ECHELON CIRCULAR CDH23PRecalls & reported problems
Ethicon Endo-Surgery, LLC · ECHELON CIRCULAR Powered Stapler · Class 2
All models of ECHELON CIRCULAR (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
83 adverse event reports name this ECHELON CIRCULAR powered stapler, a surgical stapler used to join tissue, with no deaths, 16 injuries and 67 malfunctions reported. Most reports describe the stapler failing to fire during surgery, staples forming incorrectly, or problems ejecting or cutting, sometimes requiring a different device or extra stitches; some reports mention bleeding at the staple line. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 83
- Deaths reported
- 0
- Injuries reported
- 16
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 83 adverse event reports that list this exact device (DI 10705036015369), received between Feb 4, 2020 and Jun 29, 2026 — classified as 16 injuries, 67 malfunctions.
The most frequently reported specific device problem is “Failure to Fire” (50 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Fire50
- Failure to Form Staple8
- Ejection Problem7
- Adverse Event Without Identified Device or Use Problem6
- Failure to Cut4
- Patient Device Interaction Problem4
- Mechanical Jam3
- Self-Activation or Keying3
Patient problems
- No Clinical Signs, Symptoms or Conditions71
- Failure to Anastomose4
- Hemorrhage/Blood Loss/Bleeding4
- Insufficient Information4
- Obstruction/Occlusion2
- Stenosis2
- Unspecified Tissue Injury2
- No Code Available1
Recent adverse event reports
The 25 most recent of 83 reports, in the reporter's own words as published by the FDA.
InjuryJun 29, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURES, FOUR DIFFERENT PATIENTS HAD TO BRING BACK FOR BLEEDING AT THE STAPLE LINE. THE PATIENTS WERE DOING FINE RIGHT NOW.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 25653591 ↗InjuryJun 22, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURES, FOUR DIFFERENT PATIENTS HAD TO BRING BACK FOR BLEEDING AT THE STAPLE LINE. THE PATIENTS WERE DOING FINE RIGHT NOW.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 25585432 ↗InjuryJun 19, 2026Failure to Fire; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, DURING ESOPHAGOJEJUNOSTOMY, DURING USE FOR ESOPHAGOJEJUNOSTOMY, THE INDICATOR WAS ILLUMINATED, BUT THE DEVICE COULD NOT BE FIRED. DUE TO A CHANGE IN THE ANASTOMOTIC TECHNIQUE FOR THE ESOPHAGOJEJUNOSTOMY (FROM CIRCULAR PROCEDURE TO OVERLAP PROCEDURE USING AN END CUTTER), ADDITIONAL RESECTION WAS REQUIRED AND THE OPERATION TIME WAS PROLONGED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Unspecified Tissue Injury
MAUDE report 25561448 ↗MalfunctionJun 11, 2026Failure to Fire
IT WAS REPORTED THAT, DURING USE FOR ESOPHAGEAL ANASTOMOSIS IN OPEN THORACIC SURGERY, FIRING COULD NOT BE PERFORMED. THE ANVIL WAS LEFT IN PLACE, AND THE SURGERY WAS COMPLETED SUCCESSFULLY USING ANOTHER DEVICE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25476893 ↗MalfunctionMay 23, 2026Failure to Fire
IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25264717 ↗MalfunctionApr 11, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING THE ESOPHAGUS SURGERY SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. SUTURE WAS USED TO OVERSEW. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24852633 ↗MalfunctionApr 2, 2026Failure to Fire
IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24770269 ↗MalfunctionJan 23, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24167444 ↗InjuryJan 21, 2026Patient Device Interaction Problem
THIS COMPLAINT IS FROM A LITERATURE SOURCE. THE FOLLOWING LITERATURE CITE HAS BEEN REVIEWED: SAITO A, OTSUKA K, GOTO S, ARIYOSHI T, YAMASHITA T, MOTEGI K, KOMOTO M, KISHIMOTO Y, MURAKAMI M, AOKI T. COMPARISON OF ANASTOMOTIC STRICTURE RATES BETWEEN 23- AND 25-MM POWERED CIRCULAR STAPLERS IN CERVICAL ESOPHAGOGASTRIC ANASTOMOSIS: A PROPENSITY-MATCHED STUDY. ESOPHAGUS. 2025 DEC 22. DOI: 10.1007/S10388-025-01171-2. EPUB AHEAD OF PRINT. THIS STUDY AIMED TO COMPARE STENOSIS RATES BETWEEN NEWLY INTRODUCED 23- AND 25-MM POWERED CIRCULAR STAPLER (CSS) IN CERVICAL ESOPHAGOGASTRIC ANASTOMOSIS. FROM (B)(6) 2022 TO (B)(6) 2024, 126 PATIENTS WHO UNDERWENT THORACOSCOPIC MCKEOWN ESOPHAGECTOMY WITH RETROSTERNAL GASTRIC CONDUIT RECONSTRUCTION WERE RETROSPECTIVELY ANALYZED. THEY WERE CATEGORIZED INTO THE 23 MM (N = 52) AND 25-MM (N = 74) CS GROUPS. REPORTED COMPLICATIONS INCLUDE THE INCIDENCE OF ANASTOMOTIC STRICTURE WAS 18.0% (7 CASES) AND 12.8% (5 CASES) IN THE 23- AND 25-MM CS GROUPS, RESPECTIVELY. THE INCIDENCE OF ANASTOMOTIC LEAKAGE AND THE LEAKAGE LOCATION REVEALED NO SIGNIFICANT DIFFERENCES BETWEEN THE TWO GROUPS. IN CONCLUSION, 23- AND 25-MM POWERED CSS ACHIEVED COMPARABLE ANASTOMOTIC STRICTUR
Patient problems: Failure to Anastomose; Obstruction/Occlusion; Stenosis
MAUDE report 24137099 ↗InjuryJan 21, 2026Patient Device Interaction Problem
THIS COMPLAINT IS FROM A LITERATURE SOURCE. THE FOLLOWING LITERATURE CITE HAS BEEN REVIEWED: SAITO A, OTSUKA K, GOTO S, ARIYOSHI T, YAMASHITA T, MOTEGI K, KOMOTO M, KISHIMOTO Y, MURAKAMI M, AOKI T. COMPARISON OF ANASTOMOTIC STRICTURE RATES BETWEEN 23- AND 25-MM POWERED CIRCULAR STAPLERS IN CERVICAL ESOPHAGOGASTRIC ANASTOMOSIS: A PROPENSITY-MATCHED STUDY. ESOPHAGUS. 2025 DEC 22. DOI: 10.1007/S10388-025-01171-2. EPUB AHEAD OF PRINT. THIS STUDY AIMED TO COMPARE STENOSIS RATES BETWEEN NEWLY INTRODUCED 23- AND 25-MM POWERED CIRCULAR STAPLER (CSS) IN CERVICAL ESOPHAGOGASTRIC ANASTOMOSIS. FROM MAY 2022 TO FEBRUARY 2024, 126 PATIENTS WHO UNDERWENT THORACOSCOPIC MCKEOWN ESOPHAGECTOMY WITH RETROSTERNAL GASTRIC CONDUIT RECONSTRUCTION WERE RETROSPECTIVELY ANALYZED. THEY WERE CATEGORIZED INTO THE 23 MM (N = 52) AND 25-MM (N = 74) CS GROUPS. REPORTED COMPLICATIONS INCLUDE THE INCIDENCE OF ANASTOMOTIC STRICTURE WAS 18.0% (7 CASES) AND 12.8% (5 CASES) IN THE 23- AND 25-MM CS GROUPS, RESPECTIVELY. THE INCIDENCE OF ANASTOMOTIC LEAKAGE AND THE LEAKAGE LOCATION REVEALED NO SIGNIFICANT DIFFERENCES BETWEEN THE TWO GROUPS. IN CONCLUSION, 23- AND 25-MM POWERED CSS ACHIEVED COMPARABLE ANASTOMOTIC STRICTURE
Patient problems: Failure to Anastomose; Obstruction/Occlusion; Stenosis
MAUDE report 24137086 ↗MalfunctionOct 3, 2025Ejection Problem; Ejection Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, NOTED THE JAW LOOSE AND COULD NOT CLAMP THE TISSUE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23208384 ↗MalfunctionAug 13, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22794257 ↗MalfunctionJul 16, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN ESOPHAGECTOMY SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22519127 ↗MalfunctionMay 19, 2025Ejection Problem; Ejection Problem
IT WAS REPORTED THAT DURING AN ESOPHAGEAL SURGERY, ANVIL COULD NOT LOCKED ON TROCAR AFTER FIRED. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22044439 ↗MalfunctionApr 1, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT DURING A TOTAL GASTRECTOMY, THE GREEN CHECK MARK ILLUMINATED WHEN THE BATTERY WAS INSERTED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21738184 ↗InjuryMar 31, 2025Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Failure to Eject; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Failure to Eject
IT WAS REPORTED THAT DURING A MEDIASTINOSCOPIC SUBTOTAL ESOPHAGECTOMY, THE KNOB DID NOT TURN AND THE ANASTOMOSIS COULD NOT BE COMPLETED WHEN THE ANVIL AND THE DEVICE WERE CONNECTED DURING ANASTOMOSIS AND THE KNOB WAS TURNED TO CLOSE. CDH23P WAS USED FOR ANASTOMOSIS OF THE CERVICAL ESOPHAGUS AND GASTRIC TUBE. ADDITIONAL RESECTION WAS PERFORMED AND THE PROCEDURE WAS CHANGED TO HAND-SEWN ANASTOMOSIS BECAUSE THE ANVIL AND THE DEVICE COULD NOT BE SEPARATED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Tissue Injury
MAUDE report 21730741 ↗MalfunctionFeb 6, 2025Ejection Problem; Ejection Problem
IT WAS REPORTED THAT DURING A THORACOSCOPIC ESOPHAGECTOMY, THE DEVICE COULD NOT CLOSE AFTER THE ANVIL AND THE TROCAR WERE CONNECTED IN THE ESOPHAGUS FOR CERVICAL ANASTOMOSIS. AND ALSO THE ANVIL AND THE TROCAR COULD NOT SEPARATE, AND CDH WAS REMOVED BY RESECTION. ADDITIONAL HAND SUTURE WAS PERFORMED TO COMPLETE THE CASE. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21326405 ↗MalfunctionJan 28, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A GASTRIC PROCEDURE SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21255378 ↗MalfunctionJan 6, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING A RECTUM PROCEDURE SURGERY WHEN FIRING, THE STAPLES COULDN'T BE DEPLOYED. CHANGED ANOTHER DEVICE TO COMPLETE THE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21079226 ↗MalfunctionAug 23, 2024Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20054608 ↗MalfunctionAug 7, 2024Self-Activation or Keying; Self-Activation or Keying
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE CDH23P WAS FIRED EVEN THE NODE WAS NOT CLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19928506 ↗MalfunctionJul 15, 2024Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING A LAPAROSCOPIC ESOPHAGECTOMY, USED DURING ANASTOMOSIS OF ESOPHAGUS AND STOMACH. THE DEVICE COULD NOT BE FIRED WITH PULLING THE FIRING KNOB. NO ELECTROCAUTERY WAS USED IMMEDIATELY PRIOR TO THE PROCEDURE. THERE WAS NOTHING OUT OF THE ORDINARY INCLUDING THE SETTING. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19743951 ↗MalfunctionJul 8, 2024Self-Activation or Keying; Self-Activation or Keying
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE WAS FIRED EVEN THE FIRING TRIGGER WAS NOT PRESSED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19689137 ↗InjuryMay 13, 2024Use of Device Problem; Failure to Fire; Adverse Event Without Identified Device or Use Problem; Use of Device Problem; Failure to Fire; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, AFTER THE FIRE, NOTED THE DEVICE CUT THE TISSUE WITHOUT STAPLE DEPLOYED. CHANGED TO OPEN OPERATION TO COMPLETE THE SURGERY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 19297603 ↗MalfunctionMay 6, 2024Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN INTESTINAL SURGERY THE DEVICE COULD NOT FIRE. CHANGED TO THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19250246 ↗
FDA clearance and classification
The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the ECHELON CIRCULAR CDH23P
Has the ECHELON CIRCULAR CDH23P been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705036015369), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ECHELON CIRCULAR CDH23P?
The FDA's MAUDE database lists 83 adverse event reports for this device, received between Feb 4, 2020 and Jun 29, 2026. In those reports, the most common specific device problems are “Failure to Fire” (50), “Failure to Form Staple” (8) and “Ejection Problem” (7); the most common patient problems are “Failure to Anastomose” (4), “Hemorrhage/Blood Loss/Bleeding” (4) and “Obstruction/Occlusion” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ECHELON CIRCULAR CDH23P?
None of the reports are classified as deaths, 16 as injuries and 67 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ECHELON CIRCULAR CDH23P MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the ECHELON CIRCULAR CDH23P, and is it still sold?
The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand ECHELON CIRCULAR, model CDH23P). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Staple, implantable devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,660 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,727 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,350 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,279 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,006 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,967 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,910 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,044 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.