Skip to content

LIGACLIP LT400Recalls & reported problems

Ethicon Endo-Surgery, LLC · LIGACLIP EXTRA Ligating Clip Cartridges: 6 Titanium Clips/ Cartridge (Large) · Class 2 · Implant

All models of LIGACLIP (26)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10705036012832), received between Jun 28, 2018 and Dec 4, 2025 — classified as 3 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Delivered as Unsterile Product” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K830503); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Appropriate Device Problem Term/Code Not Available2
  • No Apparent Adverse Event2
  • Delivered as Unsterile Product1
  • Detachment of Device or Device Component1
  • Difficult to Open or Close1
  • Loose or Intermittent Connection1

Patient problems

  • Failure to Anastomose2
  • Hemorrhage/Blood Loss/Bleeding2
  • No Clinical Signs, Symptoms or Conditions2
  • No Code Available2
  • No Consequences Or Impact To Patient2
  • Abdominal Distention1
  • Fistula1
  • Ischemia1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 4, 2025No Apparent Adverse Event; Appropriate Device Problem Term/Code Not Available; No Apparent Adverse Event; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT, DURING A VASCULAR PROCEDURE, CUTTING THE VESSEL RESULTED IN BLEEDING INSTEAD OF CLAMPING. NO FURTHER INFORMATION PROVIDED. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23716076 ↗
  • MalfunctionFeb 28, 2024Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT DURING INVENTORY IT WAS NOTICED THAT THERE WAS A PERFORATION IN DEVICES SECONDARY PACKAGING. DEVICE NOT USED IN PROCEDURE. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18797635 ↗
  • InjuryJul 11, 2023Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT POST OP TO A CHOLANGIOGRAPHY BY VIDEO LAPAROSCOPY, THE PATIENT EXPERIENCED SEVERE ABDOMINAL PAIN, DISTENSION, NAUSEA AND VOMITING. THE PATIENT UNDERWENT A CT SCAN OF THE ABDOMEN AND IT WAS DISCOVERED THAT THERE WAS A SMALL AMOUNT OF FLUID IN THE ABDOMINAL CAVITY AND NOTABLY IN THE PELVIC REGION, AS WELL AS OBLITERATION OF ADJACENT FAT, RELATED TO PREVIOUS MANIPULATION. THERE WAS A NEED FOR SURGICAL RE-APPROACH, WHICH WAS IDENTIFIED THAT THE CLIP WAS OPEN.

    Patient problems: Abdominal Distention; Failure to Anastomose; Nausea; Pain; Vomiting

    MAUDE report 17302454 ↗
  • MalfunctionSep 23, 2022No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT POST OP TO A LAPAROSCOPIC CHOLECYSTECTOMY THE PATIENT WAS FOUND TO HAVE 3,000 ML OF BLOOD IN THE ABDOMEN. THE PATIENT WAS REOPERATED ON AND IT WAS DISCOVERED THE CLIPS DID NOT CLOSE IN THE MIDDLE. THIS WAS CORRECTED AND FOUR UNITS OF BLOOD WERE GIVEN AND THE PATIENT IS IMPROVING.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 15474083 ↗
  • InjuryJul 18, 2019Loose or Intermittent Connection

    IT WAS REPORTED THAT PATIENT WITH ACUTE PANCREATITIS, HOSPITALIZED ON (B)(6) 2019, UNDERWENT LAPAROSCOPIC SURGERY CONVERTED TO LAPAROTOMY WITH BILEODIGESTIVE DERIVATION (HEPATICOJEJUNOSTOMY IN Y-DE-ROUX). CLIPS APPLIED IN THE LIGATION OF THE BILIARY TRACT WERE LOOSEN. PREOPERATIVE DIAGNOSIS: BILIARY FISTULA. POSTOPERATIVE DIAGNOSIS: LOOSENING OF THE CYSTIC DUCT LIGATURE + PARTIAL ISCHEMIA OF THE COLÉDOCO AND HEPATIC DUCT. INTERCURRENT: CONVERSION TO OPEN SURGERY. DESCRIPTION OF THE SURGERY: PATIENT IN DORSAL DECUBITUS UNDER GENERAL ANESTHESIA. ANTISEPSIS AND PLACEMENT OF STERILE FIELDS. WITHDRAWAL OF THE STITCHES OF THE PREVIOUS SURGERY AND PNEUMOPERITONEUM DONE BY THE UMBILICAL INCISION BY OPEN TECHNIQUE. CAVITY REVIEW: SMALL AMOUNT OF FREE BILIARY SECRETION IN HYPOCHONDRIUM D, INTENSE INFLAMMATORY BLOCK IN HCD. THE BLOCKAGE WAS DISASSEMBLED WITH EXTREME CAUTION WITH VISUALIZATION OF LOOSENING OF THE CYSTIC LIGATION CLIP, PARTIAL ISCHEMIA OF THE COMMON BILE DUCT AND HEPATIC DUCT APPROXIMATELY 20 MM PROXIMAL AND DISTAL TO THE LOCAL OF THE LOOSEN CLIP, INTENSE LOCAL INFLAMMATORY PROCESS. DUE TO THE IMPOSSIBILITY OF CONTINUING THE PROCEDURE BY LAPAROSCOPY, IT WAS DECIDED TO CONVERT T

    Patient problems: Failure to Anastomose; Fistula; Ischemia; No Code Available

    MAUDE report 8804490 ↗
  • InjurySep 12, 2018Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED VIA JOURNAL ARTICLE: TITLE: MID-TERM OUTCOMES OF PATIENTS UNDERGOING ADJUSTABLE PULMONARY ARTERY BANDING AUTHORS: SACHIN TALWAR, NEERAJ ARAVIND KAMAT, SHIV KUMAR CHOUDHARY, SIVASUBRAMANIAN RAMAKRISHNAN, ANITA SAXENA, RAJNISH JUNEJA, SHYAM SUNDER KOTHARI, AND BALRAM AIRAN CITATION: I N D I A N H E A R T J O U R N A L 6 8 ( 2 0 1 6 ) 7 2 ¿ 7 6, HTTP://DX.DOI.ORG/10.1016/J.IHJ.2015.06.036. THE AIM OF THIS STUDY WAS TO ANALYZE THE OUTCOMES OF APAB IN THE MID-TERM IN TERMS OF THE REDUCTION OF PA PRESSURES AND SUITABILITY FOR DEFINITIVE PALLIATION AND TO STUDY THE BAND RELATED COMPLICATIONS ENCOUNTERED AT DEFINITIVE SURGERY. BETWEEN JANUARY 2002 UNTIL DECEMBER 2002, 73 PATIENTS UNDERWENT ADJUSTABLE PULMONARY ARTERY BAND (APAB) WERE INCLUDED IN THE STUDY. DURING THE PROCEDURE, THE MPA WAS LOOPED WITH A NO. 2 ETHIBOND (JOHNSON AND JOHNSON INC, SOMERVILLE, NJ) SUTURE AFTER PASSING A RIGHT-ANGLE FORCEPS BETWEEN THE AORTA AND THE MPA. BOTH ARMS OF THE SUTURE WERE CLIPPED TOGETHER WITH A LIGACLIP (LT 200; ETHICON ENDOSURGERY INC, CINCINNATI, OH) JUST ON THE PLEDGET. THESE SUTURES WERE BROUGHT OUT THROUGH THE PERICARDIUM AND THE LATERAL EDGE OF THE STERNUM AND THEN THROUGH SUBCU

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 7869127 ↗
  • MalfunctionJul 9, 2018Difficult to Open or Close

    THE CLIPS DO NOT CLOSE PROPERLY WITH CONSEQUENT ISSUE FOR THE HEMOSTASIS. PROCEDURE IS UNKNOWN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7669251 ↗
  • MalfunctionJun 28, 2018Detachment of Device or Device Component

    AFTER THE APPLICATION THE CLIP DETACHED. THE ISSUE WAS SOLVED USING NEW ONE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7646936 ↗

FDA clearance and classification

Cleared through 510(k) K830503 on Mar 8, 1983 (decision: Substantially Equivalent). Submitted by ETHICON, Inc. as “Ligaclip Titanium Ligating Clip”. The FDA review took 20 days. FDA record ↗

The FDA regulates product code FZP (clip, implantable) under 21 CFR 878.4300 as a Class II device. 21 CFR 878.4300 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K830503. One submission often covers many models, so check the product and lot details in each record.

  • Z-1775-2013 · Apr 12, 2013 · Terminated
    LIGACLIP Endoscopic Rotating Multiple Clip Applier (ER320), 10mm M/L Product Usage: The LIGACLIP Endoscopic Rotating Multiple Clip Applier is a sterile, single patient use instru
  • Z-1777-2013 · Apr 12, 2013 · Terminated
    LIGACLIP Endoscopic Rotating Multiple Clip Applier (ER320), 10mm M/L, bulk non-sterile Product Usage: The LIGACLIP Endoscopic Rotating Multiple Clip Applier is a single patient us
  • Z-1776-2013 · Apr 12, 2013 · Terminated
    LIGACLIP Endoscopic Rotating Multiple Clip Applier Convenience Kit (Flex Trays), 10mm M/L Product Usage: The LIGACLIP Endoscopic Rotating Multiple Clip Applier is a sterile, sing

Common questions about the LIGACLIP LT400

Has the LIGACLIP LT400 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705036012832), based on the FDA recall file dated Nov 27, 2026. However, 3 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the LIGACLIP LT400?

The FDA's MAUDE database lists 8 adverse event reports for this device, received between Jun 28, 2018 and Dec 4, 2025. In those reports, the most common specific device problems are “Delivered as Unsterile Product” (1), “Detachment of Device or Device Component” (1) and “Difficult to Open or Close” (1); the most common patient problems are “Failure to Anastomose” (2), “Hemorrhage/Blood Loss/Bleeding” (2) and “Abdominal Distention” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the LIGACLIP LT400?

None of the reports are classified as deaths, 3 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the LIGACLIP LT400 FDA approved?

It was cleared through 510(k) K830503 on Mar 8, 1983. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the LIGACLIP LT400 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the LIGACLIP LT400, and is it still sold?

The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand LIGACLIP, model LT400). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CLIP, IMPLANTABLE devices

All CLIP, IMPLANTABLE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page