Bard® X-Force® Nephrostomy Balloon Dilation Catheter 996101Recalls & reported problems
C. R. Bard, Inc. · Bard® X-Force® Nephrostomy Balloon Dilation Catheter · Class U
All models of Bard X-Force Nephrostomy Balloon Dilation Catheter (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Bard X-Force nephrostomy balloon catheter is used to widen the tract for kidney drainage procedures. Reports mostly describe balloons bursting, developing holes or leaks and failing to inflate or stay inflated, along with missing instruction sheets and damaged pumps. There are 24 reports naming its identifier, with no deaths and 2 injuries mentioned. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 10801741127660), received between Aug 9, 2017 and Dec 9, 2025 — classified as 2 injuries, 22 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Burst Container or Vessel7
- Inflation Problem4
- Material Puncture/Hole4
- Break3
- Component Missing3
- Fluid/Blood Leak2
- Use of Device Problem2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- No Known Impact Or Consequence To Patient9
- No Consequences Or Impact To Patient7
- No Patient Involvement3
- Hemorrhage/Blood Loss/Bleeding2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Information1
- Perforation1
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 9, 2025Component Missing; Component Missing
IT WAS REPORTED THAT THE CUSTOMER ORDERED 2 BOXES REF: (B)(4), LOT: BMKZFM16, CUSTOMER NORMALLY RECEIVES A GREEN BOX, HOWEVER THEY RECEIVED A BLUE BOX WITH THE SAME MATERIAL AND LOT NUMBER LISTED ON IT. THE GREEN BOX HAS THE PTA SHEET IN IT, WHILE THE BLUE BOX RECEIVED, HAS PICTURES SHOWING THE SHEET WAS SUPPOSED TO BE IN IT. HOWEVER, THEY WERE NO SHEETS CONTAINED IN THE TWO BOXES. CUSTOMER WOULD LIKE SOMEONE REACH OUT TO THEM AS SOON AS POSSIBLE AS THEY HAD AN UPCOMING PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23764877 ↗MalfunctionNov 6, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT PRIOR TO PERFORMING URETERAL STRICTURE DILATION SURGERY, WHEN OPENING THE URETERAL STRICTURE BALLOON FOR USE, A NOTICEABLE LEAK WAS OBSERVED IN THE PRESSURE PUMP. REPLACED THE BALLOON SET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23485367 ↗MalfunctionApr 10, 2025Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT DURING USE, BALLOON DILATION CATHETER BALLOON DEFLATED. AN ISSUE THEY ENCOUNTERED WITH ONE OF NEPHROSTOMY BALLOON DILATION CATHETER BALLOONS DURING A PERCUTANEOUS NEPHROLITHOTOMY PROCEDURE. AFTER FULL INFLATION, THEY OBSERVED A GRADUAL DECREASE IN PRESSURE, AND UPON FURTHER INSPECTION, THEY DISCOVERED A HOLE IN THE BALLOON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21813678 ↗MalfunctionOct 19, 2023Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT THE CUSTOMER WAS IN PROCEDURE ON 12JUNE, THE BALLOON DILATION DEVICE WOULD NOT LOCK. SEPARATE KIT OPENED TO OBTAIN DILATION PART. THERE WAS ONLY A DELAY IN CARE NO PATIENT HARM. THE CUSTOMER WAS GOING TO CONTACT CAROL PRICE TERRITORY MANAGER FOR REPLACEMENT PRODUCT. CUSTOMER HAD AFFECTED PRODUCT FOR INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17966819 ↗MalfunctionOct 9, 2023Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT ONCE THE PACKAGE WAS OPENED, THE DOCTOR FOUND THAT THE BALLOON PUMP WAS DAMAGED ON THE BALLOON DILATION CATHETER. THE DEVICE WAS NOT USED IN THE PATIENT. IT WAS STATED THAT THEY REPLACED WITH A NEW BALLOON. PER ADDITIONAL INFORMATION VIA EMAIL FROM IBC ON 26SEP2023, THE BALLOON PUMP INDICATING BALLOON OF THE DILATION CATHTER AND THE DAMAGE INDICATING BALLON BURST.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17894825 ↗InjuryJun 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT IN AN ONLINE SURVEY A PHYSICIAN STATED A PERFORATION ADVERSE EVENT, DIRECTLY ATTRIBUTABLE TO THE DEVICE 996101 - X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER, 10MM X 15CM, W/ INFLATION DEVICE, STERILE. ALSO STATING THAT THE PATIENT WITH THE DEVICE 996101 - X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER, 10MM X 15CM, W/ INFLATION DEVICE, STERILE PLACEMENT DID RESULT IN PATIENT INJURY REQUIRING MEDICAL OR SURGICAL INTERVENTION BECAUSE SURGICAL REPAIR. PER FOLLOW UP INFORMATION RECEIVED VIA EMAIL ON 14JUN2023, IN AN ONLINE SURVEY, A PHYSICIAN STATED A PERFORATION ADVERSE EVENT, DIRECTLY ATTRIBUTABLE TO THE DEVICE 996101 - X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER, 10MM X 15CM, W/ INFLATION DEVICE, STERILE. PHYSICIAN ALSO INDICATED PERFORATION WAS PRESENT PRIOR TO PLACEMENT OF THE DEVICE. ALSO STATING THAT THE PATIENT WITH THE DEVICE 996101 - X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER, 10MM X 15CM, W/ INFLATION DEVICE, STERILE PLACEMENT DID RESULT IN PATIENT INJURY REQUIRING MEDICAL OR SURGICAL INTERVENTION BECAUSE SURGICAL REPAIR.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Perforation
MAUDE report 17208560 ↗MalfunctionMay 25, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT A PHYSICIAN IN AN ONLINE SURVEY STATED THAT THEY WERE NOT ABLE TO SUCCESSFULLY DILATE THE NEPHROSTOMY TRACT. THE PHYSICIAN WAS NOT ABLE TO SUCCESSFULLY DILATE THE NEPHROSTOMY TRACT FOR THE DEVICE 996101 - BALLOON DILATION CATHETER BECAUSE OF PERSONAL EXPECTATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17002019 ↗MalfunctionApr 13, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AFTER OPENING THE PRODUCT, WHEN THE CUSTOMER STARTED TO INFLATE THE BALLOON, THE DOCTOR NOTICED THE BALLOON WAS LEAKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16732507 ↗MalfunctionJan 27, 2022Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT THE BALLOON IN THE PERCUTANEOUS DILATION CATHETER HAD A HOLE IN IT. IT WAS STATED THAT THE DOCTOR HAD TO RESTART WITH A DIFFERENT BALLOON FROM A COMPETITOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13374864 ↗MalfunctionDec 15, 2021Nonstandard Device; Nonstandard Device
IT WAS REPORTED THAT THE BALLOON IN THE DILATION CATHETER WAS PINCHED. CUSTOMER HAD TO OPEN A NEW BALLOON AND THEN IT WORKED FINE. STATED THE PRODUCT WAS DEFECTIVE AND NO IMPACT TO THE PATIENT. THE BALLOON WAS PINCHED IN THE MIDDLE WHILE INFLATING DURING A PERCUTANEOUS NEPHROLITHONOMY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13010764 ↗MalfunctionJul 28, 2021Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT THE X-FORCE N30 NEPHROSTOMY BALLOON DILATION CATHETER BALLOON WAS FOUND DAMAGED WHEN THE PACKAGE WAS OPENED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12239197 ↗MalfunctionJun 22, 2021Failure to Capture; Failure to Capture
IT WAS REPORTED THAT THE DOCTOR CANNOT SEE THE BALLOON ON THE FLUORO IMAGING WHEN THE CONTRAST WAS INJECTED INTO THE BALLOON. NO IMPACT TO THE PATIENT AND DOCTOR OPENED NEW BALLOON AND CASE WENT WELL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12041632 ↗MalfunctionFeb 10, 2020Break; Material Puncture/Hole; Break
IT WAS REPORTED THAT WHILE TRYING TO TIGHTEN THE X-FORCE DEVICE, A HOLE WAS DISCOVERED.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 9687692 ↗MalfunctionFeb 8, 2020Burst Container or Vessel; Component Missing; Burst Container or Vessel
IT WAS REPORTED THAT PART OF THE BALLOON WAS MISSING.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9684979 ↗MalfunctionDec 20, 2019Use of Device Problem; Inflation Problem; Material Puncture/Hole; Use of Device Problem
IT WAS REPORTED THAT THERE WAS A HOLE IN THE BALLOON ON THE INFLATION DEVICE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9503842 ↗MalfunctionJun 20, 2019Inflation Problem; Inflation Problem; Infusion or Flow Problem; Physical Resistance/Sticking
IT WAS REPORTED THAT THE BALLOON WAS USED IN PCNL SURGERY, THE PUMP PUSHER CANNOT PUSH, AND THE FLUID CANNOT BE INJECTED INTO THE BALLOON.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8718988 ↗MalfunctionJun 18, 2019Use of Device Problem; Leak/Splash; Use of Device Problem
IT WAS REPORTED THAT THE BALLOON WAS USED DURING A PCNL SURGERY AND LEAKAGE OCCURRED FROM THE PUMP TO THE BALLOON. IT WAS ALSO NOTED THAT FLUID COULD NOT BE INJECTED INTO THE BALLOON.
Patient problems: No Consequences Or Impact To Patient; No Information
MAUDE report 8708256 ↗MalfunctionFeb 14, 2019Burst Container or Vessel; Material Rupture; Burst Container or Vessel
IT WAS REPORTED THAT THE BALLOON RUPTURED UPON INFLATION.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8340010 ↗MalfunctionFeb 14, 2019Break
IT WAS REPORTED THAT UPON PRESSING THE BALLOON WITH A PUMP, THE PRESSURE GAUGE WAS FOUND BROKEN.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 8339621 ↗MalfunctionJan 29, 2019Component Missing
IT WAS REPORTED THAT A COMPONENT WAS MISSING FROM THE XFORCE PACKAGING. ADDITIONAL INFORMATION WAS RECEIVED FROM THE COMPLAINANT VIA EMAIL ON 16JAN2019, THAT THE INFLATION DEVICE WAS MISSING FROM THE PACKAGING.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 8285440 ↗MalfunctionMay 29, 2018Break; Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT THE BALLOON WAS DAMAGED WHILE DILATING PRESSURE WAS BEING APPLIED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7550283 ↗InjuryDec 12, 2017Burst Container or Vessel
IT WAS REPORTED THAT THE BALLOON HAD BURST WITH A SMALL AMOUNT OF SELF LIMITED BLEEDING. THE PROCEDURE WAS URETHRAL STRICTURE BALLOON DILATATION SURGERY, PRESSURE TO 25 ATM.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Consequences Or Impact To Patient
MAUDE report 7107949 ↗MalfunctionAug 31, 2017Burst Container or Vessel
IT WAS REPORTED THAT THE URETHRAL STRICTURE BALLOON COMPRESSION WAS 25 ATMOSPHERIC PRESSURE, AND THE BALLOON BURST. A SMALL AMOUNT OF BLEEDING WAS OBSERVED IN THE PATIENT. HOWEVER, THERE WERE NO MISSING PIECES FROM THE BALLOON.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 6835075 ↗MalfunctionAug 9, 2017Inflation Problem
IT WAS REPORTED THAT THE BALLOON WOULD NOT INFLATE PROPERLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6780503 ↗
Other CATHETER, NEPHROSTOMY devices
- Cook G09501Cook Incorporated · CATHETER, NEPHROSTOMY280 reports
- Dawson Mueller G09705Cook Incorporated · CATHETER, NEPHROSTOMY2 deaths215 reports
- Cook G09497Cook Incorporated · CATHETER, NEPHROSTOMY129 reports
- Dawson Mueller G11020Cook Incorporated · CATHETER, NEPHROSTOMY100 reports
- Cook G09502Cook Incorporated · CATHETER, NEPHROSTOMY86 reports
- NephroMax Kit M0062101180Boston Scientific Corporation · CATHETER, NEPHROSTOMY1 recall25 reports
- NephroMax Kit M0062101600Boston Scientific Corporation · CATHETER, NEPHROSTOMY1 recall11 reports
- Dawson Mueller G13768Cook Incorporated · CATHETER, NEPHROSTOMY59 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page