NA 359.213Recalls & reported problems
Synthes GmbH · AWL · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886982197146), received between Jan 17, 2017 and Jan 8, 2026 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Contamination /Decontamination Problem2
- Failure to Cut1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Device Embedded In Tissue or Plaque1
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 8, 2026Break
THE TIP OF THE INSTRUMENT BROKE OFF DURING USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24010295 ↗MalfunctionJun 28, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT ON (B)(6) 2024, THE PATIENT UNDERWENT AN UNKNOWN SURGERY. NUMEROUS STAINS WERE ADHERING TO THE RUBBER PART OF SEVERAL PRODUCTS. THE INSTRUMENTATION NURSE, AT THE DISCRETION OF THE PRIMARY SURGEON, WIPED OFF THE STAINS WITH A TOWEL, AND THE SAID PRODUCTS WERE USED DURING THE SURGERY. IT WAS CONFIRMED THAT THE PRODUCTS WERE ALSO MORE WATER-REPELLENT THAN USUAL DURING POST-OPERATIVE CLEANING. THIS REPORT IS FOR ONE AWL F/TEN THIS IS REPORT 2 OF 4 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 19636874 ↗MalfunctionJun 28, 2023Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE INSTRUMENTS WERE BROKEN. THERE WERE NO PATIENT CONSEQUENCES. NO FURTHER DETAILS ARE AVAILABLE. THIS REPORT INVOLVES ONE AWL. THIS IS REPORT 5 OF 7 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17216716 ↗MalfunctionDec 7, 2022Failure to Cut; Failure to Cut
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN COLOMBIA AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2022, THAT THE DEVICE DOES NOT HAVE A SHARP EDGE. THIS REPORT IS FOR ONE (1) AWL F/TEN THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15931057 ↗MalfunctionOct 17, 2022Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED THAT ON(B)(6)2022 , THE ITEMS WERE BROKEN DURING A PROCEDURE. THERE WAS A SURGICAL DELAY OF 15 MINUTES AS A RESULT. ANOTHER KIT WAS OPENED TO COMPLETE THE PROCEDURE. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE AWL. THIS IS REPORT 1 OF 5 FOR PC-(B)(4). .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15616535 ↗InjuryMay 17, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021 DURING AN UNKNOWN PROCEDURE, A BROWN SUBSTANCE LEAKED FROM THE INSTRUMENTS THAT RESULTED IN CONTAMINATION OF THE HAMMER, INSERTER AND AWL. THERE WAS A SURGICAL DELAY OF (30) MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING REPLACEMENT DEVICES. IT WAS STATED THAT THE PATIENT REQUIRED AN OPEN REDUCTION INTERNAL FIXATION PROCEDURE DUE TO THE INSTRUMENTS NOT BEING AVAILABLE. THIS REPORT IS FOR (1) AWL. THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11837856 ↗InjuryJan 17, 2017Break; Break
DEVICE REPORT FROM SYNTHES ON AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT IS WAS AN OPEN FRACTURE REDUCTION IN HUMERAL SHAFT WITH INTERNAL FIXATION. WHEN THE SPECIALIST PERFORMED THE PIERCING WITH THE PUNCH, THE TIP BROKE. NO PROLONGATION OF THE SURGERY WAS REPORTED. PATIENT STATUS IS NORMAL. THE PUNCTURE TIP, THAT WAS SPLIT, WAS INSIDE THE CHANNEL OF THE PATIENT'S BONE; HOWEVER, THE SPECIALIST STATED THAT IT DOES NOT REPRESENT ANY RISK TO THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO OTHER MEDICAL INTERVENTION WAS REQUIRED. THIS COMPLAINT INVOLVES 1 PART. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: Device Embedded In Tissue or Plaque; No Known Impact Or Consequence To Patient
MAUDE report 6258797 ↗
FDA clearance and classification
The FDA regulates product code HWJ (awl) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 359.213
Has the NA 359.213 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982197146), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 359.213?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Jan 17, 2017 and Jan 8, 2026. In those reports, the most common specific device problems are “Break” (4), “Contamination /Decontamination Problem” (2) and “Failure to Cut” (1); the most common patient problem is “Device Embedded In Tissue or Plaque” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 359.213?
None of the reports are classified as deaths, 2 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 359.213 FDA approved?
Devices of this type (product code HWJ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.
Who makes the NA 359.213, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 359.213). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other AWL devices
- Meridian RA2-003015Seaspine Orthopedics Corporation · AWL1 recall1 report
- AMP 117-165Alphatec Spine, Inc. · AWL1 recall0 reports
- SYNFIX Evolution 03.835.032Synthes GmbH · AWL10 reports
- NA 03.010.041Synthes GmbH · AWL7 reports
- ENDOSKELETON® TAS 2300-1110Medtronic Sofamor Danek, Inc. · AWL7 reports
- Endoskeleton™ TCS Interbody System 5210-1063Medtronic Sofamor Danek, Inc. · Awl5 reports
- NA 03.820.100Synthes GmbH · AWL5 reports
- SKYLINE 286820400Depuy Spine, LLC · AWL4 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page