ECHELON; ENDOPATH ECR60BRecalls & reported problems
Ethicon Endo-Surgery, LLC · Endoscopic Linear Cutter Reloads Blue 3.5mm · Class 2 · Implant
All models of ECHELON ENDOPATH (31)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Echelon stapler cartridge from Ethicon holds staples used in minimally invasive surgery to join and cut tissue. The FDA database holds 545 reports for it, most describing staples that do not form properly or fail to fire, with 60 injuries and 1 death. No recalls name this device. Reports fell from 73 in the previous 12 months to 8 in the last 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 545
- Deaths reported
- 1
- Injuries reported
- 60
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 545 adverse event reports that list this exact device (DI 10705036001775), received between Dec 14, 2017 and Mar 13, 2026 — classified as 1 death, 60 injuries, 484 malfunctions.
The most frequently reported specific device problem is “Failure to Form Staple” (283 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Form Staple283
- Failure to Fire105
- Adverse Event Without Identified Device or Use Problem33
- No Apparent Adverse Event27
- Material Integrity Problem22
- Manufacturing, Packaging or Shipping Problem14
- Device Misassembled During Manufacturing /Shipping12
- Ejection Problem12
Patient problems
- No Clinical Signs, Symptoms or Conditions357
- No Consequences Or Impact To Patient133
- Hemorrhage/Blood Loss/Bleeding35
- No Code Available25
- Failure to Anastomose23
- Blood Loss15
- Fistula6
- Insufficient Information6
Recent adverse event reports
The 25 most recent of 545 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 13, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE FIRING SEQUENCE SUDDENLY STOPPED; AND THE REVERSE BUTTON WAS USED. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24591537 ↗MalfunctionMar 3, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24500626 ↗MalfunctionFeb 11, 2026Device Misassembled During Manufacturing /Shipping
IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, BEFORE USED ON THE PATIENT, NOTED THERE ARE FOREIGN MATTERS IN THE PACKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24316038 ↗MalfunctionFeb 11, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24314618 ↗MalfunctionJan 9, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24022647 ↗MalfunctionDec 21, 2025Failure to Form Staple; No Apparent Adverse Event; Failure to Form Staple; No Apparent Adverse Event
IT WAS REPORTED THAT DURING A LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT SOME STAPLES MALFORMED AND OOZING OCCURRED. SUTURE WAS USED TO OVERSEW. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 23860607 ↗MalfunctionNov 13, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LAP GASTRIC SLEEVE PROCEDURE, IT WAS OBSERVED THAT THE RELOAD DID NOT LEAD TO A B-FORMED SHAPE AS 4-5 PINS LOOKED OPEN ONCE FIRING COMPLETED. THE SURGEON DID OMENTOPEXY FOR SAFETY AND COMPLETED THE CASE. THERE WAS NO BLEEDING OR OOZING REPORTED. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23542232 ↗MalfunctionNov 13, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LAP GASTRIC SLEEVE PROCEDURE, IT WAS OBSERVED THAT THE RELOAD DID NOT LEAD TO A B-FORMED SHAPE AS 4-5 PINS LOOKED OPEN ONCE FIRING COMPLETED. THE SURGEON DID OMENTOPEXY FOR SAFETY AND COMPLETED THE CASE. THERE WAS NO BLEEDING OR OOZING REPORTED. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23542205 ↗MalfunctionOct 31, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED ON THE THIRD, FOURTH AND FIFTH FIRING. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23438966 ↗MalfunctionOct 31, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED ON THE THIRD, FOURTH AND FIFTH FIRING. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23438818 ↗MalfunctionOct 31, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED ON THE THIRD, FOURTH AND FIFTH FIRING. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23438814 ↗MalfunctionOct 29, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO ANOTHER THREE TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23417427 ↗MalfunctionOct 29, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO ANOTHER THREE TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23417426 ↗MalfunctionOct 29, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO ANOTHER THREE TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23417421 ↗MalfunctionOct 29, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO ANOTHER THREE TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23417418 ↗MalfunctionOct 27, 2025Ejection Problem; Ejection Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE OF THE ESOPHAGUS, THE SURGEON NOTED THAT IT WAS VERY DIFFICULT TO INSTALL THE CARTRIDGE AFTER OPENING THE PACKAGE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23397198 ↗MalfunctionSep 10, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
THE DEVICE WAS RECEIVED WITH NO COMPLAINT INFORMATION. ATTEMPTS WERE MADE TO RECONCILE THE DEVICE WITH NO SUCCESS RESULTING IN THE CREATION OF A NEW COMPLAINT. AWB 391120895591 RECEIVED 07/28/2025. BOX MARKED AS (B)(4) ON AWB 391120895591; RECD ECR60B; LOT#: 965A82; ECM HAS GST60B; LOT#: 887C43 FROM COUNTRY CHINA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23014873 ↗MalfunctionSep 8, 2025Failure to Fire; Failure to Fire
THE CARTRIDGE WAS NOT USED BECAUSE IT DID NOT COVER THE TISSUE. IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT BE FIRED. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22992443 ↗MalfunctionAug 20, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A COLON RESECTION PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22850002 ↗MalfunctionAug 18, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LUNG RESECTION PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22833414 ↗MalfunctionAug 18, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LUNG RESECTION PROCEDURE, IT WAS OBSERVED THAT SOME STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22833151 ↗MalfunctionAug 18, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LUNG RESECTION PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22828577 ↗MalfunctionAug 18, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT, DURING A LUNG RESECTION PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22827086 ↗MalfunctionAug 18, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LUNG RESECTION PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22826927 ↗MalfunctionAug 16, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. THEY CHANGED TO ANOTHER TWO RELOADS TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22823190 ↗
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,680 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,730 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,356 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,294 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,011 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,965 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,909 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,046 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page