PDS Z509GRecalls & reported problems
Ethicon, LLC · Undyed Monofilament (Polydioxanone), Absorbable Synthetic Surgical Suture · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10705031061323), received between Nov 22, 2021 and Dec 5, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Aug 10, 2023Open, Classified
Internal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength requirement.
Z-2617-2023 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Break4
- Material Separation3
- Defective Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 5, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UROLOGICAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN UROLOGICAL SURGERY, WHEN PULLING THE NEEDLE OUT DURING SUTURING, THE NEEDLE AND SUTURE HAD BEEN DETACHED BEFORE USE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WAS NO EFFECT ON THE PATIENT. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23726885 ↗MalfunctionApr 22, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN RADIOLOGY SURGERY ON (B)(6) 2025 AND SUTURE WAS USED. DURING AN UNKNOWN RADIOLOGY SURGERY, THE NEEDLE WAS BROKEN AT THE PROXIMAL END DURING SUTURING. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21884622 ↗MalfunctionMar 13, 2024Material Separation; Defective Component; Material Separation; Defective Component
IT WAS REPORTED THAT A PATIENT UNDERWENT A GASTROENTEROLOGY PROCEDURE ON AN UNKNOWN DATE; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE SURGEON FELT THAT THE JOINT OF THE NEEDLE-SUTURE WAS THICK, AND WHEN THE SURGEON HELD IT, THE SUTURE WAS DETACHED FROM THE NEEDLE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18895024 ↗MalfunctionMar 13, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A GASTROENTEROLOGY PROCEDURE ON AN UNKNOWN DATE; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE DURING USE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18894911 ↗MalfunctionMar 22, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A PLASTIC SURGERY ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13846299 ↗MalfunctionDec 3, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CESAREAN SECTION PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. THE NEEDLE WAS DETACHED FROM THE SUTURE DURING INTERRUPTED SUTURING ON THE DERMIS. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12925570 ↗MalfunctionNov 22, 2021Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE ADN SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE TIP BROKE. THE NURSE SAID IT MAY BE BECAUSE THE SURGEON GRASPED THE NEEDLE TIP WITH FORCEPS, BUT THE SURGEON SAID HE DIDN'T. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12851726 ↗
Other Suture, surgical, absorbable, polydioxanone devices
- STRATAFIX SXPP1A405Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE617 reports
- STRATAFIX SXPP1A400Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE485 reports
- STRATAFIX SXPP1A401Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE446 reports
- STRATAFIX SXPP1A301Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE443 reports
- STRATAFIX SXPP2B405Ethicon, LLC · Suture, surgical, absorbable, polydioxanone406 reports
- STRATAFIX SXPP1A406Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE339 reports
- STRATAFIX SXPP1A404Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death319 reports
- STRATAFIX SXPP1B420Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death229 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.