SYNFIX Evolution 03.835.032Recalls & reported problems
Synthes GmbH · SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE · Class 1
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10705034814179), received between Dec 20, 2016 and Sep 29, 2026 — classified as 1 injury, 15 malfunctions.
The most frequently reported specific device problem is “Failure to Cut” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Cut8
- Break2
- Device Operates Differently Than Expected2
- Difficult to Advance2
- Dull, Blunt2
- Material Twisted/Bent2
- Adverse Event Without Identified Device or Use Problem1
- Contamination /Decontamination Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient3
- No Known Impact Or Consequence To Patient2
- No Patient Involvement2
- No Code Available1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Failure to Cut
It was reported the awls in the synfix evo trays are dull. No patient harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50933052 ↗MalfunctionSep 29, 2026Failure to Cut
It was reported the awls in the synfix evo trays are dull. No patient harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50933026 ↗MalfunctionSep 29, 2026Failure to Cut
It was reported the awls in the synfix evo trays are dull. No patient harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50933018 ↗MalfunctionSep 29, 2026Failure to Cut
It was reported the awls in the synfix evo trays are dull. No patient harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932988 ↗MalfunctionSep 29, 2026Failure to Cut
It was reported the awls in the synfix evo trays are dull. No patient harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932973 ↗MalfunctionSep 29, 2026Failure to Cut
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932791 ↗MalfunctionMay 5, 2026Failure to Cut; Material Twisted/Bent
IT WAS REPORTED ON AN UNKNOWN DATE, THAT WHEN INSPECTING INSTRUMENTS RECEIVED THROUGH DECONTAMINATION, PRIOR TO STERILIZATION - THE SYNFIX EVOLUTION AWLS AND 2 SCREWDRIVER SHAFTS WERE OBSERVED TO BE WORN. THE AWL WAS NO LONGER SHARP AND THE DRIVER TIPS WERE BEGINNING TO STRIP. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25082861 ↗MalfunctionMay 5, 2026Failure to Cut; Material Twisted/Bent
IT WAS REPORTED ON AN UNKNOWN DATE, THAT WHEN INSPECTING INSTRUMENTS RECEIVED THROUGH DECONTAMINATION, PRIOR TO STERILIZATION - THE SYNFIX EVOLUTION AWLS AND 2 SCREWDRIVER SHAFTS WERE OBSERVED TO BE WORN. THE AWL WAS NO LONGER SHARP AND THE DRIVER TIPS WERE BEGINNING TO STRIP. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25082860 ↗MalfunctionJun 5, 2020Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING A ROUTINE INCOMING INSPECTION OF THE LOANER SET, IT WAS OBSERVED THAT THE SYNFIX EVOLUTION AWL WITHOUT SLEEVE WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR 1 SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Patient Involvement
MAUDE report 10125397 ↗MalfunctionMay 18, 2020Dull, Blunt; Dull, Blunt
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING AN ANTERIOR LUMBAR INTERBODY FUSION PROCEDURE ON (B)(6) 2020, A SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE WAS NOTED TO BE BLUNT. THE SURGEON ATTEMPTED TO USE A SECOND SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE THAT WAS OBSERVED TO BE BLUNT. THE SURGEON WAS UNSUCCESSFUL IN USING THE SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICE REPORTED: UNK - SCREWS: TRAUMA (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR ONE (1) SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10067151 ↗MalfunctionMay 15, 2020Dull, Blunt; Dull, Blunt
UPDATED EVENT DESCRIPTION IT WAS REPORTED THAT DURING AN ANTERIOR LUMBAR INTERBODY FUSION PROCEDURE ON (B)(6) 2020, A SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE WAS NOTED TO BE BLUNT. THE SURGEON ATTEMPTED TO USE A SECOND SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE THAT WAS OBSERVED TO BE BLUNT. THE SURGEON WAS UNSUCCESSFUL IN USING THE SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE REPORTED. CONCOMITANT DEVICE REPORTED: UNK - SCREWS: TRAUMA (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) SYNFIX® EVOLUTION AWL/WITHOUT SLEEVE. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10063499 ↗MalfunctionNov 4, 2019Break; Break
IT WAS REPORTED THAT DURING A ROUTINE INCOMING INSPECTION OF A LOANER SET ON AN UNKNOWN DATE, THE SYNFIX EVOLUTION AWL WAS OBSERVED TO BE BROKEN. THERE WERE NO PATIENT AND SURGICAL INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS IS 1 OF 1 FOR REPORT (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9273358 ↗MalfunctionMay 8, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE SYNFIX EVOLUTION SET, STANDARD INSTRUMENTS-EVAL WAS USED IN SURGERY ON (B)(6) 2019. ON (B)(6) 2019, FIELD STOCKING LOCATION (FSL) WAS INFORMED THAT THE SET WAS READY FOR PICK UP ON THAT DATE. THE SCREW LOADING STATION FOR SYNFIX EVOLUTION, SYNFIX EVOLUTION AWL/WITHOUT SLEEVE, HANDLE WITH QUICK COUPLING-SMALL, AND TWO (2) SYNFIX EVOLUTION SCREWDRIVER WITHOUT THREAD LOCK SLEEVE WAS PICKED UP BY OWENS AND MINORS. UPON INSPECTION, QWENS AND MINORS FOUND BLOOD ON THE PARTS OF THE SET AND REPORTED TO FSL. IT WAS FURTHER CLARIFIED THAT THESE ARE INSTRUMENT TRAYS USED BY FSAL AND THE TRAYS WERE DISCOVERED TO BE "SOILED". THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) SYNFIX EVOLUTION AWL/WITHOUT SLEEVE. THIS IS REPORT 2 OF 5 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8591934 ↗InjuryNov 6, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2018, ANTERIOR LUMBAR INTERBODY FUSION (ALIF) WAS PERFORMED AT L4-S1 WITH SYNTHES SYNFIX EVOLUTION. THERE WERE DIFFICULTIES PLACING ONE OF THE S1 SCREWS ON THE L5-S1 LEVEL. AFTER INSERTING THE SPACER AT L5-S1, THE AWL WAS PLACED THROUGH THE AIMING DEVICE TO CANNULATE S1. IT WAS VERY DIFFICULT TO FULLY ADVANCE THE AWL INTO THE "SCLEROTIC" BONE THAT THE PATIENT HAD AT S1. THE AWL APPEARED TO BE 15MM IN LENGTH BUT THE SCREWS BEING USED WERE 25MM. THIS LEFT 10MM OF BONE NOT PREPPED FOR A SCREW. THIS WAS CONFIRMED BY VIEWING THE CT SCAN ON THE IMAGING IN THE ROOM - THE BONE WAS DENSE AND SCLEROTIC. AFTER ADVANCING THE AWL BY MALLETING AND TWISTING, A FINE TIP SCREW WAS PLACED INTO THE HOLE OF S1. AFTER ADVANCING THE SCREW ABOUT 2/3 OF THE WAY THROUGH THAT PLATE ON THE SPACER, THE SCREW WOULD NOT ADVANCE ANYMORE. THE SCREW WAS TAKEN OUT. THE SURGEON WANTED TO DRILL INTO RECANNULATED HOLE DEEPER INTO S1 SO HE COULD ALLOW THE SCREW TO ADVANCE. WITH NO DRILL OPTION, THE AIMING DEVICE WAS ELECTED TO BE TAKEN OFF AND THE AWL "FREEHAND" WAS USED WHICH IS NOT SUGGESTED IN THE TECHNIQUE GUIDE. THE SCREW WAS PLACED WITHOUT THE AIMING DEVICE ATTACHED TO THE SPACER. U
Patient problems: No Code Available
MAUDE report 8044105 ↗MalfunctionDec 20, 2016Device Operates Differently Than Expected; Difficult to Advance; Device Operates Differently Than Expected; Difficult to Advance
IT WAS REPORTED THAT DURING AN UNKNOWN SPINE PROCEDURE, WHILE USING AN ANGLED AWL TO PENETRATE THE VERTEBRAL BODY, THE SURGEON HAD DIFFICULTY WHILE PREPARING THREE (3) SITES FOR IMPLANTATION. THE SURGEON BELIEVES THAT THE SHOULDER OF THE AWL WAS BECOMING HUNG UP ON THE AIMING DEVICE, CAUSING THE AWL TO FAIL TO ADVANCE THROUGH THE VERTEBRAE. AFTER MULTIPLE ATTEMPTS AT EACH SITE, THE SURGEON UTILIZED A MALLET TO SUCCESSFULLY PENETRATE THE VERTEBRAL BODY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE PATIENT WAS REPORTED TO BE STABLE. NO ADVERSE EVENTS WERE REPORTED. CONCOMITANT MEDICAL PRODUCTS: UNKNOWN HOSPITAL MALLET (PART #: UNKNOWN, LOT #: UNKNOWN, QTY. 1). THIS IS REPORT NUMBER 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6191335 ↗MalfunctionDec 20, 2016Device Operates Differently Than Expected; Difficult to Advance; Device Operates Differently Than Expected; Difficult to Advance
IT WAS REPORTED THAT DURING AN UNKNOWN SPINE PROCEDURE, WHILE USING AN ANGLED AWL TO PENETRATE THE VERTEBRAL BODY, THE SURGEON HAD DIFFICULTY WHILE PREPARING THREE (3) SITES FOR IMPLANTATION. THE SURGEON BELIEVES THAT THE SHOULDER OF THE AWL WAS BECOMING HUNG UP ON THE AIMING DEVICE, CAUSING THE AWL TO FAIL TO ADVANCE THROUGH THE VERTEBRAE. AFTER MULTIPLE ATTEMPTS AT EACH SITE, THE SURGEON UTILIZED A MALLET TO SUCCESSFULLY PENETRATE THE VERTEBRAL BODY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE PATIENT WAS REPORTED TO BE STABLE. NO ADVERSE EVENTS WERE REPORTED. CONCOMITANT DEVICES REPORTED: UNKNOWN HOSPITAL MALLET (PART #: UNKNOWN, LOT #: UNKNOWN, QTY. 1). THIS IS REPORT NUMBER 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6191175 ↗
Other AWL devices
- Meridian RA2-003015Seaspine Orthopedics Corporation · AWL1 recall1 report
- AMP 117-165Alphatec Spine, Inc. · AWL1 recall0 reports
- NA 03.010.041Synthes GmbH · AWL7 reports
- ENDOSKELETON® TAS 2300-1110Medtronic Sofamor Danek, Inc. · AWL7 reports
- NA 359.213Synthes GmbH · AWL7 reports
- NA 03.820.100Synthes GmbH · AWL6 reports
- Endoskeleton™ TCS Interbody System 5210-1063Medtronic Sofamor Danek, Inc. · Awl5 reports
- na 7080911Medtronic Sofamor Danek, Inc. · AWL4 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.