Zilver 635 G23816Recalls & reported problems
Cook Ireland Ltd · Zilver 635 Biliary Stent, Expandable Metal Biliary Stent System · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10827002238166), received between Oct 2, 2020 and Jul 17, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break3
- Activation, Positioning or Separation Problem2
- Material Integrity Problem2
- Activation Problem1
- Improper or Incorrect Procedure or Method1
- Material Deformation1
- Off-Label Use1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 17, 2023Improper or Incorrect Procedure or Method; Activation, Positioning or Separation Problem; Improper or Incorrect Procedure or Method; Activation, Positioning or Separation Problem
SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION ON 25-OCT-2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17333000 ↗MalfunctionApr 3, 2023Break; Material Integrity Problem; Break; Material Integrity Problem
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE COMPLETION OF THE INVESTIGATION ON 24-MAY-2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16662758 ↗MalfunctionApr 3, 2023Break; Material Integrity Problem; Break; Material Integrity Problem
SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION ON 25-MAY-23.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16662756 ↗MalfunctionMay 11, 2021Break; Activation, Positioning or Separation Problem; Activation Problem; Break; Activation, Positioning or Separation Problem; Activation Problem
SUPPLEMENTAL FOLLOW-UP REPORT IS BEING SUBMITTED DUE TO THE RECEIPT ADDITIONAL INFORMATION AND RECEIPT OF THE COMPLAINT DEVICE FOR EVALUATION. ADDITIONAL INFORMATION RECEIVED 17-MAY-2021 AS FOLLOWS: THE REP'S REPLY ON ADDITIONAL INFORMATION REQUEST WAS RECEIVED FROM THE REP ON 17/5/21 JIL01: 1. ARE IMAGES OF THE DEVICE OR PROCEDURE AVAILABLE? NO. 2. AT WHAT STAGE OF THE PROCEDURE DID THE COMPLAINT OCCUR? DURING STENT PLACEMENT . 3. WAS A SPHINCTEROTOMY PERFORMED PRIOR TO USE OF THE STENT DEVICE? YES. 4. WAS BALLOON DILATION PERFORMED PRIOR TO USE OF THE STENT DEVICE? NO. 5. WHAT WAS THE LENGTH AND DIAMETER OF THE STRICTURE? UNKNOWN. 6. WHAT BRAND OF ENDOSCOPE WAS USED WITH THE DEVICE? OLYMPUS. 7. WAS THE PRODUCT INSPECTED FOR KINKS OR DAMAGE BEFORE USE? YES. 8. WHAT WAS THE POSITION OF THE ELEVATOR DURING DEPLOYMENT? , OPEN. 9. WAS THE DELIVERY SYSTEM TRACKED AROUND A TIGHT ANGLE IN THE PATIENT ANATOMY OR AROUND A TIGHT ANGLE IN AN ENDOSCOPE? , YES. 10. WAS THE DEVICE FLUSHED THROUGH BOTH FLUSHING PORTS BEFORE THE PROCEDURE, AS PER IFU? YES, 11. DETAILS OF THE WIRE GUIDE USED (NAME, DIAMETER, HYDROPHYLLIC)? UNKNOWN. 12. WHAT WAS THE TARGET LOCATION FOR THE STENT? HILAR. 13. WAS THE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11804439 ↗MalfunctionOct 2, 2020Off-Label Use; Material Deformation; Off-Label Use; Material Deformation
THE PATIENT PREVIOUSLY HAD CHOLEDOCHOJEJUNOSTOMY UNDERWENT METALIC BILIARY STENT PLACEMENT. THE USER ADVANCED THE DELIVERY SYSTEM TO THE HEPATIC PORTAL REGION USING OLYMPUS' SINGLE BALLOON SHORT TIE SCOPE (SIF-290), AND ATTEMPTED TO DEPLOY THE STENT. HOWEVER, THE HANDLE WAS TOO TIGHT AND WOULD NOT MOVE/PULL AND THE DISTAL SIDE OF THE SHEATH GOT DAMAGED DURING ATTEMPT OF DEPLOYMENT. HE REPLACED IT WITH ANOTHER ZILVER 635 BILIARY SELF EXPANDING METAL STENT(LOT UNKNOWN) AND PLACED THE STENT AT THE TARGET SITE SUCCESSFULLY. THE HANDLE OF THIS REPLACEMENT WAS FAIRLY TIGHT AS WELL. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. PREFIX ZILBS: WAS A SPHINCTEROTOMY OR BALLOON DILATION PERFORMED PRIOR TO USE OF THE STENT DEVICE? - NO. WAS POST-DILATION PERFORMED AFTER THE PLACEMENT OF THE STENT? - NO. WHAT BRAND OF ENDOSCOPE WAS USED WITH THE DEVICE? OLYMPUS' SIF-290. WHAT WAS THE TARGET LOCATION FOR THE COMPLAINT DEVICE? HEPATIC PORTAL. REGION WAS THE DEVICE FLUSHED THROUGH BOTH FLUSHING PORTS BEFORE THE PROCEDURE, AS PER IFU? YES. DETAILS OF THE WIRE GUIDE USED (NAME, DIAMETER, HYRDOPHYLLIC)? ASKU. WAS RESISTANCE ENCOUNTERED WHEN ADVANCING THE WIRE GUIDE OR DELI
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10618561 ↗
Other Stents, drains and dilators for the biliary ducts devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.