Skip to content

NA 03.615.100Recalls & reported problems

Synthes GmbH · RETRACTOR FRAME CRANIAL/CAUDAL · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10705034715360), received between Nov 15, 2017 and Feb 18, 2026 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Compatibility Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Compatibility Problem3
  • Device-Device Incompatibility3
  • Difficult to Open or Close2
  • Mechanical Problem2
  • Device Operates Differently Than Expected1
  • Device Slipped1
  • Failure To Adhere Or Bond1
  • Material Twisted/Bent1

Patient problems

  • No Consequences Or Impact To Patient5
  • No Clinical Signs, Symptoms or Conditions4
  • No Patient Involvement3
  • No Code Available1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 18, 2026Mechanical Problem

    IT WAS REPORTED THE SMALL SCREWS THAT SECURE THE PIECE FOR RETRACTION OF THE BLADES KEEP LOOSENING AND ALLOWING THE PIECE FOR RETRACTION TO DETACH FROM THE RETRACTOR FRAME. NO PATIENT HARM. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24376704 ↗
  • MalfunctionSep 2, 2025Mechanical Problem; Compatibility Problem; Mechanical Problem; Compatibility Problem

    IT WAS REPORTED THAT PATIENT UNDERWENT A & P LUMBAR FUSION ON (B)(6) 2025. THE RETRACTOR FRAME GOT STUCK REPEATED TIMES DURING RETRACTION AND THE BLADES WOULD NOT ANGULATE WHILE USING THE HANDLES. THE BLADES WERE USED WITHOUT THE PROPER RETRACTION. NO FRAGMENTS WERE GENERATED. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22938327 ↗
  • MalfunctionOct 6, 2023Compatibility Problem; Compatibility Problem

    EVENT OCCURRED INTRAOPERATIVELY ON (B)(6) 2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17890033 ↗
  • MalfunctionSep 26, 2023Compatibility Problem; Compatibility Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE MIS CRANIAL/CAUDAL DISTRACTOR IN QUESTION WAS NOT HOLDING PROPERLY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO PATIENT CONSEQUENCES. NO FRAGMENTS WERE GENERATED. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS REPORT IS FOR A RETRACTOR FRAME CRANIAL/CAUDAL. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17818365 ↗
  • MalfunctionFeb 26, 2020Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A ROUTINE INSPECTION OF CONSIGNMENT SETS ON (B)(6) 2020, THE INSIGHT OF A RETRACTOR FRAME WOULD NOT HOLD THE ARTICULATION WITH THE ARM IS TOED OUT AND THE TEETH MUST BE WORN OUT. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9758936 ↗
  • MalfunctionFeb 21, 2020Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2002, THE SURGEON WAS PERFORMING A L3-L5 MICRO DISC ON A PATIENT WITH THE SYNTHES INSIGHT BLADE RETRACTOR. THE SURGEON COULD NOT EXPAND OR TOE OUT EITHER OF THE RETRACTOR FRAMES, TRIED BOTH THE RETRACTOR FRAME MEDIAL/LATERAL AND RETRACTOR FRAME CRANIAL/CAUDAL. THE SURGEON ALSO DEMONSTRATED HOW ON 1 OF THE 2 ARMS THAT RATCHET THAT HOLDS THE ARM ARTICULATED MUST BE WORN OUT. THE ARM ON BOTH OF THESE RETRACTOR FRAMES WILL NOT HOLD. SURGICAL DELAY WAS ABOUT 4 MINUTES WHILE THE SURGEON ADJUSTED THE RETRACTOR TO MAKE IT WORK. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO HARM TO THE PATIENT. THIS IS REPORT 02 OF 02 OF (B)(4).

    Patient problems: No Code Available

    MAUDE report 9741269 ↗
  • MalfunctionSep 13, 2019Device Slipped; Device Slipped

    IT WAS REPORTED THAT ON (B)(6) 2019 DURING SURGERY THE RETRACTOR FRAME MEDIAL LATERAL AND RETRACTOR FRAME CRANIAL CAUDAL DISTRACTORS OF THE MINIMALLY INVASIVE SURGICAL SUPPORT SYSTEM HAVE WORN OUT AND SLIPPING DURING SURGERY. THEY NO LONGER HOLD THE POSITION TIGHTLY THEY ARE INTENDED TO. SURGICAL DELAY IS UNKNOWN. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY. PATIENT STATUS IS UNKNOWN. THIS IS REPORT 2 FOR 3 (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9035285 ↗
  • MalfunctionAug 17, 2018Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE RETRACTOR FRAME CRANIAL/CAUDAL IS NOT HOLDING. IT IS UNKNOWN WHEN AND HOW THE MALFUNCTION WAS DISCOVERED. IT IS UNKNOWN IF THERE WAS PATIENT AND PROCEDURAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) RETRACTOR FRAME CRANIAL/CAUDAL. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement

    MAUDE report 7794764 ↗
  • MalfunctionMay 10, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected

    IT WAS REPORTED THAT ONE (1) RETRACTOR FRAME CRANIAL/CAUDAL FROM A SYNTHES FRAMES AND HANDLES CADDY DOES NOT HOLD POSITION AND IT SLOWLY SLIPS/ GIVES OVER TIME. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient; No Patient Involvement

    MAUDE report 7504919 ↗
  • MalfunctionNov 28, 2017Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT A RETRACTOR FRAME CRANIAL/CAUDAL IS NOT HOLDING THE RETRACTOR. THE DEVICE WILL NOT STAY OPEN. THE LOCKING MECHANISM WILL NOT HOLD AND THE GEAR INSIDE THE DEVICE MAY BE STRIPPED. THIS WAS DISCOVERED IN CENTRAL STERILE PROCESSING. THERE IS NO PATIENT OR PROCEDURE INVOLVEMENT. CONCOMITANT DEVICES REPORTED: RETRACTOR (PART NUMBER UNKNOWN, LOT NUMBER: UNKNOWN, QTY. 1) THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7062979 ↗
  • MalfunctionNov 27, 2017Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT A RETRACTOR FRAME CRANIAL/CAUDAL IS NOT HOLDING THE RETRACTOR. THE DEVICE WILL NOT STAY OPEN. THE LOCKING MECHANISM WILL NOT HOLD AND THE GEAR INSIDE THE DEVICE MAY BE STRIPPED. THIS WAS DISCOVERED IN CENTRAL STERILE PROCESSING. THERE IS NO PATIENT OR PROCEDURE INVOLVEMENT. CONCOMITANT DEVICES REPORTED: RETRACTOR (PART NUMBER UNKNOWN, LOT NUMBER: UNKNOWN, QTY. 1). THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7062068 ↗
  • MalfunctionNov 15, 2017Failure To Adhere Or Bond; Failure To Adhere Or Bond

    DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THE FLEX ARMS ARE NOT HOLDING THE RETRACTOR FRAMES (CRANIAL CAUDAL AND MEDIAL LATERAL). IT IS IMPOSSIBLE TO PUSH THE FRAMES INTO THE HOLDING RING SLEEVE, SOMETHING IS BLOCKING INSIDE. THIS WAS DETECTED BEFORE THE SURGERY. NO PATIENT WAS INVOLVED. CONCOMITANT DEVICE REPORTED: HOLDING SLEEVE RING (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) RETRACTOR FRAME CRANIAL/CAUDAL. THIS IS REPORT 4 OF 4 FOR COMPLAINT (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 7033308 ↗

FDA clearance and classification

The FDA regulates product code GAD (retractor) under 21 CFR 878.4800 as a Class I device. 21 CFR 878.4800 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA 03.615.100

Has the NA 03.615.100 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705034715360), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA 03.615.100?

The FDA's MAUDE database lists 12 adverse event reports for this device, received between Nov 15, 2017 and Feb 18, 2026. In those reports, the most common specific device problems are “Compatibility Problem” (3), “Device-Device Incompatibility” (3) and “Difficult to Open or Close” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 03.615.100?

None of the reports are classified as deaths, 0 as injuries and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 03.615.100 FDA approved?

Devices of this type (product code GAD) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.

Who makes the NA 03.615.100, and is it still sold?

The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 03.615.100). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other RETRACTOR devices

All RETRACTOR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page