Skip to content

NA 03.010.041Recalls & reported problems

Synthes GmbH · 14.0MM CANNULATED AWL · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886982067142), received between Dec 21, 2016 and Apr 10, 2026 — classified as 3 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Entrapment of Device1
  • Material Fragmentation1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1
  • Injury1
  • Insufficient Information1
  • No Code Available1
  • No Patient Involvement1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 10, 2026Break

    IT WAS REPORTED ON (B)(6) 2026 THAT THE AWL HAS BROKEN TIP, AND THE INSERTION INSTRUMENT WAS BENT. NO OPERATION INVOLVED. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24844748 ↗
  • MalfunctionApr 8, 2024Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2024, AT THE TIME OF MAKING THE INITIAL CHANNEL TO THE INTRAMEDULLARY AREA, THE TIP OF THE PUNCH WAS BROKEN, BUT IT WAS MANAGED AND THE SURGERY WAS PERFORMED WITHOUT ANY PROBLEM. THIS REPORT IS FOR ONE AWL Ø14/3.2 CANN. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 19060092 ↗
  • MalfunctionOct 10, 2022Material Integrity Problem; Material Integrity Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN PORTUGAL AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT - 03.010.041 B 14.0/3.2 MM PUNZÓN, DAMAGED CANULATED. 03.010.063 HYSTIC PROTECTION PLANE 12.0/8.0, LONGITUD 188 MM, AND A 03.010.065 BIT GUIDE 8.0/4.2, FOR PART NO. 03.010.063, VERY DIFFICULT TO REMOVE FROM EACH OTHER. THERE WAS NO SURGICAL DELAY. THERE WAS ALTERATION OF THE TECHNIQUE BY ANOTHER SURGICAL OPTION EXISTING IN THE SURGICAL INSTRUMENT, WITH NO IMPACT ON THE SURGERY. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR A 14.0MM CANNULATED AWL THIS IS REPORT 3 OF 3 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15573464 ↗
  • InjurySep 29, 2020Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT INTRAMEDULLARY NAIL, DURING THE PROCEDURE, A PIECE OF THE CANNULATED AWL BROKE OFF IN THE GREATER TROCHANTER. FRAGMENTS WERE GENERATED. THE FRAGMENT WAS NOT ABLE TO BE REMOVED AND REMAINED IN THE PATIENT. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. PATIENT STATUS IS UNKNOWN. THIS REPORT INVOLVES ONE (1) 14.0MM CANNULATED AWL. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Foreign Body In Patient; Injury

    MAUDE report 10599447 ↗
  • InjuryMar 6, 2020Break; Break

    IT WAS REPORTED THAT DURING AN INTRAMEDULLARY (IM) NAILING OF THE FEMUR FOR ROTATIONAL OSTEOTOMY OF THE FEMUR ON (B)(6) 2020, WHILE GETTING THE STARTING POINT IN THE FEMUR FOR IM NAIL, THE TIP OF THE CANNULATED AWL BROKE OFF. IT TOOK ABOUT 30-40 MINUTES TO RETRIEVED THE BROKEN PIECE OUT OF THE PATIENT AS THE SURGEON HAD TO MAKE A LARGER INCISION IN ORDER TO ACCESS THE PIRIFORMIS FOSSA WHERE THE FRAGMENT WAS LODGE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THE USE OF AN OPENING REAMER. PATIENT STATUS WAS UNKNOWN. CONCOMITANT DEVICE REPORTED: UNKNOWN GUIDEWIRE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) 14.0MM CANNULATED AWL. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Code Available

    MAUDE report 9799810 ↗
  • InjurySep 28, 2017Break; Material Fragmentation; Break; Material Fragmentation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A FEMORAL NAILING DUE TO A FOOTBALL INJURY. UPON OPENING THE PATIENT'S PROXIMAL FEMUR, IT WAS NOTED THAT HE HAD EXTREMELY HARD BONE SO THE SURGEON USED A 14.0 MM CANNULATED AWL AFTER THE OPENING REAMER WOULDN'T ADVANCE. AFTER SINKING THE 14.0 MM CANNULATED AWL IT WAS NOTED THAT A PIECE OF THE AWL BROKE OFF IN THE BONE. THE SURGEON DECIDED NOT TO RETRIEVE THE PIECE AND THE SURGERY WASN'T DELAYED. THEN, WHEN LOCKING THE NAIL PROXIMALLY, THE DRILL SLEEVES DIDN'T PASS THROUGH THE HANDLE EASILY, AND THEY WERE DIFFICULT TO REMOVE. THERE WASN'T A CONSIDERABLE DELAY DUE TO THIS, BUT THE SURGEON MENTIONED HOW DIFFICULT IT WAS TO PASS THE SLEEVES. CONCOMITANT DEVICE REPORTED: DRILL SLEEVES, UNKNOWN NAIL. THIS REPORT IS FOR ONE (1) 14.0 MM CANNULATED AWL. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 6901821 ↗
  • MalfunctionDec 21, 2016Break; Break

    IT WAS REPORTED THAT AN AWL WAS DISCOVERED BROKEN IN A STORAGE WAREHOUSE OUTSIDE OF HOSPITAL. NO PATIENT INVOLVEMENT WAS REPORTED. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 6193328 ↗

Other AWL devices

All AWL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page