Skip to content

ECHELON FLEXRecalls & reported problems across all models

Ethicon Endo-Surgery, LLC · 7 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9,393
Deaths reported
59
Injuries reported
1,151
Recalls
1

We found 7 models of ECHELON FLEX by Ethicon Endo-Surgery, LLC in the FDA's device registry (GUDID). Together they are named in 9,393 adverse event reports in the FDA's MAUDE database, received between Feb 26, 2016 and Sep 30, 2026 — classified as 59 deaths, 1,151 injuries, 8,183 malfunctions.

The most frequently reported specific device problem is “Failure to Fire” (3,014 reports).

1 FDA recall record name at least one of these models by its device identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 117 reportsDec 2023: 109 reportsJan 2024: 100 reportsFeb 2024: 113 reportsMar 2024: 83 reportsApr 2024: 91 reportsMay 2024: 66 reportsJun 2024: 93 reportsJul 2024: 104 reportsAug 2024: 110 reportsSep 2024: 139 reportsOct 2024: 136 reportsNov 2024: 133 reportsDec 2024: 85 reportsJan 2025: 110 reportsFeb 2025: 93 reportsMar 2025: 79 reportsApr 2025: 68 reportsMay 2025: 170 reportsJun 2025: 98 reportsJul 2025: 134 reportsAug 2025: 91 reportsSep 2025: 75 reportsOct 2025: 70 reportsNov 2025: 84 reportsDec 2025: 90 reportsJan 2026: 106 reportsFeb 2026: 73 reportsMar 2026: 90 reportsApr 2026: 99 reportsMay 2026: 43 reportsJun 2026: 66 reportsJul 2026: 24 reportsAug 2026: 49 reportsSep 2026: 52 reportsOct 2026: 0 reports202420252026170
Last 36 months. The most recent month may be incomplete.

ECHELON FLEX recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 3, 2021Terminated

    Mislabeled: Labeled on the package as 60mm contains 45mm devices

    Z-2326-2021 · Root cause: Packaging change control · Recall details · FDA record ↗

Most reported ECHELON FLEX problems

Device problems

  • Failure to Fire3,014
  • Difficult to Open or Close2,700
  • Failure to Form Staple1,688
  • Mechanical Jam1,002
  • Adverse Event Without Identified Device or Use Problem756
  • Activation Failure586
  • No Apparent Adverse Event459
  • Material Integrity Problem366

Patient problems

  • No Clinical Signs, Symptoms or Conditions6,038
  • No Consequences Or Impact To Patient2,125
  • Hemorrhage/Blood Loss/Bleeding645
  • No Code Available361
  • Failure to Anastomose280
  • Blood Loss263
  • Insufficient Information104
  • Unspecified Tissue Injury73

ECHELON FLEX models

7 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent ECHELON FLEX adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Ethicon Endo-Surgery, LLC products

All Ethicon Endo-Surgery, LLC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page