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength requirement.
Z-2617-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Break11
- Material Separation4
- Component Misassembled1
- Unsealed Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
- No Consequences Or Impact To Patient3
- Foreign Body In Patient1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 12, 2024Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN OB-GYN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING CONTINUOUS SUTURING, THE SUTURE BROKE THE PRODUCT WAS USED ON SUBCUTANEOUS. FURTHER DETAILS ARE NOT PROVIDED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20665356 ↗MalfunctionNov 12, 2024Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN OB-GYN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING CONTINUOUS SUTURING, THE SUTURE BROKE THE PRODUCT WAS USED ON SUBCUTANEOUS. FURTHER DETAILS ARE NOT PROVIDED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20664310 ↗MalfunctionNov 12, 2024Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN OB-GYN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING CONTINUOUS SUTURING, THE SUTURE BROKE THE PRODUCT WAS USED ON SUBCUTANEOUS. FURTHER DETAILS ARE NOT PROVIDED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20663568 ↗MalfunctionAug 11, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(4) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE-SUTURE WAS BROKEN IMMEDIATELY DURING USE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17514644 ↗MalfunctionAug 11, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE-SUTURE WAS BROKEN IMMEDIATELY DURING USE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17513908 ↗MalfunctionMar 9, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A KNEE REPLACEMENT PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN INTERRUPTED SUTURE WAS PERFORMED WITH THE PRODUCT ON DERMIS FOR WOUND CLOSURE, THE SURGEON REALIZED THAT THE NEEDLE TIP WAS BROKEN. THE OPERATION WAS TEMPORARILY STOPPED. AFTER THAT, THE SURGERY WAS CONTINUED, AND ANOTHER PRODUCT WAS USED TO CLOSE THE WOUND. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16518895 ↗MalfunctionMar 9, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A KNEE REPLACEMENT PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN INTERRUPTED SUTURE WAS PERFORMED WITH THE PRODUCT ON DERMIS FOR WOUND CLOSURE, THE SURGEON REALIZED THAT THE NEEDLE TIP WAS BROKEN. THE OPERATION WAS TEMPORARILY STOPPED. AFTER THAT, THE SURGERY WAS CONTINUED, AND ANOTHER PRODUCT WAS USED TO CLOSE THE WOUND. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16518892 ↗MalfunctionMar 9, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A KNEE REPLACEMENT PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN INTERRUPTED SUTURE WAS PERFORMED WITH THE PRODUCT ON DERMIS FOR WOUND CLOSURE, THE SURGEON REALIZED THAT THE NEEDLE TIP WAS BROKEN. THE OPERATION WAS TEMPORARILY STOPPED. AFTER THAT, THE SURGERY WAS CONTINUED, AND ANOTHER PRODUCT WAS USED TO CLOSE THE WOUND. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16518891 ↗InjuryNov 23, 2021Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN DERMATOLOGIC SURGERY ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE TIP WAS BROKEN. AT THAT TIME, IT WAS JUDGED THAT THERE WAS NO RESIDUE IN THE BODY AND THE WOUND WAS CLOSED. HOWEVER, ONE WEEK AFTER THE SURGERY, AN X-RAY REVEALED THAT IT WAS IN THE DERMIS. HOW TO PROCEED IS UNDER DISCUSSION CURRENTLY, AND IT IS HIGHLY LIKELY THAT A REOPERATION WILL BE PERFORMED FOR REMOVAL. THE PATIENT'S CONDITION WAS GOOD AFTER THE INITIAL SURGERY. NO SECONDARY HEALTH HAZARD DUE TO RETAINED NEEDLE HAS BEEN IDENTIFIED. THOUGH THE NEEDLE PIECE IS STILL LEFT IN PATIENT, THE PATIENT IS STABLE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 12865555 ↗MalfunctionSep 16, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE DETACHED FROM THE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12484593 ↗MalfunctionAug 27, 2021Component Misassembled; Component Misassembled
IT WAS REPORTED THAT A PATIENT UNDERWENT A CESAREAN SECTION ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS FOUND THAT THERE HAD BEEN SUTURE FROM A DIFFERENT PRODUCT CODE IN A BOX OF SUTURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12380374 ↗MalfunctionJul 27, 2021Break; Break
IT WAS REPORTED THAT A LAPAROSCOPIC GASTRECTOMY AND CHOLECYSTECTOMY PROCEDURE WAS PERFORMED AND SUTURE WAS USED. DURING THE DERMAL SUTURING, THE NEEDLE COULD NOT BE PASSED SMOOTHLY THROUGH THE SECOND STITCH. WHEN CHECKING THE NEEDLE IT APPEARED BROKEN OR CRUSHED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12235068 ↗MalfunctionJul 13, 2021Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. WHEN CLOSING THE SURGICAL WOUND, AFTER PUTTING 3 STITCHES, IT BECAME HARD TO PASS THE NEEDLE THROUGH THE TISSUE. SO, WHEN CHECKING THE NEEDLE, THE NEEDLE TIP HAD BEEN BROKEN. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12162377 ↗MalfunctionApr 14, 2021Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN SURGERY ON UNKNOWN DATE AND THE SUTURE WAS USED. DURING SURGERY, WHEN PUTTING THE FIRST STITCH, IT WAS FOUND THAT A PART OF THE SUTURE HAD BEEN TORN, SO USE WAS STOPPED. IT IS UNKNOWN WHETHER THE SUTURE HAD BEEN TORN WHEN OPENING THE PACKAGE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11668785 ↗MalfunctionAug 28, 2020Unsealed Device Packaging; Unsealed Device Packaging
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2020 AND THE SUTURE WAS USED. DURING THE SURGERY, IT WAS FOUND THAT THE SUTURE HAD BEEN CAUGHT AT THE EDGE OF THE PACKAGE. WHEN OPENING THE PACKAGE, THE SUTURE FELL ONTO AN UNCLEAN AREA WITH THE PACKAGE. THE SURGEON WAS GUESSING THAT THE SUTURE MIGHT HAVE BEEN CAUGHT AT THE EDGE OF THE PACKAGE WHEN THE PACKAGE HAD BEEN SEALED. THE PROVIDED PHOTO WAS SHOWING THE SUTURE PASSED THROUGH THE SEALING AREA. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10466568 ↗MalfunctionJul 20, 2019Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN DERMATOLOGIC SURGERY ON (B)(6) 2019 AND SUTURE WAS USED. THE SUTURE WAS BROKEN EASILY DURING USE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8809858 ↗MalfunctionOct 29, 2018Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2018 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS DETACHED FROM THE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8015140 ↗
FDA clearance and classification
The FDA regulates product code NEW (suture, surgical, absorbable, polydioxanone) under 21 CFR 878.4840 as a Class II device. 21 CFR 878.4840 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the PDS Z507G
Has the PDS Z507G been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2617-2023 (Class II), began on Aug 10, 2023: Internal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength requirement. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the PDS Z507G?
The FDA's MAUDE database lists 17 adverse event reports for this device, received between Oct 29, 2018 and Nov 12, 2024. In those reports, the most common specific device problems are “Break” (11), “Material Separation” (4) and “Component Misassembled” (1); the most common patient problem is “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the PDS Z507G?
None of the reports are classified as deaths, 1 as an injury and 16 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the PDS Z507G MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.
Who makes the PDS Z507G, and is it still sold?
The FDA's device database lists Ethicon, LLC as the company responsible for it (brand PDS, model Z507G). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices
- STRATAFIX SXPP1A405Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE611 reports
- STRATAFIX SXPP1A400Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE478 reports
- STRATAFIX SXPP1A401Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE442 reports
- STRATAFIX SXPP1A301Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE442 reports
- STRATAFIX SXPP2B405Ethicon, LLC · Suture, surgical, absorbable, polydioxanone400 reports
- STRATAFIX SXPP1A406Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE334 reports
- STRATAFIX SXPP1A404Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death319 reports
- STRATAFIX SXPP1B420Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death230 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page