Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10705036013792), received between Dec 20, 2019 and Nov 11, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Failure to Form Staple” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Form Staple3
- Failure to Fire1
- Material Integrity Problem1
- Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 11, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23523012 ↗MalfunctionMar 21, 2024Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT DURING AN PERITONECTOMY PROCEDURE, A RELOAD WAS OPENED AND FOUND TO BE BENT/BROKEN. NO FURTHER DETAILS WERE PROVIDED. NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18949530 ↗MalfunctionAug 4, 2022Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING THE SURGERY, WHEN SEVERING VASCULAR TISSUE, AFTER FIRED, NOTED SOME STAPLES MALFORMED . CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15167092 ↗MalfunctionMar 31, 2022Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING THE SURGERY, COULD NOT FIRE FARTHER AFTER FIRED 2/3. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13971529 ↗MalfunctionJan 8, 2021Separation Problem; Separation Problem
IT WAS REPORTED THAT DURING A NEPHRECTOMY PROCEDURE, UPON OPENING THE RELOAD, THE METAL TRAY ON THE RELOAD WAS NOTED TO BE DISLODGED FROM THE PLASTIC CARTRIDGE. A NEW RELOAD WAS OPENED. THE PROCEDURE WAS COMPLETED WITH NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11142503 ↗MalfunctionDec 20, 2019Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, ¿MD HAD USED SEVERAL GRAY ECHELON STAPLE RELOADS WITH THE POWERED ECHELON 60 STAPLER. WHEN SHE USED THE 5TH GRAY RELOAD, THAT WAS FROM A DIFFERENT LOT NUMBER FOR THE 4 PREVIOUS, IT MALFUNCTIONED. THE STAPLER FIRED NORMALLY, BUT THE STAPLE LINE WAS NOT SEALED FULLY WITH STAPLES. MD HAD TO QUICKLY USE SEVERAL HEMO-CLIPS ON THE VESSELS THAT WERE SPRAYING BLOOD. THIS COULD HAVE BEEN A VERY SERIOUS COMPLICATION HAD SHE NOT BEEN ABLE TO REACT QUICKLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9505550 ↗
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,680 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,730 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,356 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,294 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,011 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,965 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,909 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,046 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page