Skip to content

ProDisc PDL-M-SP11SRecalls & reported problems

Synthes (U.S.A.) LP · SUPERIOR END PLATE MEDIUM 11°-STERILE · Class 3 · Implant

All models of ProDisc (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10705034700182), received between May 2, 2018 and May 14, 2026 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Expulsion” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Expulsion3
  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Bone Fracture(s)1
  • No Code Available1
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2026Expulsion

    A PATIENT WAS IMPLANTED WITH A PRODISC L DEVICE APPROXIMATELY 15 YEARS AGO. THE PATIENT FOLLOWED UP WITH THE IMPLANTATION SURGEON AFTER EXPERIENCING INCREASED LOWER BACK PAIN WHILE SQUATTING. IMAGING WAS REVIEWED ON A PEER-TO-PEER DISCUSSION IT WAS DETERMINED THAT THE POLY INLAY WAS EXPULSED OR EXPULSING. THE PATIENT IS ASYMPTOMATIC AND THIS WAS CONSIDERED AN ANCILLARY RADIOGRAPHIC FINDING. SURGEON WAS GOING TO DISCUSS POSSIBLE MEDICAL OR SURGICAL INTERVENTIONS WITH THE PATIENT. AT THIS TIME NO INTERVENTION HAS BEEN SCHEDULED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25171582 ↗
  • MalfunctionMay 14, 2026Expulsion

    A PATIENT WAS IMPLANTED WITH A PRODISC L DEVICE APPROXIMATELY 15 YEARS AGO. THE PATIENT FOLLOWED UP WITH THE IMPLANTATION SURGEON AFTER EXPERIENCING INCREASED LOWER BACK PAIN WHILE SQUATTING. IMAGING WAS REVIEWED ON A PEER-TO-PEER DISCUSSION IT WAS DETERMINED THAT THE POLY INLAY WAS EXPULSED OR EXPULSING. THE PATIENT IS ASYMPTOMATIC AND THIS WAS CONSIDERED AN ANCILLARY RADIOGRAPHIC FINDING. SURGEON WAS GOING TO DISCUSS POSSIBLE MEDICAL OR SURGICAL INTERVENTIONS WITH THE PATIENT. AT THIS TIME NO INTERVENTION HAS BEEN SCHEDULED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25171495 ↗
  • MalfunctionMay 14, 2026Expulsion

    A PATIENT WAS IMPLANTED WITH A PRODISC L DEVICE APPROXIMATELY 15 YEARS AGO. THE PATIENT FOLLOWED UP WITH THE IMPLANTATION SURGEON AFTER EXPERIENCING INCREASED LOWER BACK PAIN WHILE SQUATTING. IMAGING WAS REVIEWED ON A PEER-TO-PEER DISCUSSION IT WAS DETERMINED THAT THE POLY INLAY WAS EXPULSED OR EXPULSING. THE PATIENT IS ASYMPTOMATIC AND THIS WAS CONSIDERED AN ANCILLARY RADIOGRAPHIC FINDING. SURGEON WAS GOING TO DISCUSS POSSIBLE MEDICAL OR SURGICAL INTERVENTIONS WITH THE PATIENT. AT THIS TIME NO INTERVENTION HAS BEEN SCHEDULED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25171494 ↗
  • InjuryJan 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A PATIENT WAS IMPLANTED WITH A PRODISC L IMPLANT AT L5-S1 ON (B)(6) 2019. PATIENT EXPERIENCED INSTABILITY AT THE LEVEL DUE TO A BILATERAL PARS FRACTURE. THE PDL IMPLANT WAS REMOVED AND THE PATIENT WAS FUSED ON (B)(6) 2025.

    Patient problems: Bone Fracture(s)

    MAUDE report 21273771 ↗
  • InjuryMay 2, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PATIENT UNDERWENT A PRODISC-L REMOVAL AT L5-S1 ON (B)(6) 2018, DUE TO PAIN. THE PATIENT WAS IMPLANTED WITH THE PRODISC-L ON AN UNKNOWN DATE APPROXIMATELY ONE (1) YEAR AGO AND THE ADJACENT LEVELS ARE FUSED WITH NON-SYNTHES DEVICES. NO OTHER LEVELS WERE OPERATED ON. THIS WAS NOT AN ADJACENT LEVEL CASE. THE REMOVAL WENT SMOOTHLY AND THE PRODISC APPEARED TO BE IN GOOD CONDITION AND FUNCTIONING AS INTENDED. IT WAS REMOVED EASILY AND A FUSION DEVICE WAS SUCCESSFULLY IMPLANTED. THE REVISION WAS DUE TO THE PATIENT'S CONTINUED BACK PAIN WHICH THE SURGEON BELIEVED WOULD BE RESOLVED THROUGH A FUSION. NO SURGICAL DELAY AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. CONCOMITANT DEVICES REPORTED: UNKNOWN FUSION HARDWARE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # UNKNOWN).

    Patient problems: No Code Available; Pain

    MAUDE report 7481874 ↗

Other PROSTHESIS, INTERVERTEBRAL DISC devices

All PROSTHESIS, INTERVERTEBRAL DISC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page