Skip to content

ECHELON; ENDOPATH ECR45GRecalls & reported problems

Ethicon Endo-Surgery, LLC · Endoscopic Linear Cutter Reloads Green 4.1 mm · Class 2 · Implant

All models of ECHELON ENDOPATH (31)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Endopath Echelon stapler reload has 109 malfunction reports, with no deaths and 3 injuries. Many reports describe staples that were malformed after firing or did not form, and some describe the reload failing to fire, with staff switching to a new reload. Reports fell to 5 in the last 12 months from 16 in the 12 months before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
108
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 108 adverse event reports that list this exact device (DI 10705036001744), received between Jun 19, 2018 and Apr 11, 2026 — classified as 3 injuries, 105 malfunctions.

The most frequently reported specific device problem is “Failure to Form Staple” (67 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 2 reportsApr 2024: 4 reportsMay 2024: 0 reportsJun 2024: 4 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 4 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 4 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 3 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Form Staple67
  • Failure to Fire15
  • Device Markings/Labelling Problem7
  • Separation Problem5
  • Ejection Problem3
  • No Apparent Adverse Event3
  • Structural Problem3
  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions83
  • No Consequences Or Impact To Patient20
  • Hemorrhage/Blood Loss/Bleeding2
  • No Known Impact Or Consequence To Patient2
  • Pneumothorax2
  • Adhesion(s)1
  • Failure to Anastomose1
  • Foreign Body In Patient1

Recent adverse event reports

The 25 most recent of 108 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 11, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24847360 ↗
  • MalfunctionJan 26, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24177492 ↗
  • MalfunctionDec 15, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23803695 ↗
  • MalfunctionNov 13, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23546739 ↗
  • MalfunctionOct 15, 2025Structural Problem; Structural Problem

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THEY OPENED THE PACKING AND NOTED THAT SOME STAPLES FELL OFF. CHANGED TO A NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23290592 ↗
  • MalfunctionJun 4, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING A WEDGE RESECTION PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT BE FIRED. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22144488 ↗
  • InjuryMay 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING THE POST-OPERATIVE COURSE FOLLOWING AN OPEN LOBECTOMY IN WHICH THE FISSURE HAD BEEN COMPLETED WITH ECHELON 3000 AND BLUE RELOADS THE PATIENT HAD AN AIR LEAK. LONG HOSPITAL STAY.

    Patient problems: Failure to Anastomose; Hemorrhage/Blood Loss/Bleeding; Pneumothorax

    MAUDE report 21953420 ↗
  • InjuryMay 1, 2025Failure to Form Staple; Adverse Event Without Identified Device or Use Problem; Failure to Form Staple; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON THE AFTERNOON OF THE SECOND DAY AFTER SURGERY, THE PATIENT COMPLAINED OF BUBBLE OVERFLOW IN WATER-SEALED BOTTLE WHEN COUGHING, WHICH WAS CONSIDERED TO BE LUNG WOUND LEAKAGE, AND THE AIR LEAKAGE WAS AGGRAVATED GRADUALLY, WHICH WAS CONSIDERED TO BE POOR NAILING OF OCCLUDER. THE SYMPTOMS WERE IMPROVED AFTER INTRATHORACIC INFUSION OF HIGH GLUCOSE, AND THERE WAS STILL A SMALL AMOUNT OF AIR LEAKAGE. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: Pneumothorax

    MAUDE report 21945997 ↗
  • MalfunctionMay 1, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT AFTER FIRING THE DEVICE, THE STAPLES WERE MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21945187 ↗
  • MalfunctionApr 18, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21868200 ↗
  • MalfunctionMar 3, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21505281 ↗
  • MalfunctionMar 3, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21505144 ↗
  • MalfunctionFeb 10, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21346469 ↗
  • MalfunctionJan 27, 2025Failure to Form Staple; No Apparent Adverse Event; Failure to Form Staple; No Apparent Adverse Event

    IT WAS REPORTED THAT DURING A PNEUMONECTOMY PROCEDURE, IT WAS OBSERVED THAT SOME STAPLES MALFORMED AND OOZING OCCURRED AFTER FIRED. SUTURE WAS USED TO OVERSEW. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21241821 ↗
  • MalfunctionJan 22, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING A WEDGE RESECTION SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21208603 ↗
  • MalfunctionJan 21, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21197638 ↗
  • MalfunctionJan 21, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING A PANCREAS PROCEDURE SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21197497 ↗
  • MalfunctionDec 30, 2024Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. USED SUTURE TO OVERSEW. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21037381 ↗
  • MalfunctionNov 8, 2024Separation Problem; Separation Problem

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, NOTED THAT SOME SLED FELL OFF. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20643030 ↗
  • MalfunctionNov 6, 2024Separation Problem; Separation Problem

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, NOTED THE CARTRIDGE PAN SEPARATION. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20622608 ↗
  • MalfunctionOct 8, 2024Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20395982 ↗
  • MalfunctionSep 26, 2024Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING A LUNG RESECTION SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20312427 ↗
  • MalfunctionSep 6, 2024Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING THE RECTUM PROCEDURE, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED ANOTHER TWO TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. USED SUTURE TO OVERSEW. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20156029 ↗
  • MalfunctionSep 6, 2024Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING THE RECTUM PROCEDURE SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED ANOTHER TWO TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. USED SUTURE TO OVERSEW. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20156003 ↗
  • MalfunctionSep 6, 2024Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING THE RECTUM PROCEDURE, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED ANOTHER TWO TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. USED SUTURE TO OVERSEW. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20155987 ↗

FDA clearance and classification

The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the ECHELON; ENDOPATH ECR45G

Has the ECHELON; ENDOPATH ECR45G been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705036001744), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the ECHELON; ENDOPATH ECR45G?

The FDA's MAUDE database lists 108 adverse event reports for this device, received between Jun 19, 2018 and Apr 11, 2026. In those reports, the most common specific device problems are “Failure to Form Staple” (67), “Failure to Fire” (15) and “Device Markings/Labelling Problem” (7); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (2), “Pneumothorax” (2) and “Adhesion(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ECHELON; ENDOPATH ECR45G?

None of the reports are classified as deaths, 3 as injuries and 105 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ECHELON; ENDOPATH ECR45G MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the ECHELON; ENDOPATH ECR45G, and is it still sold?

The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand ECHELON; ENDOPATH, model ECR45G). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other STAPLE, IMPLANTABLE devices

All STAPLE, IMPLANTABLE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page