Ethicon Endo-Surgery, LLC · Endoscopic Linear Cutter Reloads White 2.5 mm · Class 2 · Implant
All models of ECHELON ENDOPATH (31)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Echelon Endopath stapler reload has 118 adverse event reports, with 14 injuries and no deaths. Many describe staples that were not formed correctly, and some describe the reload not firing. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 117
- Deaths reported
- 0
- Injuries reported
- 13
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 117 adverse event reports that list this exact device (DI 10705036001768), received between Dec 1, 2017 and Mar 23, 2026 — classified as 13 injuries, 104 malfunctions.
The most frequently reported specific device problem is “Failure to Form Staple” (69 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Form Staple69
- Failure to Fire22
- Adverse Event Without Identified Device or Use Problem7
- No Apparent Adverse Event6
- Separation Problem4
- Delivered as Unsterile Product3
- Material Integrity Problem3
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions73
- No Consequences Or Impact To Patient33
- No Code Available7
- Blood Loss6
- Hemorrhage/Blood Loss/Bleeding5
- Fistula2
- Failure to Anastomose1
- Hypovolemic Shock1
Recent adverse event reports
The 25 most recent of 117 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 23, 2026Separation Problem
IT WAS REPORTED THAT PREOPERATIVELY TO A LAPAROSCOPIC LOBECTOMY, IT WAS OBSERVED THAT THERE WAS CARTRIDGE PAN SEPARATION IN THE PACKAGING OF THE DEVICES. THERE WERE FOUR DEVICES WITH THE SAME ISSUE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24663823 ↗MalfunctionMar 23, 2026Separation Problem
IT WAS REPORTED THAT PREOPERATIVELY TO A LAPAROSCOPIC LOBECTOMY, IT WAS OBSERVED THAT THERE WAS CARTRIDGE PAN SEPARATION IN THE PACKAGING OF THE DEVICES. THERE WERE FOUR DEVICES WITH THE SAME ISSUE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24663806 ↗MalfunctionMar 23, 2026Separation Problem
IT WAS REPORTED THAT PREOPERATIVELY TO A LAPAROSCOPIC LOBECTOMY, IT WAS OBSERVED THAT THERE WAS CARTRIDGE PAN SEPARATION IN THE PACKAGING OF THE DEVICES. THERE WERE FOUR DEVICES WITH THE SAME ISSUE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24663803 ↗MalfunctionJan 6, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING A LUNG RESECTION PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23977241 ↗MalfunctionJan 2, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING A LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE CLOSED STAPLES UNRAVELED WHEN VESSEL WAS CLOSED INTRAOPERATIVELY. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23949789 ↗MalfunctionDec 15, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23804492 ↗MalfunctionDec 15, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23804440 ↗MalfunctionDec 15, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23803621 ↗MalfunctionNov 4, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23466396 ↗MalfunctionOct 30, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE TISSUE WAS NOT COMPLETELY CLOSED DUE TO THE FAULTY B FORMATION. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23426421 ↗InjuryOct 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT POST OP TO A LAPAROSCOPIC LOBECTOMY THORACIC PROCEDURE, BLEEDING ON THE SECOND POSTOPERATIVE DAY. EMERGENCY THORACOSCOPIC EXPLORATION, THORACIC BLOOD CLOT DEBRIDEMENT, HEMOSTASIS. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 23403911 ↗MalfunctionOct 27, 2025Failure to Form Staple; Material Integrity Problem; No Apparent Adverse Event; Failure to Form Staple; Material Integrity Problem; No Apparent Adverse Event
IT WAS REPORTED THAT DURING A LOBECTOMY, IT WAS OBSERVED THAT THE CARTRIDGE HAD A BULGE IN THE MIDDLE PORTION; AND THERE WAS BLEEDING FROM THE CARTRIDGE FACE. THE PHYSICIAN IMMEDIATELY STOPPED THE BLEEDING WITH A VASCULAR CLIP. THERE WAS AN UNSPECIFIED INCREASED IN THE PROCEDURE TIME. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 23393019 ↗MalfunctionSep 5, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO ANOTHER ONE TO CONTINUE THE PROCEDURE BUT THE SAME PROBLEM HAPPENED AGAIN. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22980115 ↗MalfunctionAug 15, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22815133 ↗MalfunctionJul 10, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. THEY CHANGED TO ANOTHER ONE TO CONTINUE THE PROCEDURE BUT THE SAME PROBLEM HAPPENED AGAIN. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22471017 ↗MalfunctionJul 10, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. THEY CHANGED TO ANOTHER ONE TO CONTINUE THE PROCEDURE BUT THE SAME PROBLEM HAPPENED AGAIN. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22471014 ↗MalfunctionJul 10, 2025Failure to Form Staple; No Apparent Adverse Event; Failure to Form Staple; No Apparent Adverse Event
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE WAS USED TO DISCONNECT THE LEFT INFERIOR PULMONARY VEIN. AFTER DISCONNECTING THE LEFT INFERIOR PULMONARY VEIN, IT WAS FOUND THAT THE NAIL EDGE AT THE DISTAL END OF THE VEIN WAS NOT FULLY CLOSED, RESULTING IN BLEEDING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. ADDITIONAL INFORMATION REQUESTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 22470941 ↗MalfunctionDec 30, 2024Failure to Form Staple; Failure to Form Staple
THIS CASE IS FROM HEALTH AUTHORITY. IT WAS REPORTED THAT DURING A LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE STAPLES MALFORMED AFTER FIRING. SUTURE WAS USED TO OVERSEW. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21036946 ↗MalfunctionNov 7, 2024Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, COULD NOT FIRE FARTHER AFTER FIRED A LITTLE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20632730 ↗InjurySep 18, 2024Failure to Fire; Adverse Event Without Identified Device or Use Problem; Failure to Fire; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE LEFT UPPER LOBECTOMY, AFTER A FIRE, NOTED THE STAPLES ONLY DEPLOYED ON ONE SIDE OF THE RELOAD, WHICH CAUSED MASSIVE BLEEDING. CHANGED TO OPEN OPERATION TO COMPLETE THE SURGERY. THE SURGERY WAS PROLONGED FOR ABOUT 1 HOUR.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 20250793 ↗MalfunctionJun 26, 2024Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19615093 ↗MalfunctionJun 25, 2024Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING PROCEDURE, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19606459 ↗MalfunctionJun 24, 2024Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19593870 ↗MalfunctionMay 8, 2024Difficult or Delayed Separation; Difficult or Delayed Separation
IT WAS REPORTED THAT THE RELOAD WAS USED WITH THE STAPLER, BUT AS THE SCRUB NURSE TRIED TO LOAD THE RELOAD, THE BLUE PLASTIC PIECE AT POSTERIOR END OF RELOAD, FELL OUT. SHE SUBSEQUENTLY ASKED FOR A NEW RELOAD TO BE OPENED AS SHE WAS NOT COMFORTABLE PUSHING BACK THE LITTLE BLUE PIECE. NO PATIENT CONSEQUENCES REPORTED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19272086 ↗MalfunctionMay 3, 2024Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING THE SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19242474 ↗
FDA clearance and classification
The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the ECHELON; ENDOPATH ECR45W
Has the ECHELON; ENDOPATH ECR45W been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705036001768), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ECHELON; ENDOPATH ECR45W?
The FDA's MAUDE database lists 117 adverse event reports for this device, received between Dec 1, 2017 and Mar 23, 2026. In those reports, the most common specific device problems are “Failure to Form Staple” (69), “Failure to Fire” (22) and “Separation Problem” (4); the most common patient problems are “Blood Loss” (6), “Hemorrhage/Blood Loss/Bleeding” (5) and “Fistula” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ECHELON; ENDOPATH ECR45W?
None of the reports are classified as deaths, 13 as injuries and 104 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ECHELON; ENDOPATH ECR45W MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the ECHELON; ENDOPATH ECR45W, and is it still sold?
The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand ECHELON; ENDOPATH, model ECR45W). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,660 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,727 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,350 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,279 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,006 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,967 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,910 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,044 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page