LIGACLIP LT200Recalls & reported problems
Ethicon Endo-Surgery, LLC · LIGACLIP EXTRA Ligating Clip Cartridges: 6 Titanium Clips/ Cartridge (Medium) · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon LIGACLIP LT200 is a cartridge of six medium titanium ligating clips used to tie off blood vessels and tissue during surgery. The 76 reports mostly describe clips that fail to close properly, snap in half when loaded, sit crooked or too loose on the applier, or scatter when the cartridge is opened, sometimes delaying the operation. Eight reports describe injuries and none describe deaths. No recalls name this device, and the number of reports is steady at 6 over each of the past 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 75
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 75 adverse event reports that list this exact device (DI 10705036012801), received between Jan 4, 2018 and Jan 16, 2026 — classified as 8 injuries, 67 malfunctions.
The most frequently reported specific device problem is “Failure to Form Staple” (45 reports).
3 other recalls cover devices cleared under the same FDA submission (K830503); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Form Staple45
- Appropriate Device Problem Term/Code Not Available12
- Unintended Ejection12
- Break7
- Mechanical Problem7
- Defective Device5
- Difficult to Insert5
- Device Slipped4
Patient problems
- No Clinical Signs, Symptoms or Conditions41
- No Consequences Or Impact To Patient26
- No Code Available5
- Pain5
- Blood Loss4
- Cardiac Arrest2
- Hemorrhage/Blood Loss/Bleeding2
- Unspecified Infection2
Recent adverse event reports
The 25 most recent of 75 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 16, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, WHEN INSTALLING CLIP WITH APPLIER, CLIP IS CROOKED. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24096739 ↗MalfunctionJan 16, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, WHEN INSTALLING CLIP WITH APPLIER, CLIP IS CROOKED. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24096731 ↗MalfunctionJan 16, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, WHEN INSTALLING CLIP WITH APPLIER, CLIP IS CROOKED. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24096709 ↗MalfunctionJan 15, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THE SCISSORING EFFECT SEEN IN THE APPLICATORS. THE PROCEDURE WAS DELAYED BY 10 MINUTES AND COMPLETED SUCCESSFULLY. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24082253 ↗MalfunctionNov 24, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE CLIPS ARE SLIPPING AND NOT HOLDING SECURELY ON VESSELS DURING PROCEDURES. NO OTHER DETAILS PROVIDED. NO PATEINT CONSEQUENCE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23638374 ↗MalfunctionJun 3, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE WHEN OPENED, THE NAIL CLIPS ARE ALL SCATTERED ON THE OPERATING TABLE, AND THE NAIL CLIPS ARE MANUALLY FIXED ON THE CLAMPING FORCEPS, AND THE TISSUE IS TRIED TO CLAMP, BUT BECAUSE THE NAIL CLAMPS ARE TOO LOOSE, THE TISSUE CAN NOT BE FIXED AND CLAMPED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22131269 ↗MalfunctionMar 20, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TIP DOESN'T SCREW ON TIGHTLY, SO THE CLIP IS MOUNTED AT AN ANGLE. NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21652170 ↗InjuryDec 3, 2024Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, DEVICE HAS ASYMMETRICAL JAWS AND THE CLIPS ARE NOT FORMED PROPERLY.
Patient problems: Cardiac Arrest; Hemorrhage/Blood Loss/Bleeding
MAUDE report 20833030 ↗InjuryDec 3, 2024Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, DEVICE HAS ASYMMETRICAL JAWS AND THE CLIPS ARE NOT FORMED PROPERLY.
Patient problems: Cardiac Arrest; Hemorrhage/Blood Loss/Bleeding
MAUDE report 20833017 ↗MalfunctionNov 19, 2024Failure to Form Staple; Defective Device; Failure to Form Staple; Defective Device
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, STAPLES TACK TOGETHER AT AN ANGLE AND THE WHITE HOUSINGS IN WHICH THE STAPLES ARE INSERTED ARE COMING APART. NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20727009 ↗MalfunctionNov 19, 2024Failure to Form Staple; Defective Device; Failure to Form Staple; Defective Device
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, STAPLES TACK TOGETHER AT AN ANGLE AND THE WHITE HOUSINGS IN WHICH THE STAPLES ARE INSERTED IS COMING APART. NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20725226 ↗MalfunctionOct 16, 2024Failure to Form Staple; Defective Device; Failure to Form Staple; Defective Device
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, CLIPS NOT MAINTAINING INTEGRITY CARTRIDGE BREAKING IN HALF AND CLIPS FALLING OUT. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20466497 ↗MalfunctionSep 24, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE LT200 LIGA CLIPS BROKE WHILE FIRING. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20292278 ↗MalfunctionSep 24, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE LT200 LIGA CLIPS BROKE WHILE FIRING. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20292227 ↗MalfunctionSep 24, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE LT200 LIGA CLIPS BROKE WHILE FIRING. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20292183 ↗MalfunctionSep 24, 2024Mechanical Problem; Failure to Form Staple; Mechanical Problem; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE LT200 LIGA CLIPS BROKE WHILE FIRING. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20292096 ↗MalfunctionSep 24, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE LT200 LIGA CLIPS BROKE WHILE FIRING. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20292092 ↗MalfunctionSep 24, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE LT200 LIGA CLIPS BROKE WHILE FIRING. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20292009 ↗MalfunctionAug 8, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIPS BREAK. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19938429 ↗MalfunctionDec 22, 2023Unintended Ejection; Difficult to Insert; Manufacturing, Packaging or Shipping Problem; Unintended Ejection; Difficult to Insert; Manufacturing, Packaging or Shipping Problem
IT WAS REPORTED THAT DURING A MASTECTOMY PROCEDURE, WHEN REMOVE THE CLIP BANK FROM THE STERILE PACKAGING, IT ALREADY FALLS APART. THE CLIPS CAN THEN NO LONGER BE PROPERLY LOADED. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED. PROCEDURE FINISHED WITH SAME LIKE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18388591 ↗MalfunctionDec 15, 2023Difficult to Insert; Failure to Form Staple; Difficult to Insert; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIP WAS MALFORMED. THE CLIP COULD NOT BE FED INTO THE JAW PROPERLY. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18336666 ↗MalfunctionNov 30, 2023Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, CLIP COULD NOT CLOSE. CHANGED TO ANOTHER DEVICE TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCE REPORTED, NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18239898 ↗MalfunctionNov 9, 2023Unintended Ejection; Failure to Form Staple; Unintended Ejection; Failure to Form Staple
IT WAS REPORTED THAT DURING A VASECTOMY PROCEDURE, THE NEEDED TO TAKE OUT EXTRA MAGAZINES OF CLIPS. THE CLIPS WHERE SOMETIMES FORMED AN "S" ALREADY IN THE MAGAZINE, SOMETIMES THE CLIPS JUST FELL OUT OF THE MAGAZINE. NO SURGICAL DELAY. PROCEDURE COMPLETED SUCCESSFULLY. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18104965 ↗MalfunctionNov 9, 2023Unintended Ejection; Failure to Form Staple; Unintended Ejection; Failure to Form Staple
IT WAS REPORTED THAT DURING A VASECTOMY PROCEDURE, THE NEEDED TO TAKE OUT EXTRA MAGAZINES OF CLIPS. THE CLIPS WHERE SOMETIMES FORMED AN "S" ALREADY IN THE MAGAZINE, SOMETIMES THE CLIPS JUST FELL OUT OF THE MAGAZINE. NO SURGICAL DELAY. PROCEDURE COMPLETED SUCCESSFULLY. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18104850 ↗MalfunctionNov 9, 2023Unintended Ejection; Failure to Form Staple; Unintended Ejection; Failure to Form Staple
IT WAS REPORTED THAT DURING A VASECTOMY PROCEDURE, THE NEEDED TO TAKE OUT EXTRA MAGAZINES OF CLIPS. THE CLIPS WHERE SOMETIMES FORMED AN "S" ALREADY IN THE MAGAZINE, SOMETIMES THE CLIPS JUST FELL OUT OF THE MAGAZINE. NO SURGICAL DELAY. PROCEDURE COMPLETED SUCCESSFULLY. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18104439 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K830503. One submission often covers many models, so check the product and lot details in each record.
- Z-1775-2013 · Apr 12, 2013 · TerminatedLIGACLIP Endoscopic Rotating Multiple Clip Applier (ER320), 10mm M/L Product Usage: The LIGACLIP Endoscopic Rotating Multiple Clip Applier is a sterile, single patient use instru
- Z-1777-2013 · Apr 12, 2013 · TerminatedLIGACLIP Endoscopic Rotating Multiple Clip Applier (ER320), 10mm M/L, bulk non-sterile Product Usage: The LIGACLIP Endoscopic Rotating Multiple Clip Applier is a single patient us
- Z-1776-2013 · Apr 12, 2013 · TerminatedLIGACLIP Endoscopic Rotating Multiple Clip Applier Convenience Kit (Flex Trays), 10mm M/L Product Usage: The LIGACLIP Endoscopic Rotating Multiple Clip Applier is a sterile, sing
Other CLIP, IMPLANTABLE devices
- LIGAMAX EL5MLEthicon Endo-Surgery, LLC · CLIP, IMPLANTABLE3 deaths3,808 reports
- Endo Clip III 176630Covidien LP · CLIP, IMPLANTABLE1 death1,959 reports
- Epix® Universal Clip Applier CA500Applied Medical Resources Corporation · Clip, Implantable1 recall1 death1,183 reports
- Endo Clip II 176657Covidien LP · CLIP, IMPLANTABLE1,183 reports
- LIGACLIP ER320Ethicon Endo-Surgery, LLC · CLIP, IMPLANTABLE1,115 reports
- Direct Drive® Clip Applier CA090Applied Medical Resources Corporation · Clip, Implantable1 death806 reports
- LAPRA-TY XC200Ethicon Inc. · CLIP, IMPLANTABLE791 reports
- Endo Clip 176625Covidien LP · CLIP, IMPLANTABLE1 death768 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page