N/A ECS33ARecalls & reported problems
Ethicon Endo-Surgery, LLC · Endoscopic Curved Intraluminal Stapler (ILS) 28 Titanium Adjustable Height Staples · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon endoscopic curved stapler (ECS33A) is used to join tissue inside the body during surgery. Most of the 52 reports describe the stapler failing mid-procedure — staples that would not form or fire, staples coming out misshapen, or a loose anvil that would not attach — so surgeons swapped in a new device to finish the operation. The reports include 1 death and 14 injuries, with problems such as a connection that failed to hold and a postoperative leak. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 53
- Deaths reported
- 1
- Injuries reported
- 14
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 53 adverse event reports that list this exact device (DI 10705036012412), received between Dec 1, 2017 and Sep 8, 2026 — classified as 1 death, 14 injuries, 38 malfunctions.
The most frequently reported specific device problem is “Failure to Form Staple” (18 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Form Staple18
- Failure to Fire10
- Adverse Event Without Identified Device or Use Problem8
- Ejection Problem5
- No Apparent Adverse Event4
- Firing Problem2
- Misfire2
- Activation, Positioning or Separation Problem1
Patient problems
- No Consequences Or Impact To Patient24
- No Clinical Signs, Symptoms or Conditions16
- Failure to Anastomose7
- No Code Available7
- Blood Loss3
- Insufficient Information3
- Hemorrhage/Blood Loss/Bleeding2
- Sepsis2
Recent adverse event reports
The 25 most recent of 53 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 8, 2026No Apparent Adverse Event
It was reported that during an Unknown surgery, after fired, the staple line and cut line are both complete. noted oozing. To stop oozing with compression hemostasis. There were no adverse consequences to the patient. No additional information could be provided.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 50669923 ↗MalfunctionNov 12, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A RECTUM PROCEDURE SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23533148 ↗MalfunctionMar 13, 2025Ejection Problem; Ejection Problem
IT WAS REPORTED THAT DURING A COLECTOMY SURGERY, NOTED THE ANVIL LOCK-OUT LOOSE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21596903 ↗MalfunctionMar 13, 2025Ejection Problem; Ejection Problem
IT WAS REPORTED THAT DURING A PROCTECTOMY SURGERY, NOTED THE ANVIL LOCK-OUT LOOSE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21596656 ↗MalfunctionMar 5, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING A GASTRIC PROCEDURE SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21525150 ↗MalfunctionJul 18, 2023Ejection Problem; Ejection Problem
IT WAS REPORTED THAT DURING A ESOPHAGEAL PROCEDURE THEY COULD NOT ATTACH THE ANVIL. OPENED THE PACKING AND FOUND COULD NOT ATTACH THE ANVIL. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17344504 ↗MalfunctionJun 5, 2023Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING A RECTUM PROCEDURE SURGERY, THE DEVICE COULD NOT FIRE. CHANGED TO THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17057816 ↗MalfunctionMay 24, 2023Ejection Problem; Ejection Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE HEAD OF THE DEVICE CAME OFF. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16994392 ↗InjuryMay 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THERE WAS A POST-OP LEAKAGE. 1-DAY AFTER SURGERY A LEAKAGE WAS NOTED. THE SURGERY PROCEDURE WAS NORMAL. PROCEDURE: UNKNOWN.
Patient problems: Failure to Anastomose
MAUDE report 16919020 ↗MalfunctionMar 30, 2023Failure to Form Staple; Failure to Fire; No Apparent Adverse Event; Failure to Form Staple; Failure to Fire; No Apparent Adverse Event
IT WAS REPORTED WHO WAS IN THE CASE THAT DURING A ROBOTIC LOW ANTERIOR RESECTION PROCEDURE THAT ON THE INITIAL ATTEMPT BY THE SURGICAL TECH DID NOT FULLY DEPRESS THE HANDLE AND THE STAPLES DID NOT FULLY DEPLOY. THE TECH THEN STARTED THE PROCESS TO REMOVE THE DEVICE WHEN IT WAS NOTICED THAT THE DEVICE DID NOT FULLY FIRE. SURGEON THEN ATTEMPTED TO COMPLETE THE FIRING SEQUENCE BUT DID NOT WAIT THE FULL FIFTEEN SECONDS BEFORE FIRING THE DEVICE. DEVICE WAS REMOVED, A LEAK TEST WAS PREFORMED, AND A LEAK WAS FOUND. SURGEON RESECTED MORE TISSUE AND USED A COMPETITOR¿S DEVICE TO COMPLETE THE ANASTOMOSIS. WHILE SURGEON WAS COMPLETING THE NEW ANASTOMOSIS THE REP EXAMINED THE INITIAL DEVICE AND FOUND SOME MALFORMED STAPLES LEFT BEHIND IN THE DEVICE. LEAK TEST WAS PERFORMED ON THE NEW ANASTOMOSIS AND NO LEAKS WERE FOUND. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16643989 ↗InjuryDec 21, 2022Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT THERE WAS A PATIENT WITH A HISTORY OF DIVERTICULOSIS WHO UNDERWENT LAPAROSCOPIC SIGMOIDECTOMY ON (B)(6) 2020. ACCORDING TO MEDICAL RECORDS PROVIDED, A 33 MM EEA DILATOR WAS PLACED IN THE PATIENT¿S RECTUM AND A 33 MM STAPLER ADVANCED INTO THE RECTAL STUMP AND FIRED. THE SURGEON NOTED TWO INTACT TISSUE DONUTS, AND A LEAK TEST WAS NEGATIVE FOR LEAKS. THE PATIENT TOLERATED THE PROCEDURE WELL AND WAS DISCHARGED ON POSTOP DAY 2. SHE RETURNED ON POSTOP DAY 3 WITH PAIN AND FEVER AND SUBSEQUENTLY DEVELOPED AN ANASTOMOTIC LEAK, WHICH REQUIRED EXPLORATORY LAPAROTOMY, RESECTION OF THE ANASTOMOSIS WITH END COLOSTOMY. SHE ALSO DEVELOPED OTHER POST OPERATIVE COMPLICATIONS WITH EXTENDED RECOVERY AND THE COLOSTOMY WAS REVERSED THE FOLLOWING YEAR.
Patient problems: Failure to Anastomose; Fever; Insufficient Information; Pain; Peritonitis; Sepsis
MAUDE report 16022927 ↗InjuryNov 29, 2022Failure to Form Staple; Adverse Event Without Identified Device or Use Problem; Failure to Form Staple; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED VIA THE RISK MANAGER RECEIVED NOTICE OF LITIGATION FROM ATTORNEY FOR PATIENT WITH HISTORY OF DIVERTICULITIS WHO UNDERWENT SIGMOID RESECTION FOR ¿THE REMOVAL OF A LARGE MASS IN HIS COLON.¿ PLEADING FURTHER ALLEGES THAT ON POST OPERATIVE DAY FIVE, PATIENT WAS DIAGNOSED WITH SEPSIS AND RUSHED TO THE OPERATING ROOM FOR EMERGENCY SURGERY TO REPAIR A COLORECTAL ANASTOMOTIC LEAK. IT IS ALLEGED THAT THAT THE ANASTOMOTIC LEAK WAS CAUSED BY AN AREA OF NON-FUNCTIONAL STAPLES. PLEADING FURTHER STATES PATIENT HAD TO UNDERGO NUMEROUS OTHER SURGERIES, INCLUDING ¿THE PLACEMENT OF A VERY LARGE AREA OF MESH THROUGHOUT HIS ENTIRE ABDOMEN.¿
Patient problems: Failure to Anastomose; Sepsis
MAUDE report 15880209 ↗MalfunctionOct 3, 2022Failure to Fire; No Apparent Adverse Event; Failure to Fire; No Apparent Adverse Event
IT WAS REPORTED THAT DURING A LAPAROTOMY CONVERSION OF BILLROTH TO GASTRIC BYPASS PROCEDURE THAT WHILE CREATING THE ANASTOMOSIS ON THE GASTRIC POUCH THAT ONLY HALF OF THE OUTER ROW OF STAPLES DEPLOYED. THE STAPLES THAT WERE DEPLOYED WERE FULLY FORMED. A LITTLE OF THE TISSUE WAS TORN WHILE REMOVING THE DEVICE. ANASTOMOSIS WAS THEN OVER SEWN WITH SUTURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED.
Patient problems: Insufficient Information
MAUDE report 15528192 ↗MalfunctionSep 15, 2022Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15429477 ↗MalfunctionSep 9, 2022Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping
IT WAS REPORTED THAT DURING THE RECTUM PROCEDURE, OPEN THE PACKING AND NOTED THAT SOME STAPLES FELL OFF. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15391332 ↗DeathMar 8, 2022Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE OPERATION WAS (B)(6) 2022 IN THE MEDICAL CENTER OF (B)(6). THE PATIENT DIED. THE AUTOPSY FOUND OUT THAT THE CIRCULAR SEAM WAS 0.8 CM LONG. THE DATE OF DEATH IS CLARIFIED. PROCEDURE: LAPAROTOMY
Patient problems: Fluid Discharge; Insufficient Information; Multiple Organ Dysfunction Syndrome
MAUDE report 13695574 ↗InjuryFeb 1, 2022Failure to Form Staple; Adverse Event Without Identified Device or Use Problem; Failure to Form Staple; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A SIGMOID COLECTOMY PROCEDURE THAT AFTER COMPLETING THE FIRING SEQUENCE THE STAPLER SEEMED TO FIRE AS NORMAL AND THERE WERE TWO INTACT DONUTS. WHEN THEY WENT TO LOOK IN THE ABDOMEN THERE WAS ¿STOOL EVERYWHERE¿, THE STAPLES NEVER GRABBED THE TISSUE TO DO THE ANASTOMOSIS. SURGEON ENDED UP PREFORMED A COLOSTOMY ON THE PATIENT.
Patient problems: Failure to Anastomose
MAUDE report 13417377 ↗InjuryJan 20, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT POST OP OF A LOW ANTERIOR RESECTION, PATIENT WITH POST-OP BLEED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 13308850 ↗MalfunctionJan 4, 2022No Apparent Adverse Event; Failure to Eject; No Apparent Adverse Event; Failure to Eject
IT WAS REPORTED THAT DURING A COLECTOMY THE CIRCULAR STAPLER MISFIRED. CASE WAS COMPLETED WITH ANOTHER LIKE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13154276 ↗MalfunctionAug 9, 2021Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT DURING THE SURGERY, BEFORE USED ON THE PATIENT, NOTED THE PACKING WAS DAMAGED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12293403 ↗MalfunctionJun 14, 2021Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING A SIGMOID RESECTION PROCEDURE THAT THE DEVICE MISFIRED AND ONLY DEPLOYED A HALF CIRCLE. ANASTOMOSIS WAS COMPLETED USING SILK SUTURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11991277 ↗MalfunctionMar 1, 2021Ejection Problem; Ejection Problem
IT WAS REPORTED THAT DURING A HEMICOLECTOMY, THE ANCILLARY TROCAR COULD BE FITTED ONTO THE TROCAR OF ECS. BUT REMOVAL WAS VERY DIFFICULT. REMOVED IT BY LOT OF STRUGGLE. LATER THEY USED LINEAR CUTTER TO COMPLETE THE PROCEDURE. NO FRAGMENTS WERE GENERATED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE SURGERY WAS COMPLETED AS USUAL. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11394048 ↗MalfunctionOct 15, 2020Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING THE ENDOSCOPIC BOWEL SURGERY, AFTER FIRED, NOTED STAPLES MALFORMED WITH IRREGULAR SHAPE (WITH STRAIGHT LEG). CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10683154 ↗MalfunctionAug 10, 2020Failure to Form Staple; Failure to Form Staple
SEE ATTACHED USER FACILITY MEDWATCH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10384755 ↗MalfunctionJul 24, 2020Firing Problem; Firing Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE WOULDN'T FIRE. NO OTHER INFORMATION IS AVAILABLE. THERE WERE NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10323301 ↗
FDA clearance and classification
The FDA regulates product code OCW (endoscopic tissue approximation device) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the N/A ECS33A
Has the N/A ECS33A been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705036012412), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the N/A ECS33A?
The FDA's MAUDE database lists 53 adverse event reports for this device, received between Dec 1, 2017 and Sep 8, 2026. In those reports, the most common specific device problems are “Failure to Form Staple” (18), “Failure to Fire” (10) and “Ejection Problem” (5); the most common patient problems are “Failure to Anastomose” (7), “Blood Loss” (3) and “Hemorrhage/Blood Loss/Bleeding” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the N/A ECS33A?
1 report is classified as a death, 14 as injuries and 38 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the N/A ECS33A MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the N/A ECS33A, and is it still sold?
The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand N/A, model ECS33A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endoscopic tissue approximation device devices
- Endo Stitch 173016Covidien LP · Endoscopic tissue approximation device1 death1,332 reports
- ProTack 174006Covidien LP · Endoscopic tissue approximation device705 reports
- N/A ECS29AEthicon Endo-Surgery, LLC · Endoscopic tissue approximation device1 death517 reports
- N/A ECS25AEthicon Endo-Surgery, LLC · Endoscopic tissue approximation device233 reports
- OverStitch™ M00505530Boston Scientific Corporation · Endoscopic tissue approximation device1 recall75 reports
- Polysorb 170053Covidien LP · Endoscopic tissue approximation device1 recall63 reports
- Surgidac 173024Covidien LP · Endoscopic tissue approximation device70 reports
- NA M00505550Boston Scientific Corporation · Endoscopic tissue approximation device1 recall12 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page