Skip to content

SYNFIX Evolution 03.835.032Recalls & reported problems

Synthes GmbH · SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE · Class 1

All models of SYNFIX Evolution (21)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10705034814179), received between Dec 20, 2016 and May 5, 2026 — classified as 1 injury, 9 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Device Operates Differently Than Expected2
  • Difficult to Advance2
  • Dull, Blunt2
  • Failure to Cut2
  • Material Twisted/Bent2
  • Adverse Event Without Identified Device or Use Problem1
  • Contamination /Decontamination Problem1

Patient problems

  • No Consequences Or Impact To Patient3
  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2
  • No Patient Involvement2
  • No Code Available1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 5, 2026Failure to Cut; Material Twisted/Bent

    IT WAS REPORTED ON AN UNKNOWN DATE, THAT WHEN INSPECTING INSTRUMENTS RECEIVED THROUGH DECONTAMINATION, PRIOR TO STERILIZATION - THE SYNFIX EVOLUTION AWLS AND 2 SCREWDRIVER SHAFTS WERE OBSERVED TO BE WORN. THE AWL WAS NO LONGER SHARP AND THE DRIVER TIPS WERE BEGINNING TO STRIP. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25082861 ↗
  • MalfunctionMay 5, 2026Failure to Cut; Material Twisted/Bent

    IT WAS REPORTED ON AN UNKNOWN DATE, THAT WHEN INSPECTING INSTRUMENTS RECEIVED THROUGH DECONTAMINATION, PRIOR TO STERILIZATION - THE SYNFIX EVOLUTION AWLS AND 2 SCREWDRIVER SHAFTS WERE OBSERVED TO BE WORN. THE AWL WAS NO LONGER SHARP AND THE DRIVER TIPS WERE BEGINNING TO STRIP. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25082860 ↗
  • MalfunctionJun 5, 2020Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING A ROUTINE INCOMING INSPECTION OF THE LOANER SET, IT WAS OBSERVED THAT THE SYNFIX EVOLUTION AWL WITHOUT SLEEVE WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR 1 SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 10125397 ↗
  • MalfunctionMay 18, 2020Dull, Blunt; Dull, Blunt

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING AN ANTERIOR LUMBAR INTERBODY FUSION PROCEDURE ON (B)(6) 2020, A SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE WAS NOTED TO BE BLUNT. THE SURGEON ATTEMPTED TO USE A SECOND SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE THAT WAS OBSERVED TO BE BLUNT. THE SURGEON WAS UNSUCCESSFUL IN USING THE SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICE REPORTED: UNK - SCREWS: TRAUMA (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR ONE (1) SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10067151 ↗
  • MalfunctionMay 15, 2020Dull, Blunt; Dull, Blunt

    UPDATED EVENT DESCRIPTION IT WAS REPORTED THAT DURING AN ANTERIOR LUMBAR INTERBODY FUSION PROCEDURE ON (B)(6) 2020, A SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE WAS NOTED TO BE BLUNT. THE SURGEON ATTEMPTED TO USE A SECOND SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE THAT WAS OBSERVED TO BE BLUNT. THE SURGEON WAS UNSUCCESSFUL IN USING THE SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE REPORTED. CONCOMITANT DEVICE REPORTED: UNK - SCREWS: TRAUMA (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10063499 ↗
  • MalfunctionNov 4, 2019Break; Break

    IT WAS REPORTED THAT DURING A ROUTINE INCOMING INSPECTION OF A LOANER SET ON AN UNKNOWN DATE, THE SYNFIX EVOLUTION AWL WAS OBSERVED TO BE BROKEN. THERE WERE NO PATIENT AND SURGICAL INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS IS 1 OF 1 FOR REPORT (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9273358 ↗
  • MalfunctionMay 8, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE SYNFIX EVOLUTION SET, STANDARD INSTRUMENTS-EVAL WAS USED IN SURGERY ON (B)(6) 2019. ON (B)(6) 2019, FIELD STOCKING LOCATION (FSL) WAS INFORMED THAT THE SET WAS READY FOR PICK UP ON THAT DATE. THE SCREW LOADING STATION FOR SYNFIX EVOLUTION, SYNFIX EVOLUTION AWL/WITHOUT SLEEVE, HANDLE WITH QUICK COUPLING-SMALL, AND TWO (2) SYNFIX EVOLUTION SCREWDRIVER WITHOUT THREAD LOCK SLEEVE WAS PICKED UP BY OWENS AND MINORS. UPON INSPECTION, QWENS AND MINORS FOUND BLOOD ON THE PARTS OF THE SET AND REPORTED TO FSL. IT WAS FURTHER CLARIFIED THAT THESE ARE INSTRUMENT TRAYS USED BY FSAL AND THE TRAYS WERE DISCOVERED TO BE "SOILED". THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) SYNFIX EVOLUTION AWL/WITHOUT SLEEVE. THIS IS REPORT 2 OF 5 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 8591934 ↗
  • InjuryNov 6, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2018, ANTERIOR LUMBAR INTERBODY FUSION (ALIF) WAS PERFORMED AT L4-S1 WITH SYNTHES SYNFIX EVOLUTION. THERE WERE DIFFICULTIES PLACING ONE OF THE S1 SCREWS ON THE L5-S1 LEVEL. AFTER INSERTING THE SPACER AT L5-S1, THE AWL WAS PLACED THROUGH THE AIMING DEVICE TO CANNULATE S1. IT WAS VERY DIFFICULT TO FULLY ADVANCE THE AWL INTO THE "SCLEROTIC" BONE THAT THE PATIENT HAD AT S1. THE AWL APPEARED TO BE 15MM IN LENGTH BUT THE SCREWS BEING USED WERE 25MM. THIS LEFT 10MM OF BONE NOT PREPPED FOR A SCREW. THIS WAS CONFIRMED BY VIEWING THE CT SCAN ON THE IMAGING IN THE ROOM - THE BONE WAS DENSE AND SCLEROTIC. AFTER ADVANCING THE AWL BY MALLETING AND TWISTING, A FINE TIP SCREW WAS PLACED INTO THE HOLE OF S1. AFTER ADVANCING THE SCREW ABOUT 2/3 OF THE WAY THROUGH THAT PLATE ON THE SPACER, THE SCREW WOULD NOT ADVANCE ANYMORE. THE SCREW WAS TAKEN OUT. THE SURGEON WANTED TO DRILL INTO RECANNULATED HOLE DEEPER INTO S1 SO HE COULD ALLOW THE SCREW TO ADVANCE. WITH NO DRILL OPTION, THE AIMING DEVICE WAS ELECTED TO BE TAKEN OFF AND THE AWL "FREEHAND" WAS USED WHICH IS NOT SUGGESTED IN THE TECHNIQUE GUIDE. THE SCREW WAS PLACED WITHOUT THE AIMING DEVICE ATTACHED TO THE SPACER. U

    Patient problems: No Code Available

    MAUDE report 8044105 ↗
  • MalfunctionDec 20, 2016Device Operates Differently Than Expected; Difficult to Advance; Device Operates Differently Than Expected; Difficult to Advance

    IT WAS REPORTED THAT DURING AN UNKNOWN SPINE PROCEDURE, WHILE USING AN ANGLED AWL TO PENETRATE THE VERTEBRAL BODY, THE SURGEON HAD DIFFICULTY WHILE PREPARING THREE (3) SITES FOR IMPLANTATION. THE SURGEON BELIEVES THAT THE SHOULDER OF THE AWL WAS BECOMING HUNG UP ON THE AIMING DEVICE, CAUSING THE AWL TO FAIL TO ADVANCE THROUGH THE VERTEBRAE. AFTER MULTIPLE ATTEMPTS AT EACH SITE, THE SURGEON UTILIZED A MALLET TO SUCCESSFULLY PENETRATE THE VERTEBRAL BODY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE PATIENT WAS REPORTED TO BE STABLE. NO ADVERSE EVENTS WERE REPORTED. CONCOMITANT MEDICAL PRODUCTS: UNKNOWN HOSPITAL MALLET (PART #: UNKNOWN, LOT #: UNKNOWN, QTY. 1). THIS IS REPORT NUMBER 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6191335 ↗
  • MalfunctionDec 20, 2016Device Operates Differently Than Expected; Difficult to Advance; Device Operates Differently Than Expected; Difficult to Advance

    IT WAS REPORTED THAT DURING AN UNKNOWN SPINE PROCEDURE, WHILE USING AN ANGLED AWL TO PENETRATE THE VERTEBRAL BODY, THE SURGEON HAD DIFFICULTY WHILE PREPARING THREE (3) SITES FOR IMPLANTATION. THE SURGEON BELIEVES THAT THE SHOULDER OF THE AWL WAS BECOMING HUNG UP ON THE AIMING DEVICE, CAUSING THE AWL TO FAIL TO ADVANCE THROUGH THE VERTEBRAE. AFTER MULTIPLE ATTEMPTS AT EACH SITE, THE SURGEON UTILIZED A MALLET TO SUCCESSFULLY PENETRATE THE VERTEBRAL BODY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE PATIENT WAS REPORTED TO BE STABLE. NO ADVERSE EVENTS WERE REPORTED. CONCOMITANT DEVICES REPORTED: UNKNOWN HOSPITAL MALLET (PART #: UNKNOWN, LOT #: UNKNOWN, QTY. 1). THIS IS REPORT NUMBER 1 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6191175 ↗

FDA clearance and classification

The FDA regulates product code HWJ (awl) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the SYNFIX Evolution 03.835.032

Has the SYNFIX Evolution 03.835.032 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705034814179), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the SYNFIX Evolution 03.835.032?

The FDA's MAUDE database lists 10 adverse event reports for this device, received between Dec 20, 2016 and May 5, 2026. In those reports, the most common specific device problems are “Break” (2), “Device Operates Differently Than Expected” (2) and “Difficult to Advance” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SYNFIX Evolution 03.835.032?

None of the reports are classified as deaths, 1 as an injury and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SYNFIX Evolution 03.835.032 FDA approved?

Devices of this type (product code HWJ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.

Who makes the SYNFIX Evolution 03.835.032, and is it still sold?

The FDA's device database lists Synthes GmbH as the company responsible for it (brand SYNFIX Evolution, model 03.835.032). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other AWL devices

All AWL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page