ECHELON Flex ENDOPATH PSE45ARecalls & reported problems
Ethicon Endo-Surgery, LLC · Echelon Flex45 Powered Endopath Stapler Articulating Endoscopic Linear Cutter · Class 2
All models of ECHELON Flex ENDOPATH (7)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Echelon Flex is a powered stapler used to cut and staple tissue during keyhole surgery. Its 382 reports include 1 death and 40 injuries, and often describe the stapler failing to fire, staples not forming properly and jaws that are hard to open or close. Many reports record no patient effect; some describe bleeding. No recalls name this device. Reports fell to 12 over the last year from 30 in the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 384
- Deaths reported
- 1
- Injuries reported
- 40
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 384 adverse event reports that list this exact device (DI 10705036014317), received between Mar 28, 2013 and Sep 17, 2026 — classified as 1 death, 40 injuries, 343 malfunctions.
The most frequently reported specific device problem is “Failure to Fire” (116 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Fire116
- Failure to Form Staple90
- Difficult to Open or Close72
- Mechanical Jam46
- Retraction Problem32
- Activation Failure24
- Adverse Event Without Identified Device or Use Problem23
- Mechanical Problem9
Patient problems
- No Clinical Signs, Symptoms or Conditions193
- No Consequences Or Impact To Patient150
- Hemorrhage/Blood Loss/Bleeding19
- No Code Available17
- Failure to Anastomose6
- Blood Loss5
- Low Blood Pressure/ Hypotension3
- Unspecified Infection3
Recent adverse event reports
The 25 most recent of 384 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026Failure to Cut; Difficult to Open or Close; Activation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50807772 ↗MalfunctionAug 14, 2026Failure to Cut
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50260359 ↗MalfunctionAug 14, 2026Positioning Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50254402 ↗MalfunctionJun 24, 2026Difficult to Open or Close
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THEY COULD NOT OPEN THE JAW AFTER FIRING. REPEATED MANY TIMES TO OPEN THE JAW (FOLLOWED UP IFU); AND FINALLY, THEY OPENED THE JAW SUCCESSFULLY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25610940 ↗MalfunctionJun 24, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE DEVICE DID NOT FIRE; AND HAD NO MOTOR SOUND. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25610899 ↗MalfunctionJun 24, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE DEVICE DID NOT FIRE; AND HAD NO MOTOR SOUND. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25610856 ↗MalfunctionApr 17, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24915727 ↗MalfunctionApr 16, 2026Failure to Fire
IT WAS REPORTED THAT DURING A PROSTATECTOMY PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24897936 ↗MalfunctionMar 24, 2026Material Integrity Problem
IT WAS REPORTED THAT PREOPERATIVELY TO AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE MANUAL OVERRIDE DOOR FELL OFF IN THE PACKAGING. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24674749 ↗MalfunctionMar 17, 2026Failure to Form Staple; No Apparent Adverse Event
IT WAS REPORTED THAT DURING A PULMONARY PROCEDURE, IT WAS OBSERVED THAT THERE WAS BLEEDING AFTER A FIRE OF WHITE RELOAD (NOT THE 1ST FIRE). THE VOLUME OF BLEEDING WAS ABOUT 300CC. SUTURE SEWING WAS USED TO CONTROL THE BLEEDING. AFTER A FIRE OF A BLUE RELOAD, NOTED MALFORMED STAPLES. CHANGED TO ANOTHER COMPETITIVE DEVICE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 24618433 ↗MalfunctionJan 26, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24178207 ↗MalfunctionJan 5, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23957445 ↗InjuryDec 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A LAP GASTRIC BYPASS THE DOCTOR PERFORMED THE BYPASS PROCEDURE WITH OUR BARIATRIC KIT (B)(6), BUT THE PATIENT RETURNED TO THE ROOM WITH INTENSE CRAVING AND VOMITING, PRESENTING BLEEDING. PATIENT HAD TO GO TO THE ICU (INTENSIVE CARE UNIT). DID THE PATIENT EXPERIENCE AN ADVERSE EVENT SUCH AS INFECTION, NON-UNION, ALLERGIC REACTION, OSTEOPOROSIS, OVERLOADING, PAIN, DEGENERATIVE DISEASES, BLEEDING OR OOZING? YES. DID THE PATIENT REQUIRE REVISION SURGERY OR HARDWARE REMOVAL? NO. PATIENT STATUS/ OUTCOME / CONSEQUENCES: YES. PATIENT CONSEQUENCE DESCRIPTION/WAS THERE A CLINICAL OUTCOME EXPERIENCED BY THE PATIENT (INFECTION, INFLAMMATION, ETC.)? PATIENT RETURNED TO THE ROOM WITH INTENSE CRAVING AND VOMITING, PRESENTING BLEEDING. PATIENT HAD TO GO TO THE ICU.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Vomiting
MAUDE report 23837687 ↗InjuryDec 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A LAP GASTRIC BYPASS THE DOCTOR REPORTED THAT THE PATIENT PRESENTED VOMITING WITH BLOOD AND CLOT IN THE POSTOPERATIVE PERIOD. DID THE PATIENT EXPERIENCE AN ADVERSE EVENT SUCH AS INFECTION, NON-UNION, ALLERGIC REACTION, OSTEOPOROSIS, OVERLOADING, PAIN, DEGENERATIVE DISEASES, BLEEDING OR OOZING? UNKNOWN. DID THE PATIENT REQUIRE REVISION SURGERY OR HARDWARE REMOVAL? UNKNOWN. PATIENT STATUS/ OUTCOME / CONSEQUENCES: YES. PATIENT CONSEQUENCE DESCRIPTION/WAS THERE A CLINICAL OUTCOME EXPERIENCED BY THE PATIENT (INFECTION, INFLAMMATION, ETC.)? POST OPERATIVE BLEEDING.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Vomiting
MAUDE report 23837668 ↗MalfunctionOct 21, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY, THE DEVICE COULD NOT BE FIRED AND HAD NO MOTOR SOUND ON THE FIRING. THE SURGEON REMOVED THE BATTERY AND ROTATED FOR 180 DEGREE AND REINSERTED THE BATTERY. THE DEVICE STILL DID NOT FIRE AND HAD NO MOTOR SOUND. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23348312 ↗MalfunctionOct 17, 2025Ejection Problem; Ejection Problem
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY, IT WAS OBSERVED THAT THE CARTRIDGE COULD NOT BE INSTALLED. CHANGED TO A NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23327910 ↗MalfunctionAug 16, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22823199 ↗MalfunctionAug 16, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT BE FIRED. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22823187 ↗MalfunctionAug 15, 2025Failure to Form Staple; Difficult to Open or Close; Failure to Form Staple; Difficult to Open or Close
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. IT WAS FURTHER REPORTED THAT THEY COULD NOT OPEN THE JAW. OPENED THE JAW MANUALLY WITH BIG FORCE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. CHANGED ANOTHER ONE TO CONTINUE THE PROCEDURE BUT THE SAME PROBLEM HAPPENED AGAIN. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22819932 ↗MalfunctionAug 15, 2025Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE KNIFE MOVED TO THE DISTAL END BUT DID NOT RETURN AUTOMATICALLY. THEY USED REVERSE BUTTON MANY TIMES TO RETURN KNIFE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22815934 ↗MalfunctionJul 30, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT BE FIRED. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22666380 ↗MalfunctionJul 28, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT BE FIRED. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22637575 ↗MalfunctionJun 11, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING A LUNG RESECTION PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT BE FIRED. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22188215 ↗MalfunctionMay 28, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE MANUAL OVERRIDE DOOR FELL OFF FROM THE DEVICE AFTER OPENING THE PACKAGE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22100631 ↗MalfunctionMay 23, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT BE FIRED WITHOUT MOTOR SOUND ON THE FIRING. THE SURGEON REMOVED THE BATTERY AND ROTATED FOR 180 DEGREE AND REINSERTED THE BATTERY. THE DEVICE STILL COULD NOT FIRE AND WITHOUT MOTOR SOUND. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22078419 ↗
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,680 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,730 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,356 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,294 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,011 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,965 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,909 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,046 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page