ECHELON FLEX ENDOPATH EC60ARecalls & reported problems
Ethicon Endo-Surgery, LLC · ECHELON Flex 60 Articulating Endoscopic Linear Cutter (No Cartridge) · Class 2
All models of ECHELON Flex ENDOPATH (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Echelon Flex endoscopic stapler has 1,451 FDA adverse event reports, with 157 injuries, 1,286 malfunctions and 8 deaths. The reports mostly describe the stapler being hard to open or close, failing to fire, and not forming staples, with some mentioning bleeding. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,448
- Deaths reported
- 8
- Injuries reported
- 159
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,448 adverse event reports that list this exact device (DI 10705036001706), received between Feb 15, 2013 and Sep 18, 2026 — classified as 8 deaths, 159 injuries, 1,281 malfunctions.
The most frequently reported specific device problem is “Difficult to Open or Close” (510 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Open or Close510
- Failure to Fire471
- Failure to Form Staple309
- Mechanical Jam101
- Adverse Event Without Identified Device or Use Problem71
- Mechanical Problem65
- No Apparent Adverse Event61
- Firing Problem42
Patient problems
- No Clinical Signs, Symptoms or Conditions801
- No Consequences Or Impact To Patient476
- No Code Available65
- Hemorrhage/Blood Loss/Bleeding56
- Failure to Anastomose35
- Blood Loss31
- Insufficient Information30
- Unspecified Tissue Injury26
Recent adverse event reports
The 25 most recent of 1,448 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 18, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50829278 ↗InjurySep 2, 2026Adverse Event Without Identified Device or Use Problem
It was reported that post op to a hemicolectomy (sigmoid) the patient had repeated bleeding since. She was undergoing a rectosigmoidoscopy today to find out whether or not the patient needed to have another operation.Has there been any harm reported to the patient or user? YesDid the patient/user require medical or surgical intervention as a result of the event? NoDid the patient or user require hospitalization for an extended period of time as a result of the event? YesDescription: prolongation for bleeding and rectosigmoidoscopyWhat is the patient/user's status after the event? Waiting for additional information.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50572052 ↗MalfunctionAug 31, 2026Delivered as Unsterile Product
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50520037 ↗MalfunctionAug 28, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50502065 ↗MalfunctionAug 28, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50502051 ↗MalfunctionAug 14, 2026Failure to Form Staple; Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50256259 ↗MalfunctionAug 10, 2026Insufficient Device Problem Information
It was reported that during an unknown procedure, it was observed that the stapler misfired. Changed to a new one to complete the procedure. There was no patient consequence nor surgical delay reported. No additional information provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50195193 ↗InjuryAug 4, 2026Adverse Event Without Identified Device or Use Problem
It was reported that postoperatively to a Superior vena cava invasion from lung cancer, the surgery was completed without any abnormity. After chest closure, the patient experienced a drop in blood pressure in the waking room. After opening the chest, the staples anastomosed on the superior vena cava was found to have fallen apart. Suture was used to control bleeding. The patient has not waken up so far. Additional information provided.
Patient problems: Coma; Failure to Anastomose; Hemorrhage/Blood Loss/Bleeding; Low Blood Pressure/ Hypotension
MAUDE report 50127771 ↗MalfunctionAug 3, 2026Failure to Fire
It was reported that during an unknown procedure, the first firing was completed successfully. During the second firing, while closing the common enterotomy, the firing trigger felt light, and firing was completed on the fifth grasping. However, upon checking the surgical field, it was found that no staples had been formed, and the device had not fired properly. The doctor switched to manual suturing to complete the case. The cartridge color was White. There were no adverse consequences to the patient. No additional information provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50115079 ↗MalfunctionJul 17, 2026Difficult to Open or Close
IT WAS REPORTED THAT PREOPERATIVELY TO AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THEY COULD NOT CLOSE THE DEVICE BUT THEN WHEN THEY FINALLY COULD CLOSE IT, THEY COULD NOT OPEN IT AGAIN. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25860425 ↗InjuryJul 1, 2026Failure to Form Staple; Failure to Fire; Difficult to Open or Close; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Noise, Audible
IT WAS REPORTED THAT DURING A LEFT UPPER LOBECTOMY PROCEDURE, THE PATIENT¿S LOBE WAS DESCRIBED AS THICK, PREVENTING THE STAPLER FROM CLOSING INITIALLY. THE SURGEON SUBSEQUENTLY FORCED THE DEVICE TO CLOSE AND HELD COMPRESSION FOR TWO MINUTES. UPON FIRING, A LOUD POP WAS HEARD, AND THE KNIFE DID NOT ADVANCE FURTHER. TWO PLEE60A DEVICES REQUIRED MANUAL OVERRIDE TO OPEN. ADDITIONALLY, BOTH PLEE60A DEVICES RAN OUT OF BATTERY CHARGE DURING FIRING, AND THE ANVIL ON THE ECH45L AND ECH60 DEVICES WAS NOTED TO BE BENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 25681507 ↗MalfunctionJun 24, 2026Failure to Fire
IT WAS REPORTED THAT DURING A RADICAL RESECTION OF PANCREATIC CANCER, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25613177 ↗MalfunctionJun 19, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25566412 ↗MalfunctionJun 17, 2026Failure to Fire
IT WAS REPORTED THAT DURING A LAPAROSCOPIC HEPATECTOMY PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT BE FIRED. CHANGED TO ANOTHER ONE TO CONTINUE THE PROCEDURE BUT THE SAME PROBLEM HAPPENED AGAIN. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25537620 ↗MalfunctionJun 15, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING A LAP GASTRECTOMY PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. THEY USED SUTURE TO OVERSEW. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25511072 ↗MalfunctionJun 12, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25488974 ↗MalfunctionJun 9, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25447774 ↗MalfunctionJun 9, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25445921 ↗MalfunctionJun 9, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE STAPLER GOT STUCK WHILE FIRING. AS A RESULT, THE SURGEON HAD TO USE A NEW STAPLER TO COMPLETE THE CASE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25444447 ↗MalfunctionJun 8, 2026Failure to Cut
IT WAS REPORTED THAT DURING AN ENDOSCOPIC PANCREATIC PROCEDURE, IT WAS OBSERVED THAT THE TISSUE WAS NOT CUT AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25428230 ↗MalfunctionJun 1, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25349692 ↗MalfunctionJun 1, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25349430 ↗MalfunctionMay 29, 2026Failure to Fire; Difficult to Open or Close; No Apparent Adverse Event
IT WAS REPORTED THAT DURING A LAPAROSCOPIC RADICAL GASTRECTOMY PROCEDURE; IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE WHEN SEVERING DUODENAL TISSUE ON THE FIRST FIRING. IT WAS FURTHER REPORTED THAT THEY COULD NOT OPEN THE JAW. AS THE TISSUE WAS CUT OFF, THEY REMOVED THE DEVICE FROM THE PATIENT THROUGH TISSUE GAP. CHANGED ANOTHER ONE TO USE, STILL COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25337509 ↗MalfunctionMay 26, 2026Failure to Form Staple; No Apparent Adverse Event
IT WAS REPORTED THAT AFTER THE FIRST FIRING IN A RIGHT HEMICOLECTOMY, AN ACTIVE BLEEDING OCCURRED (THIS HAS HAPPENED MORE OFTEN RECENTLY). THEY WAITED >15 SECONDS BEFORE FIRING, AND IT WAS OBSERVED THAT THE STAPLES WEREN'T PROPERLY FORMED. THEY HAD TO PLACE A CLIP TO STOP THE BLEEDING. THE RELOAD IS DISPOSED, UNFORTUNATELY. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? UNKNOWN, WAS PROCEDURE SUCCESSFULLY COMPLETED? UNKNOWN, WERE FRAGMENTS GENERATED? UNKNOWN, IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? UNKNOWN, PATIENT STATUS/ OUTCOME / CONSEQUENCES NO, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN,
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 25279626 ↗MalfunctionMay 19, 2026Mechanical Jam
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE KNIFE MOVED TO THE DISTAL END AFTER FIRING BUT DID NOT RETURN KNIFE AUTOMATICALLY. THEY USED REVERSE BUTTON TO RETURN THE KNIFE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25216731 ↗
FDA clearance and classification
The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the ECHELON FLEX ENDOPATH EC60A
Has the ECHELON FLEX ENDOPATH EC60A been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705036001706), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ECHELON FLEX ENDOPATH EC60A?
The FDA's MAUDE database lists 1,448 adverse event reports for this device, received between Feb 15, 2013 and Sep 18, 2026. In those reports, the most common specific device problems are “Difficult to Open or Close” (510), “Failure to Fire” (471) and “Failure to Form Staple” (309); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (56), “Failure to Anastomose” (35) and “Blood Loss” (31). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ECHELON FLEX ENDOPATH EC60A?
8 reports are classified as deaths, 159 as injuries and 1,281 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the ECHELON FLEX ENDOPATH EC60A, and is it still sold?
The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand ECHELON FLEX ENDOPATH, model EC60A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,660 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,727 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,350 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,279 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,006 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,967 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,910 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,044 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page