N/A CDH29ARecalls & reported problems
Ethicon Endo-Surgery, LLC · Curved Intraluminal Stapler (ILS) 24 Titanium Adjustable Height Staples · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This is the Ethicon curved intraluminal stapler, a surgical stapler used inside the body to join tissue. The 1,621 reports often describe staples that do not form properly, the stapler failing to fire, and failure to cut. Some mention failed joins between tissue ends. The records include 17 deaths, 413 injuries and 1,191 malfunctions, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,612
- Deaths reported
- 17
- Injuries reported
- 413
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,612 adverse event reports that list this exact device (DI 10705036003465), received between Nov 27, 2017 and Sep 21, 2026 — classified as 17 deaths, 413 injuries, 1,182 malfunctions.
The most frequently reported specific device problem is “Failure to Form Staple” (612 reports).
13 other recalls cover devices cleared under the same FDA submission (K983536); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Form Staple612
- Failure to Fire385
- Adverse Event Without Identified Device or Use Problem181
- Failure to Cut166
- Firing Problem102
- Ejection Problem75
- Nonstandard Device73
- No Apparent Adverse Event58
Patient problems
- No Clinical Signs, Symptoms or Conditions651
- No Consequences Or Impact To Patient565
- No Code Available275
- Failure to Anastomose215
- Blood Loss42
- Hemorrhage/Blood Loss/Bleeding28
- Insufficient Information27
- Unspecified Infection18
Recent adverse event reports
The 25 most recent of 1,612 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2026Failure to Fire
It was reported that during and Unknown procedure, could not fire. Changed the new one to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50839346 ↗MalfunctionAug 26, 2026Appropriate Device Problem Term/Code Not Available
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50453845 ↗MalfunctionAug 26, 2026Appropriate Device Problem Term/Code Not Available
It was reported that during an Unknown procedure the staples came out of the clamp before use, requiring a change of clamp.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50453712 ↗MalfunctionAug 17, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50278248 ↗MalfunctionAug 13, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50235957 ↗MalfunctionAug 10, 2026Device Misassembled During Manufacturing /Shipping
It was reported that during an Unknown surgery, open the packing and noted that some staples fell off. Changed the new one to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50193953 ↗MalfunctionAug 10, 2026Device Misassembled During Manufacturing /Shipping
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50193505 ↗MalfunctionJul 7, 2026Ejection Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE WHEN PREPARING TO USE THE INSTRUMENTS, IT WAS FOUND THAT THE SHIITAKE MUSHROOM TIP AND THE PUNCTURE END DID NOT FIT PROPERLY, CAUSING THE DOCTOR TO BE UNEASY AND REPLACED THE INSTRUMENTS. HAVE THERE BEEN REPORTS OF INJURIES FROM PATIENTS OR USERS? NO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25735900 ↗MalfunctionJun 18, 2026Failure to Fire
IT WAS REPORTED THAT DURING THE RECTUM PROCEDURE SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25553316 ↗MalfunctionJun 16, 2026Failure to Fire
IT WAS REPORTED THAT DURING THE PROCTRECTOMY SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25523423 ↗MalfunctionJun 16, 2026Failure to Cut
IT WAS REPORTED THAT DURING THE PROCTRECTOMY SURGERY, AFTER FIRED, NOTED THAT THE TISSUE WAS NOT CUT . THEN CHANGED ANOTHER DEVICE TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25523333 ↗MalfunctionJun 15, 2026Failure to Fire; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING THE UNK SURGERY, THE NAILS ARE EXPOSED OUTSIDE THE WAREHOUSE AND COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25511577 ↗MalfunctionJun 15, 2026Failure to Fire
IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25510506 ↗MalfunctionJun 10, 2026Failure to Cut
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THAT THE TISSUE WAS NOT CUT, THEN CHANGED ANOTHER DEVICE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25460472 ↗MalfunctionJun 8, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, COULD NOT DRAW BACK SAFETY TRIGGER. COULD NOT DRAW BACK THE SAFETY TRIGGER DURING THE SURGERY. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25436924 ↗MalfunctionMay 27, 2026Ejection Problem
IT WAS REPORTED THAT DURING THE UNK SURGERY, OPEN THE PACKING AND NOTED ANVIL CANNOT BE ATTACHED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25302014 ↗MalfunctionMay 27, 2026Failure to Fire
IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25301794 ↗MalfunctionMay 26, 2026Failure to Form Staple; No Apparent Adverse Event
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED SOME STAPLES MALFORMED AND OOZING OCCURRED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 25284376 ↗MalfunctionMay 26, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25281088 ↗MalfunctionMay 23, 2026Failure to Fire
IT WAS REPORTED THAT DURING THE PROCTECTOMY PROCEDURE, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25264740 ↗MalfunctionMay 23, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, REMOVED THE DEVICE, NOTED THE DONUTS WERE INCOMPLETE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25264705 ↗MalfunctionMay 15, 2026Failure to Cut
IT WAS REPORTED THAT DURING THE LOW ANTERIOR RESECTION SURGERY, AFTER FIRED, NOTED THAT THE TISSUE WAS NOT CUT COMPLETELY, BUT THERE WERE SOME STAPLES DEPLOYED ON THE TISSUE. CHANGED ANOTHER DEVICE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25184232 ↗MalfunctionMay 14, 2026Failure to Fire
IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25177891 ↗MalfunctionMay 14, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25175048 ↗MalfunctionMay 14, 2026Failure to Fire
IT WAS REPORTED THAT DURING PROCTRECTOMY PROCEDURE, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25174291 ↗
FDA clearance and classification
Cleared through 510(k) K983536 on Dec 18, 1998 (decision: Substantially Equivalent). Submitted by Ethicon Endo-Surgery, Inc. as “Proximate Curved And Straight Intraluminal Staplers”. The FDA review took 70 days. FDA record ↗
The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K983536. One submission often covers many models, so check the product and lot details in each record.
- Z-1272-2019 · Apr 11, 2019 · CompletedEndoscopic Curved Intraluminal Stapler, 25 mm diameter, Model ECS25A The Ethicon Endo-Surgery Intraluminal Staplers (ILS) are anastomotic staplers available in four sizes to permi
- Z-1269-2019 · Apr 11, 2019 · CompletedCurved Intraluminal Staplers, 29 mm diameter, Model CDH29A The Ethicon Endo-Surgery Intraluminal Staplers (ILS) are anastomotic staplers available in four sizes to permit proper m
- Z-1268-2019 · Apr 11, 2019 · CompletedCurved Intraluminal Staplers, 25 mm diameter, Model CDH25A The Ethicon Endo-Surgery Intraluminal Staplers (ILS) are anastomotic staplers available in four sizes to permit proper m
- Z-1275-2019 · Apr 11, 2019 · CompletedStraight Intraluminal Staplers, 21 mm diameter, Model SDH21A The Ethicon Endo-Surgery Intraluminal Staplers (ILS) are anastomotic staplers available in four sizes to permit proper
- Z-1278-2019 · Apr 11, 2019 · CompletedStraight Intraluminal Staplers, 33 mm diameter, Model SDH33A The Ethicon Endo-Surgery Intraluminal Staplers (ILS) are anastomotic staplers available in four sizes to permit proper
- Z-1270-2019 · Apr 11, 2019 · CompletedCurved Intraluminal Staplers, 33 mm diameter, Model CDH33A The Ethicon Endo-Surgery Intraluminal Staplers (ILS) are anastomotic staplers available in four sizes to permit proper m
- Z-1274-2019 · Apr 11, 2019 · CompletedEndoscopic Curved Intraluminal Stapler, 33 mm diameter, Model ECS33A The Ethicon Endo-Surgery Intraluminal Staplers (ILS) are anastomotic staplers available in four sizes to permi
- Z-1276-2019 · Apr 11, 2019 · CompletedStraight Intraluminal Staplers, 25 mm diameter, Model SDH25A The Ethicon Endo-Surgery Intraluminal Staplers (ILS) are anastomotic staplers available in four sizes to permit proper
- Z-1277-2019 · Apr 11, 2019 · CompletedStraight Intraluminal Staplers, 29 mm diameter, Model SDH29A The Ethicon Endo-Surgery Intraluminal Staplers (ILS) are anastomotic staplers available in four sizes to permit proper
- Z-1271-2019 · Apr 11, 2019 · CompletedEndoscopic Curved Intraluminal Stapler, 21 mm diameter, Model ECS21A The Ethicon Endo-Surgery Intraluminal Staplers (ILS) are anastomotic staplers available in four sizes to permi
- Z-1273-2019 · Apr 11, 2019 · CompletedEndoscopic Curved Intraluminal Stapler, 29 mm diameter, Model ECS29A The Ethicon Endo-Surgery Intraluminal Staplers (ILS) are anastomotic staplers available in four sizes to permi
- Z-1267-2019 · Apr 11, 2019 · CompletedCurved Intraluminal Staplers, 21 mm diameter, Model CDH21A The Ethicon Endo-Surgery Intraluminal Staplers (ILS) are anastomotic staplers available in four sizes to permit proper m
- Z-0672-2018 · Oct 23, 2017 · TerminatedDisposable Curved Circular Staplers Product Usage: Staplers have application throughout the alimentary tract for end-to-end, end-to-side, and side-to-side anastomoses.
Common questions about the N/A CDH29A
Has the N/A CDH29A been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705036003465), based on the FDA recall file dated Nov 27, 2026. However, 13 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the N/A CDH29A?
The FDA's MAUDE database lists 1,612 adverse event reports for this device, received between Nov 27, 2017 and Sep 21, 2026. In those reports, the most common specific device problems are “Failure to Form Staple” (612), “Failure to Fire” (385) and “Failure to Cut” (166); the most common patient problems are “Failure to Anastomose” (215), “Blood Loss” (42) and “Hemorrhage/Blood Loss/Bleeding” (28). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the N/A CDH29A?
17 reports are classified as deaths, 413 as injuries and 1,182 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the N/A CDH29A FDA approved?
It was cleared through 510(k) K983536 on Dec 18, 1998. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the N/A CDH29A MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the N/A CDH29A, and is it still sold?
The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand N/A, model CDH29A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,660 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,727 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,350 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,279 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,006 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,967 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,910 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,044 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.