SKYLINE 286820014Recalls & reported problems
Depuy Spine, LLC · SKYLINE ANTERIOR CERVICAL PLATE SYSTEM DRILL 2.2 x 14mm · Class 1
All models of SKYLINE (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034215310), received between Jul 27, 2023 and Jul 27, 2023 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Entrapment of Device1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 27, 2023Break; Entrapment of Device; Break; Entrapment of Device
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE SURGEON WAS DRILLING THE HOLE FOR THE SCREW USING THE VARIABLE DRILL GUIDE AND THE 12MM DRILL TO THE CORRECT PROTOCOL AS PER THE SURGICAL TECHNIQUE. IT WAS NOTICED THAT THE DRILL HAD SNAPPED. THE PATIENT WAS X-RAYED AND IT WAS SEEN THAT THE DRILL HAD SNAPPED INSIDE THE VERTEBRAL BODY OF THE PATIENT. THE SURGEON DETERMINED THAT THE DRILL END THAT HAD SNAPPED OFF COULD NOT BE RETRIEVED USING THE OPERATE SET, AND THE PIECE WAS LEFT IN SITU. ALL OTHER SCREWS IN THE PLATE WERE INSERTED WITHOUT ISSUE AND FINAL TIGHTENED. THERE WAS NO DELAY IN SURGERY. IT IS UNKNOWN IF THERE WAS A PATIENT CONSEQUENCE. THIS REPORT INVOLVES ONE SKYLINE ANTERIOR CERVICAL PLATE SYSTEM DRILL 2.2 X 14MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 17406609 ↗
Other BIT, DRILL devices
- QUICKSET 2274-25-500Depuy Orthopaedics, Inc. · BIT, DRILL1,985 reports
- QUICKSET 2274-35-500Depuy Orthopaedics, Inc. · BIT, DRILL770 reports
- QUICKSET 2274-30-500Depuy Orthopaedics, Inc. · BIT, DRILL761 reports
- NA 03.037.022Synthes GmbH · BIT, DRILL339 reports
- NA 323.062Synthes GmbH · BIT, DRILL309 reports
- QUICKSET 2274-56-000Depuy (Ireland) · BIT, DRILL275 reports
- QUICKSET 2274-45-500Depuy Orthopaedics, Inc. · BIT, DRILL211 reports
- LCS 2366-83-000Depuy (Ireland) · BIT, DRILL207 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page