ETHIBOND EXCEL D10068Recalls & reported problems

Ethicon, LLC · Green/White Braided Polyester, Nonabsorbable Surgical Suture · Class 2 · Implant

Adverse event reports
22
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 10705031134416), received between Jun 12, 2023 and Jan 16, 2026 — classified as 1 injury, 21 malfunctions.

The most frequently reported specific device problem is “Break” (8 reports).

8 other recalls cover devices cleared under the same FDA submission (K946173); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 4 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 7 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break8
  • Material Separation8
  • Device Markings/Labelling Problem3
  • Delivered as Unsterile Product2
  • Device Contaminated During Manufacture or Shipping2

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 16, 2026Break; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO TIE A KNOT WITH SUTURE. BESIDES, THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED . CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24096902 ↗
  • MalfunctionNov 6, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A VALVE REPLACEMENT PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23490321 ↗
  • MalfunctionNov 4, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE BROKE WHEN SEWING IN AN UNKNOWN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23464280 ↗
  • MalfunctionOct 29, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A MITRAL VALVE REPLACEMENT PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE BROKE WHEN SEWING IN AN UNKNOWN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23413270 ↗
  • MalfunctionOct 27, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE BROKE WHEN SEWING IN AN UNKNOWN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WAS NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23391786 ↗
  • MalfunctionOct 27, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE BROKE WHEN SEWING IN AN UNKNOWN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WAS NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23391768 ↗
  • MalfunctionOct 27, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE BROKE WHEN SEWING IN AN UNKNOWN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23391745 ↗
  • MalfunctionOct 27, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A MITRAL VALVE REPLACEMENT PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE BROKE WHEN SEWING IN AN UNKNOWN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23391730 ↗
  • MalfunctionOct 15, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23294714 ↗
  • MalfunctionOct 6, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A VALVE REPLACEMENT PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO TIE A KNOT WITH SUTURE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. ENCLOSED PLEASE FIND DEFECT PHOTOS. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23227677 ↗
  • MalfunctionSep 18, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT IT WAS FOUND THAT THERE HAD BEEN FOREIGN MATTER ON THE PRODUCT BEFORE USING ON PATIENT DURING SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23083779 ↗
  • MalfunctionMay 20, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ARTIFICIAL HEART TRANSPLANTATION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN ARTIFICIAL HEART TRANSPLANTATION SURGERY. CHANGED ANOTHER THREE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22053717 ↗
  • MalfunctionMay 20, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ARTIFICIAL HEART TRANSPLANTATION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN ARTIFICIAL HEART TRANSPLANTATION SURGERY. CHANGED ANOTHER THREE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22053715 ↗
  • MalfunctionMay 20, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ARTIFICIAL HEART TRANSPLANTATION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN ARTIFICIAL HEART TRANSPLANTATION SURGERY. CHANGED ANOTHER THREE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22053709 ↗
  • MalfunctionMay 20, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ARTIFICIAL HEART TRANSPLANTATION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN ARTIFICIAL HEART TRANSPLANTATION SURGERY. CHANGED ANOTHER THREE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22053708 ↗
  • MalfunctionFeb 11, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT BEFORE USING ON PATIENT DURING SURGERY, OPENED THE PACKING AND NOTED THE NEEDLE RUSTED. CHANGED ANOTHER TWO TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. THERE WAS NO PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21355157 ↗
  • MalfunctionJan 16, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. THE PACKAGE WAS FOUND UNSEALED BEFORE USING ON PATIENT DURING SURGERY. THIS LED TO PACKAGE LEAKAGE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21167221 ↗
  • MalfunctionSep 17, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE SUTURE WAS USED. THE DIMENSIONS PRINTED IN THE PACKAGE IS WRONG, WHICH SHOULD BE "6MM*3MM*1MM" INSTEAD OF "6MM*3MM*1.5MM". NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20241204 ↗
  • MalfunctionSep 17, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE SUTURE WAS USED. THE DIMENSIONS PRINTED IN THE PACKAGE IS WRONG, WHICH SHOULD BE "6MM*3MM*1MM" INSTEAD OF "6MM*3MM*1.5MM". NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20241188 ↗
  • MalfunctionSep 16, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. PRIOR TO USE, IT WAS NOTED THAT THE DIMENSIONS PRINTED ON THE PACKAGE WERE WRONG, AND IT SHOULD BE "6MM*3MM*1MM" INSTEAD OF "6MM*3MM*1.5MM". THERE WAS NO ILLUSTRATION OF THE "PLEDGET" IN THE PRODUCT PATTERN AREA OF PACKAGE (IT'S ONLY NEEDLE THERE WITHOUT THE PLEDGET). THE CUSTOMER IS CONFUSED WHETHER THERE'S PLEDGET WITH THIS PRODUCT CODE, AS HE COMPARED IT WITH OTHER PRODUCT CODES, WHERE THERE IS ILLUSTRATION OF THE "PLEDGET" IN THE PACKAGE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20223683 ↗
  • MalfunctionDec 14, 2023Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6)2023 AND SUTURE WAS USED. BEFORE USE ON THE PATIENT, IT WAS REPORTED THAT BEFORE THE SURGERY, THE SURGEON NOTED THE PACKAGE WITH AIR LEAKAGE ISSUE. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THE LOT OF THE IP IS UNAVAILABLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18324843 ↗
  • InjuryJun 12, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART VALVE REPLACEMENT PROCEDURE IN (B)(6)2023 AND SUTURE WAS USED. THE PATIENT WAS ADMITTED TO THE HOSPITAL DUE TO "EXERTIONAL CHEST TIGHTNESS AND SHORTNESS OF BREATH FOR 3 YEARS, WORSENING FOR HALF A MONTH" AND UNDERWENT SURGERY. DURING THE SURGERY, THE DOCTOR USED THE SUTURE TO FIX THE ARTIFICIAL HEART VALVE, WHEN THE NEEDLE WAS SEWN THROUGH THE HEART AND THE ARTIFICIAL VALVE AND THREADED OUT, THE NEEDLE PULLED OFF THE SUTURE AND THE NEEDLE FELL INTO THE PERICARDIUM. IT WAS NOT FOUND AT THAT TIME. FURTHER INFORMATION PROVIDED STATED THAT AFTER CONTINUING TO SEW THE ARTIFICIAL HEART VALVE WITH OTHER SUTURE FOR ONE LOOP, WHEN PREPARING TO TIE AND FIX THE ARTIFICIAL HEART VALVE ONE BY ONE, IT WAS FOUND THAT A NEEDLE WAS MISSING FROM THE THREAD, AND THE SEARCH BEGAN, AFTER C-ARM X-RAY FLUOROSCOPY LOCALIZATION, IT WAS FOUND IN THE PERICARDIUM THAT THE CONNECTION BETWEEN THE TAIL OF THE NEEDLE AND THE HEAD OF THE SUTURE WAS UNSTABLE AND DETACHED, AND NO ARTIFICIAL SHORTENING OR TEARING MARKS WERE FOUND. THIS EVENT AFFECTED THE SURGICAL PROCESS AND PROLONGED THE DURATION OF EXTRACORPOREAL CIRCULATION, WHICH MAY EXACERBATE THE DAMAGE CAUSED BY

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 17105960 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K946173. One submission often covers many models, so check the product and lot details in each record.

  • Z-1516-2025 · Mar 7, 2025 · Open, Classified
    ETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH
  • Z-1097-2025 · Dec 20, 2024 · Open, Classified
    PERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovas
  • Z-2486-2020 · Jun 10, 2020 · Open, Classified
    PERMA-HAND/MERSILK SILK BLU and VIRGIN SILK BLU, Product Codes EH7457G EH7983G FV768 FV769 FV7771 U7058 U7059 W1782 W1784 W1819 W1820 W818 W819 W870
  • Z-1841-2019 · Apr 19, 2019 · Terminated
    ETHIBOND EXCEL Polyester Suture-ETHIBOND Green 75cm USP1 Single Armed CT-1 Product Code: X425H Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue app
  • Z-1842-2019 · Apr 19, 2019 · Terminated
    ETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX Product Code: X865W Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue appr
  • Z-0747-2019 · Dec 18, 2018 · Terminated
    PERMA-HAND Silk Suture Size 5, Black 30" (75cm) Double Armed LR Needles, Product Code 485T Product Usage: PERMA-HAND* suture is indicated for use in general soft tissue approxima
  • Z-1839-2015 · May 8, 2015 · Terminated
    Ethicon Fast Absorbing Plain Gut - MultiPass Needles, Sterile, Absorbable Surgical Suture, 5-0 (1.5 metric), 18" (45 cm), Conventional Cutting, Lot #JD6163, Exp: July 2019. Fast ab
  • Z-0492-2015 · Oct 17, 2014 · Terminated
    Ethicon PERMA-HAND Silk siliconized black braided 10 strands per packet Ethicon LLC- Made in USA.

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices