N/A 698344848Recalls & reported problems
Depuy Spine, LLC · MODIFIED INSTRUMENT DRILL,SHORT 16mm · Class 1
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034575155), received between May 22, 2024 and May 22, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 22, 2024Break; Break
IT WAS REPORTED THAT ON (B)(6) 2024, THERE WAS C3-5 ACDF WITH CODA IMPLANT SYSTEM. THE DRILL BIT WAS USED WITH A DOUBLE BARREL DTS GUIDE AND RUBS AGAINST THE DRILL BIT, SHEDDING SMALL METAL PIECES. THE PIECES WERE REMOVED WITH A SUCTION AND NO PATIENT EVENT AT THAT TIME. THE DOUBLE BARREL DTS GUIDES ARE BEING REMOVED FROM THE HOSPITAL. THE DESIGN OF THE GUIDES ALLOWS FLEXING WHILE THE DRILL WAS BEING USED, WHICH CAUSED SHEDDING FROM THE DRILL BIT. IT IS UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT INVOLVES ONE (1) MODIFIED INSTRUMENT DRILL,SHORT 16MM. THIS IS REPORT 4 OF 8 FOR (B)(4). THIS COMPLAINT IS RELATED TO (B)(4), WHICH CAPTURES ADDITIONAL IMPACTED PRODUCTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19372519 ↗MalfunctionMay 22, 2024Break; Break
IT WAS REPORTED THAT ON (B)(6) 2024, THERE WAS C3-5 ACDF WITH CODA IMPLANT SYSTEM. THE DRILL BIT WAS USED WITH A DOUBLE BARREL DTS GUIDE AND RUBS AGAINST THE DRILL BIT, SHEDDING SMALL METAL PIECES. THE PIECES WERE REMOVED WITH A SUCTION AND NO PATIENT EVENT AT THAT TIME. THE DOUBLE BARREL DTS GUIDES ARE BEING REMOVED FROM THE HOSPITAL. THE DESIGN OF THE GUIDES ALLOWS FLEXING WHILE THE DRILL WAS BEING USED, WHICH CAUSED SHEDDING FROM THE DRILL BIT. IT IS UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT INVOLVES ONE (1) MODIFIED INSTRUMENT DRILL,SHORT 16MM. THIS IS REPORT 3 OF 8 FOR (B)(4). THIS COMPLAINT IS RELATED TO (B)(4), WHICH CAPTURES ADDITIONAL IMPACTED PRODUCTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19372444 ↗
Other BIT, DRILL devices
- QUICKSET 2274-25-500Depuy Orthopaedics, Inc. · BIT, DRILL1,985 reports
- QUICKSET 2274-35-500Depuy Orthopaedics, Inc. · BIT, DRILL770 reports
- QUICKSET 2274-30-500Depuy Orthopaedics, Inc. · BIT, DRILL761 reports
- NA 03.037.022Synthes GmbH · BIT, DRILL339 reports
- NA 323.062Synthes GmbH · BIT, DRILL309 reports
- QUICKSET 2274-56-000Depuy (Ireland) · BIT, DRILL275 reports
- QUICKSET 2274-45-500Depuy Orthopaedics, Inc. · BIT, DRILL211 reports
- LCS 2366-83-000Depuy (Ireland) · BIT, DRILL207 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.