ECHELON ENDOPATH SC60Recalls & reported problems
Ethicon Endo-Surgery, LLC · Echelon 60 Endopath Stapler Compact Endoscopic Linear Cutter-Straight · Class 2
All models of ECHELON ENDOPATH (31)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10705036002703), received between Nov 22, 2017 and Oct 24, 2018 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Failure to Form Staple” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Failure to Form Staple2
- Device Inoperable1
- Difficult to Open or Close1
- Failure to Fire1
- Retraction Problem1
Patient problems
- No Consequences Or Impact To Patient4
- No Code Available1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 24, 2018Retraction Problem
IT WAS REPORTED THAT COULD NOT RETURN BLADE. DURING THE ENDOSCOPIC PULMONARY BULLAE, AFTER FIRED, COULD NOT RETURN THE BLADE, USED MANUAL RELEASE LEVER TO RETURN THE BLADE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8000989 ↗MalfunctionOct 17, 2018Failure to Fire
IT WAS REPORTED THAT DURING AN OPEN SIGMOIDECTOMY, THE DEVICE COULD NOT BE FIRED AT THE FIRST FIRING ON THE COLON. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7975875 ↗MalfunctionApr 24, 2018Failure to Form Staple
IT WAS REPORTED THAT THE STAPLES MALFORMED AFTER FIRING. DURING THE VATS LOBECTOMY, AFTER FIRED WITH ECR60B, WHEN SEVERING THE LOBE OF LUNG, FOUND THE STAPLES MALFORMED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7454672 ↗MalfunctionDec 15, 2017Device Inoperable; Difficult to Open or Close
IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTRECTOMY, THE DEVICE WAS LOCKED OUT AT THE FIRST FIRING WHEN THE DEVICE WAS USED FOR ¿DUODENUM RESECTION. THEN, THE JAWS COULD NOT BE OPENED WITH THE ANVIL RELEASE BUTTON. EVENTUALLY, THE JAWS COULD BE OPENED WITH THE BUTTON. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7120473 ↗InjuryNov 22, 2017Failure to Form Staple
IT WAS REPORTED THAT THE STAPLES MALFORMED AFTER FIRING. DURING THE THORACOSCOPIC RESECTION OF BULLAE OF SUPERIOR LOBE OF RIGHT LUNG, AFTER 1ST FIRED WITH SC60 AND ECR60B, THE STAPLES MALFORMED WITH IRREGULAR SHAPE. CHANGED ANOTHER ECR60B, THE EVENT RE-HAPPENED. CHANGED EC60 WITH ECR60D, THE EVENT RE-HAPPENED. CHANGED EC60A WITH ECR60D, THE EVENT RE-HAPPENED. THE PATIENT LOST ABOUT 200ML BLOOD WITHOUT TRANSFUSION. THE SURGERY WAS CHANGED TO OPEN SURGERY, AND DELAYED ABOUT 30 MINUTES. USED TLC75 TO COMPLETE THE SURGERY. THE PATIENT IS STABLE AND IN HOSPITAL NOW.
Patient problems: No Code Available
MAUDE report 7053051 ↗
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,680 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,730 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,356 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,294 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,011 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,965 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,909 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,046 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗