Echelon Endopath GST60BRecalls & reported problems
Ethicon Endo-Surgery, LLC · Endopath Echelon Endoscopic Linear Cutter Reloads, Blue, 60mm · Class 2 · Implant
All models of ECHELON ENDOPATH (31)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Endopath Echelon stapler reload appears in 867 adverse event reports, including 8 deaths and 68 injuries. Many describe staples not forming properly or the stapler failing to fire, with some mentioning packaging tears and pieces falling off after firing. Some reports mention bleeding and failed tissue connections in patients. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 861
- Deaths reported
- 8
- Injuries reported
- 68
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 861 adverse event reports that list this exact device (DI 10705036014706), received between Nov 20, 2017 and Sep 21, 2026 — classified as 8 deaths, 68 injuries, 785 malfunctions.
The most frequently reported specific device problem is “Failure to Form Staple” (319 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Form Staple319
- No Apparent Adverse Event203
- Failure to Fire130
- Device Markings/Labelling Problem93
- Tear, Rip or Hole in Device Packaging60
- Adverse Event Without Identified Device or Use Problem44
- Material Integrity Problem43
- Detachment of Device or Device Component19
Patient problems
- No Clinical Signs, Symptoms or Conditions648
- No Consequences Or Impact To Patient144
- Hemorrhage/Blood Loss/Bleeding58
- Failure to Anastomose26
- No Code Available12
- Abscess9
- Blood Loss8
- Foreign Body In Patient6
Recent adverse event reports
The 25 most recent of 861 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2026No Apparent Adverse Event
It was reported that during a Laparoscopic Lobectomy, the surgeon found that a large amount of the staples fell off and scattered in the thoracic cavity after the device firing was completed; and only a small amount of the staples remained in the marginal tissue of the anastomosis. Immediately and completely counted and removed all scattered staples, and repeatedly flushed the thoracic cavity with normal saline to confirm that there is no residual foreign body. Changed another one to continue the surgery but the same problem happened again. Changed another one to continue the surgery, It was reported that during the surgery, could not fire. Changed the new one to complete surgery. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50840139 ↗MalfunctionAug 30, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50514073 ↗MalfunctionAug 30, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50514069 ↗InjuryAug 21, 2026Material Integrity Problem
It was reported that during an unspecified surgical procedure, it was observed that a blue speck fell off after firing. Additional information requested.
Patient problems: Foreign Body In Patient
MAUDE report 50374676 ↗MalfunctionAug 20, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50348690 ↗MalfunctionAug 19, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50331075 ↗MalfunctionAug 19, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50331038 ↗MalfunctionAug 14, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50254421 ↗MalfunctionAug 10, 2026Structural Problem
t was reported that during an unknown procedure, it was observed that some staples fell off when they opened the packaging. Changed to a new one to complete the procedure. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50195238 ↗MalfunctionJun 19, 2026Structural Problem; Difficult to Open or Close; Mechanical Jam
IT WAS REPORTED THAT DURING A LYMPHADENECTOMY WITH RECTOSIGMOIDECTOMY, WHILE STAPLING THE SIGMOID, A GST60B EXHIBITED PREMATURE FIRING. THE SURGEON WAS INSTRUCTED TO REVERSE THE BLADE; AFTER APPROXIMATELY THREE REVERSALS THE STAPLER COULD BE OPENED. ON THE NEXT CARTRIDGE THE FIRST FIRE OCCURRED (FLEX STAPLER) AND THE INSTRUMENT JAMMED AGAIN WITH THE GST60B CARTRIDGE. WE REPEATED THE ENTIRE REVERSAL PROCESS AND THE STAPLER OPENED. THE CARTRIDGE WAS REPLACED WITH ONE FROM A DIFFERENT LOT AND THE SURGERY PROCEEDED. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25565917 ↗MalfunctionJun 16, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25523804 ↗MalfunctionJun 9, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25444613 ↗MalfunctionJun 8, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25428697 ↗MalfunctionMay 29, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25334738 ↗MalfunctionMay 19, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE; IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. THEY USED SUTURE TO OVERSEW. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25216843 ↗MalfunctionMay 4, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25071417 ↗MalfunctionMay 4, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25071313 ↗MalfunctionApr 30, 2026Failure to Fire
IT WAS REPORTED THAT DURING A LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE FARTHER AFTER FIRED A LITTLE. CHANGED THE NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25045158 ↗MalfunctionApr 30, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25043553 ↗MalfunctionApr 29, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE CUSTOMER RECEIVED ALLEGEDLY DEFECTIVE PRODUCTS. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25034320 ↗MalfunctionApr 29, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE WHEN FIRING, THE STAPLES COULDN'T BE DEPLOYED. CHANGED ANOTHER DEVICE TO COMPLETE THE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25027870 ↗MalfunctionApr 27, 2026Failure to Form Staple; No Apparent Adverse Event
IT WAS REPORTED THAT DURING A VATS LOBECTOMY, A STANDARD POWERED STAPLER WITH A BLUE RELOAD WAS USED TO TRANSECT THE PARENCHYMA. THE RESULTING STAPLE LINE WAS OBSERVED TO BE PROPERLY FORMED, WITH NO INEFFICIENCIES OR MALFORMED STAPLES, EXCEPT FOR ONE STAPLE THAT WAS NOT FORMED AT ALL. THERE WAS NO IDENTIFIED EXPLANATION FOR THE SINGLE UNFORMED STAPLE. NO INCIDENT, COMPLICATION, AND NO ADVERSE CONSEQUENCES FOR THE PATIENT. BEFORE AND AFTER THIS EVENT, THE SURGEON PERFORMED ADDITIONAL FIRINGS USING THE SAME DEVICE: ONE FIRING ON THE BRONCHUS USING A GREEN RELOAD, AND A TOTAL OF FIVE FIRINGS USING BLUE RELOADS. ALL STAPLES FROM THESE FIRINGS WERE OBSERVED TO BE CORRECTLY FORMED, WITH THE EXCEPTION OF THE SECOND FIRING, DURING WHICH THE SINGLE UNFORMED STAPLE WAS NOTED. THE UNFORMED STAPLE WAS REMOVED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A FIVE MINUTE DELAY. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25006142 ↗MalfunctionApr 27, 2026Failure to Form Staple; No Apparent Adverse Event
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT SOME STAPLES WERE MALFORMED AFTER FIRING; AND OOZING OCCURRED. CHANGED TO ANOTHER ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 25004087 ↗MalfunctionApr 27, 2026Material Integrity Problem
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE RELOADS LOOKED LIKE SOMETHING WAS MISSING. IT WAS FURTHER REPORTED THAT THE DEVICE FELT DIFFERENT WHEN FIRING; AND THAT THE STAPLE LINE LOOKED OKAY AND PATIENT WENT HOME FINE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25000518 ↗MalfunctionApr 24, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24986960 ↗
FDA clearance and classification
The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the Echelon Endopath GST60B
Has the Echelon Endopath GST60B been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705036014706), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Echelon Endopath GST60B?
The FDA's MAUDE database lists 861 adverse event reports for this device, received between Nov 20, 2017 and Sep 21, 2026. In those reports, the most common specific device problems are “Failure to Form Staple” (319), “Failure to Fire” (130) and “Device Markings/Labelling Problem” (93); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (58), “Failure to Anastomose” (26) and “Abscess” (9). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Echelon Endopath GST60B?
8 reports are classified as deaths, 68 as injuries and 785 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Echelon Endopath GST60B MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Echelon Endopath GST60B, and is it still sold?
The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand Echelon Endopath, model GST60B). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,660 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,727 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,350 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,279 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,006 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,967 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,910 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,044 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page