Ethicon Endo-Surgery, LLC · Echelon Flex60 Powered Endopath Stapler Articulating Endoscopic Linear Cutter · Class 2

Quick summary

This Ethicon Echelon Flex powered endoscopic stapler is used to cut and staple tissue during surgery. Its 329 reports describe staplers that fail to fire, staples that do not form properly, and jams or trouble opening and closing the jaws, and some say the surgeon switched to another unit to finish the procedure. The reports include 3 deaths and 20 injuries. No recalls name this device. Reports fell to 4 in the last 12 months from 13 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
329
Deaths reported
3
Injuries reported
20
Recalls naming it
0

Summary

The FDA's MAUDE database contains 329 adverse event reports that list this exact device (DI 10705036014324), received between Feb 22, 2013 and Aug 7, 2026 — classified as 3 deaths, 20 injuries, 306 malfunctions.

The most frequently reported specific device problem is “Failure to Form Staple” (95 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 13 reportsFeb 2024: 3 reportsMar 2024: 3 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 3 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Form Staple95
  • Failure to Fire75
  • Difficult to Open or Close66
  • Retraction Problem46
  • Mechanical Jam32
  • Activation Failure20
  • Firing Problem13
  • Break8

Patient problems

  • No Consequences Or Impact To Patient162
  • No Clinical Signs, Symptoms or Conditions142
  • No Code Available13
  • Blood Loss6
  • Failure to Anastomose5
  • Death3
  • Fistula3
  • Hemorrhage/Blood Loss/Bleeding3

Recent adverse event reports

The 25 most recent of 329 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2026Failure to Form Staple; Failure to Fire

    It was reported that during an unknown procedure, it was observed that the device would not fire. Changed to another one to continue the surgery but the same problem happened again. Changed to another four to continue the surgery, after fired, noted the staples malformed. Changed the new one to complete surgery. There was no patient consequence reported. No additional information can be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50179876 ↗
  • MalfunctionJun 24, 2026Failure to Fire

    IT WAS REPORTED THAT DURING A PROCTOCOLECTOMY PROCEDURE, IT WAS OBSERVED THAT THE DEVICE DID NOT FIRE. CHANGED TO ANOTHER ONE TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. CHANGED TO A NEW ONE TO COMPLETE THE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25612224 ↗
  • MalfunctionJun 9, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT FIRE AND HAD NO MOTOR SOUND. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25446127 ↗
  • MalfunctionApr 30, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE DID NOT COMPLETE THE STAPLING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25043819 ↗
  • MalfunctionAug 28, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT BE FIRED. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22912712 ↗
  • InjuryJul 18, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    TITLE:CLINICAL APPLICATION VALUE OF DOMESTIC AND IMPORTED STAPLER BASED ON REAL WORLD DATA ANALYSIS AUTHORS: WANG HONGDAN, ZHANG BENXING, WEI'ANHAI, YANG LI, ZHANG HEHUA CITATION: CHINA MEDICAL EQUIPMENT 2024 APRIL VOL.21 NO.4;DOI: 10.3969/J.ISSN .1672- 8270.2024.04.030 THE AIM OF THIS STUDY IS TO INVESTIGATE AND ANALYZE THE CLINICAL USE VALUE OF IMPORTED AND DOMESTIC BRANDS OF STAPLERS BASED ON REAL WORLD DATA. A TOTAL OF 466 PATIENTS WHO RECEIVED STAPLER ANASTOMOSIS IN DAPING HOSPITAL OF MILITARY MEDICAL UNIVERSITY IN 2022 WERE SELECTED AND DIVIDED INTO IMPORTED STAPLER GROUP (N = 294) AND DOMESTIC STAPLER GROUP (N = 172) ACCORDING TO DIFFERENT BRANDS OF STAPLER USED DURING OPERATION. IN IMPORTED STAPLER GROUP, THERE WERE 194 MALES AND 100 FEMALES, AGED 21-89 YEARS, WITH MEAN ONE OF (62.77 ± 11.88) YEARS. THE DOMESTIC STAPLER USES QELC AND OTHER MODELS OF UNIVERSAL ENDOSCOPIC LINEAR CUTTER STAPLER AND DISPOSABLE LOADING UNIT (CHONGQING MAIKEWEI MEDICAL TECHNOLOGY CO., LTD.) DEC-B160 AND OTHER MODELS OF DISPOSABLE ENDOSCOPIC LINEAR CUTTER STAPLER AND ASSEMBLY (JIANGSU TRIPLE XINGHAI MEDICAL DEVICES CO., LTD.) AND OTHER 8 BRANDS OF MULTIPLE SPECIFICATIONS; THE IMPORTED STAPLER USES

    Patient problems: Fistula; Hemorrhage/Blood Loss/Bleeding; Nausea; Pain; Restenosis; Swelling/ Edema; Vomiting

    MAUDE report 22544025 ↗
  • MalfunctionMay 29, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING A LUNG RESECTION PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT BE FIRED. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22106429 ↗
  • MalfunctionMay 19, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING A PROCTRECTOMY PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT BE FIRED. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22046565 ↗
  • MalfunctionMay 8, 2025Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE OF THE PANCREAS, IT WAS OBSERVED THAT THE JAW ON THE DEVICE COULD NOT BE OPENED AFTER FIRING. THEY USED RBRB AND MANUAL OVERRIDE TO OPEN THE JAW AND REMOVE THE DEVICE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21985399 ↗
  • MalfunctionApr 18, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21866131 ↗
  • MalfunctionApr 2, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21753160 ↗
  • MalfunctionMar 3, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE STAPLES MALFORMED AFTER FIRING THE DEVICE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21509017 ↗
  • InjuryFeb 4, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED A PATIENT UNDERWENT A GASTRIC SLEEVE PROCEDURE AND IN THE POST-OPERATIVE PERIOD THE PATIENT DEVELOPED A GASTRIC FISTULA DUE TO A FAILURE IN THE STAPLING, REQUIRING HER TO REMAIN IN THE ICU FOR A LONG PERIOD. HOWEVER, IN A SUBSEQUENT CONVERSATION WITH THE SURGEON RESPONSIBLE FOR THE PROCEDURE, SEEKING MORE CONCRETE INFORMATION, IT WAS EXPLAINED THAT THE GASTRIC FISTULA WAS NOT DUE TO EQUIPMENT FAILURE BUT RATHER DUE TO BRONCHOASPIRATION.

    Patient problems: Adhesion(s); Fistula; Insufficient Information

    MAUDE report 21303567 ↗
  • MalfunctionFeb 3, 2025Failure to Fire; Excessive Heating; Failure to Fire; Excessive Heating

    IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, OPENED THE PACKING, BEFORE USED ON THE PATIENT, NOTED THE BATTERY WAS WITH ABNORMAL HIGH TEMPERATURE AND THEN COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21297492 ↗
  • MalfunctionJan 7, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21096345 ↗
  • MalfunctionDec 26, 2024Failure to Form Staple; Firing Problem; Failure to Form Staple; Firing Problem

    IT WAS REPORTED THAT DURING A RADICAL PROCEDURE FOR COLON CANCER, IT WAS OBSERVED THAT THE DEVICE WAS DIFFICULT TO FIRE WHEN SEVERING TISSUE. THEN AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21021222 ↗
  • MalfunctionDec 5, 2024Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING THE SURGERY, COULD NOT FIRE FARTHER AFTER FIRED A LITTLE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20858073 ↗
  • MalfunctionSep 3, 2024Difficult to Open or Close; Activation Failure; Difficult to Open or Close; Activation Failure

    IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THE STAPLER LOCKED, AND IT WAS NOT POSSIBLE TO RETRACT THE BLADE WITH THE RESET BUTTON, SO IT WAS NECESSARY TO OPEN THE DEVICE TO UNLOCK IT. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20127865 ↗
  • MalfunctionAug 20, 2024Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING THE LUNG RESECTION SURGERY, COULD NOT FIRE WITHOUT MOTOR SOUND. THE SURGEON REMOVED THE BATTERY AND ROTATED FOR 180 DEGREE AND REINSERTED THE BATTERY. THE DEVICE STILL COULD NOT FIRE AND WITHOUT MOTOR SOUND. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20002265 ↗
  • MalfunctionAug 9, 2024Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING THE SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19954031 ↗
  • MalfunctionAug 5, 2024Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING THE SURGERY, WHEN SEVERING PULMONARY SEGMENT PARENCHYMA AND BRONCHI ON THE FOURTH FIRING, AFTER FIRED, NOTED THE STAPLES MALFORMED WITH IRREGULAR SHAPE. CHANGED ANOTHER TWO TO USE , THEY STILL HAD THE SAME PROBLEM. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19907757 ↗
  • MalfunctionJun 5, 2024Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING UNK PROCEDURE AFTER FIRED, NOTED THE STAPLES MALFORMED. USED SUTURE TO OVERSEW. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19471410 ↗
  • MalfunctionMay 2, 2024Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING THE SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. USED SUTURE TO OVERSEW. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19232633 ↗
  • MalfunctionApr 4, 2024Difficult to Open or Close; Mechanical Jam; Difficult to Open or Close; Mechanical Jam

    IT WAS REPORTED THAT DURING THE SURGERY, AFTER FIRED, THE KNIFE MOVED TO THE DISTAL END, BUT COULD NOT RETURN KNIFE AUTOMATICALLY. USED MANUAL OVERRIDE LEVER TO RETURN KNIFE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19041483 ↗
  • MalfunctionMar 29, 2024Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19005407 ↗

Other STAPLE, IMPLANTABLE devices

All STAPLE, IMPLANTABLE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.